Early in a program, someone asks which submission you will file, and the budget, the clinical plan and the launch date all wait on the answer. The choice between a 510(k), a De Novo request and a PMA is mostly made for you by the device type. The work is reading the evidence correctly.

The short answer: file a 510(k) when a legally marketed predicate with the same intended use exists. File a De Novo request when no predicate exists and general or special controls can give reasonable assurance of safety and effectiveness (Class I or II). Expect a PMA when the device type is Class III. The fastest way to see which path similar devices took is the Submission Type on their product code's classification record.

This post compares the three pathways side by side, explains when each applies, and shows how to check the path for a device type. For what the words cleared, approved and granted mean, see FDA cleared vs approved.

The three pathways side by side

Each pathway has a different legal question, and that question drives the evidence and the time.

510(k)De NovoPMA
Legal questionSubstantially equivalent to a legally marketed predicate? (21 CFR 807.100(b))Can general or special controls give reasonable assurance of safety and effectiveness? (21 CFR 860.260)Reasonable assurance of safety and effectiveness on the device's own data? (21 CFR Part 814)
Typical classII (some I, some preamendment III)I or II, new device typeIII
Evidence coreComparison to the predicate, plus performance testingDevice description, benefit-risk and proposed controlsNonclinical studies and clinical investigations (814.20(b))
OutcomeSE order ("cleared"), K numberGranting order, DEN number, new classification regulationApproval order, P number
MDUFA V review goal90 FDA days for 95%150 FDA days for 70%180 FDA days for 90% (320 with an advisory panel)
Average total days, FY2023 cohort141270294 (no panel)
FY2027 fee, standard / small business$28,653 / $7,163$191,020 / $47,755$636,732 / $159,183

The review goals come from the MDUFA V commitment letter. "FDA days" count only time a submission is under FDA review, so they leave out the time a company spends answering questions. The average total days come from the FDA's May 19, 2026 MDUFA V performance report, with data through March 31, 2026, and include both FDA and company time. The fees are the FDA's FY2027 rates, in effect from October 1, 2026 to September 30, 2027.

When a 510(k) applies

A 510(k) applies when your device has the same intended use as a legally marketed device and its technological differences, if any, don't raise different questions of safety and effectiveness. Under 807.81(a), it is due at least 90 days before you begin commercial distribution.

Class alone doesn't decide it. Polymer patient examination gloves (LZA) are Class 1, yet their classification record lists the Submission Type as 510(k). Infusion stands (FOX) are Class 1 and "510(K) Exempt". And the FDA's PMA page notes that some preamendment Class III devices still go through a Class III 510(k).

The 510(k) is the cheapest and usually the fastest route, but it ties your device to a predicate. If the closest predicate has a different intended use, the answer may be not substantially equivalent, which leads to the next pathway.

When De Novo applies

De Novo applies to a new device type with no legally marketed predicate that can still be Class I or II. Under 21 CFR 860.200, you can file it after a not substantially equivalent decision on a 510(k), or directly when you determine no predicate exists.

Under 860.220(a)(14)(ii), the request must show that the probable health benefit of using the device outweighs any probable injury or illness, and it proposes the controls that would make the device reasonably safe. Under 860.260(c), the FDA may decline if the device doesn't meet Class I or II, if the type is already classified, or if the type has already been approved through a PMA.

The payoff is that a grant creates a classification regulation, and the FDA's De Novo page says the device can then serve as a predicate for later 510(k)s. The integrated continuous glucose monitor is the standard example: DEN170088 created product code QBJ in 2018, and today the QBJ classification record lists Submission Type 510(k).

When a PMA applies

A PMA applies when the device type is Class III, which 21 CFR 860.3 defines as the class for which premarket approval is or will be required. The FDA's PMA page describes them as devices that support or sustain life, are of substantial importance in preventing impairment of health, or present a potential unreasonable risk.

Plan for clinical data. Section 814.20(b) requires separate sections on nonclinical laboratory studies and clinical investigations involving human subjects. Under 814.40, the FDA reviews a filed PMA within 180 days unless a major amendment arrives, but the calendar time is longer: P210015, an automated external defibrillator under product code MKJ, was received April 6, 2021 and approved October 31, 2022, 573 days later.

A PMA also brings supplements for later changes, so the approval history of a device type includes more than its original approvals.

How to see which path similar devices took

The classification record answers this for a device type in one field.

  1. Find the product code. Search the FDA's Product Classification database by device name or regulation number. Our product code guide covers the search.
  2. Read Device Class and Submission Type. JWH, a cemented knee, reads Class 2, 510(k). MKJ reads Class 3, PMA.
  3. Search the database that matches the Submission Type by product code, sorted by decision date, to see the recent decisions.
  4. Check the De Novo database for the same code. A code created by a De Novo grant shows how the first device of the type got to market.
  5. Note the decision dates and received dates of the closest devices. They tell you more about your likely timeline than the goal in the table above.
  6. If no code fits your device, that may point toward De Novo. A 513(g) request for classification information or a Pre-Submission is the formal way to ask the FDA.

The class rules behind step 2 are explained in Class II medical device: how the FDA's three risk classes work.

DeviceVigil puts steps 2 and 3 in one place. On the FDA's 510(k), De Novo, PMA and classification pages, its side panel shows the product code's class, regulation and review panel, with the code's clearance or PMA history per year, the latest decisions and the median review time. Hovering over a K, DEN or P number previews that record.

See it beside the FDA record

DeviceVigil is a Chrome and Edge extension for regulatory, quality and post-market teams. Open a 510(k), PMA, classification or recall page and see that device type’s clearances, recalls and MAUDE reports in one panel, and get told when something changes.

Add to Chrome — free See how it works

Frequently asked questions

What is the difference between a PMA and a 510(k)?

A 510(k) shows that a device is substantially equivalent to a legally marketed predicate. A PMA must show reasonable assurance of safety and effectiveness on the device's own data, usually including clinical investigations. The FY2027 standard fee is $28,653 for a 510(k) and $636,732 for a PMA.

Is De Novo faster than a PMA?

The review goals are shorter: 150 FDA days for 70% of De Novo requests against 180 FDA days for 90% of PMAs without a panel. In the FDA's FY2023 cohort, De Novo requests averaged 270 total days and PMAs without a panel 294. Your own timeline depends on how many rounds of questions the review takes.

Can a De Novo device be used as a predicate?

Yes. The FDA says a device classified through De Novo can serve as a predicate for future 510(k)s of the same type. QBJ shows it: K183206, cleared in February 2019, names the 2018 De Novo DEN170088 as its predicate.

How much does a 510(k) cost in FDA fees?

For FY2027 (October 1, 2026 to September 30, 2027), the 510(k) user fee is $28,653, or $7,163 for a qualifying small business. The FDA publishes new rates each year in the Federal Register.

Do all Class III devices need a PMA?

Most do. The FDA's PMA page notes that some preamendment Class III device types still use a Class III 510(k). The Submission Type on the product code's classification record tells you which applies.

Sources