A new device is heading for a 510(k), and the whole submission rests on one choice: which legally marketed device you claim it is substantially equivalent to. The FDA's 2023 draft guidance asks submitters to look past the legal test, at the predicate's recalls, adverse events and test methods.
A predicate device is the legally marketed device you compare yours to in a 510(k): one marketed before May 28, 1976, one reclassified from Class III to Class II or I, or one already found substantially equivalent through a 510(k) (21 CFR 807.92(a)(3)). To find candidates, start from your product code, search the FDA's 510(k) database by that code and read each candidate's 510(k) summary. Then vet the shortlist for recalls, adverse event reports and the predicate each one relied on.
This post covers the rule, the FDA's 2023 draft guidance on choosing a predicate, a step-by-step search and a vetting checklist.
What counts as a predicate device
A predicate is a device your new device can be compared with for a substantial equivalence decision. Beyond the three routes in 807.92(a)(3), the FDA's page on finding predicate devices adds that a 510(k)-exempt device can serve, and that a predicate doesn't have to be on the market today, as long as it was legally marketed.
21 CFR 807.100(b) sets the test. Your device must have the same intended use as the predicate, and either the same technological characteristics, or different ones (materials, design, energy source) with data showing it is as safe and effective and raises no different questions of safety and effectiveness. Under 807.100(b)(3), the predicate must not have been removed from the market at the FDA Commissioner's initiative or found misbranded or adulterated by a court.
The FDA's 2014 guidance, The 510(k) Program: Evaluating Substantial Equivalence, adds three terms you'll use:
- Primary predicate. When you cite more than one predicate, the FDA recommends naming the one most similar to your device.
- Reference device. A legally marketed device cited only for scientific methods or standard reference values. It isn't a predicate.
- Split predicate. Taking intended use from one device and technological characteristics from another with a different intended use. The guidance calls this inconsistent with the 510(k) standard.
What the FDA's 2023 draft guidance on predicate selection says
In September 2023 the FDA issued Best Practices for Selecting a Predicate Device to Support a Premarket Notification 510(k) Submission (docket FDA-2023-D-3134). The FDA's page still labels it a draft, not for implementation, and it doesn't change the substantial equivalence standard. It proposes four practices for narrowing a list of valid predicates:
| Draft best practice | What the draft suggests checking |
|---|---|
| Cleared using well-established methods | Whether the predicate's testing used recognized consensus standards, FDA guidance or other accepted methods, and whether those are still current |
| Meets or exceeds expected safety and performance | MAUDE, MDR and MedSun reports of injuries, deaths and malfunctions after wide distribution |
| No unmitigated use-related or design-related safety issues | FDA safety communications and emerging signals on the Medical Device Safety pages |
| No associated design-related recall | The Medical Device Recalls database; the draft separates design-related recalls from manufacturing or labeling ones |
The draft also recommends a short narrative in your 510(k) summary explaining how you chose the predicate and applied these practices. If no valid predicate meets them, it suggests saying so and describing testing that addresses the predicate's known concerns.
It sets no age cutoff. The draft notes that older devices can bring long-term data, useful for implants, while newer ones may reflect current software, cybersecurity and interoperability practice.
How to find predicate device candidates, step by step
- Get your product code. Search the Product Classification database by device name or regulation number. Note the product code, device class and regulation; the FDA's own example is 21 CFR 888.1100 for arthroscopes. Our product code guide covers this step in detail.
- Search the 510(k) database by product code alone. In the 510(k) database, enter only the Product Code and leave the other fields blank. The FDA calls this usually the most effective search. Sort by Decision Date (descending) to see recent clearances first.
- Narrow by name or applicant, one field at a time. The search matches text exactly, so use one word in Device Name, and try spelling variants of a company in Applicant Name. Records keep the original applicant and trade name, not later owners.
- Read the 510(k) summaries. For each candidate, read the indications for use, the technological comparison (807.92(a)(6)), the performance data, and the predicate it cited (807.92(a)(3)). Some records carry a 510(k) statement under 807.93 instead: a certification that the 510(k) will be supplied on request within 30 days, minus patient identifiers and trade secrets.
- Shortlist by intended use first, then technology. Same intended use is the gate in 807.100(b)(1); technological differences come second.
For the database itself, see how to search the FDA 510(k) database.
How to vet a predicate before you commit
A candidate that passes the legal test can still be a poor choice. Work through each check and keep the evidence with your predicate rationale.
- Recalls. Search the Medical Device Recalls database by the candidate's number in the PMA/510(K) Number field, then by its product code. Read each record's FDA Determined Cause. Values such as "Design", "Component design/selection" or "Software design" point to design; "Error in labeling" or "Process control" don't. Class definitions are in our recall classes guide.
- Adverse event reports. Look for a pattern of malfunctions, injuries or deaths for the brand and product code. A count of reports isn't an event rate, and MAUDE is moving into the FDA's AEMS (see /guides/maude-aems/).
- Its own predicate. Open the candidate's 510(k) summary and check what it cited. If its intended use came from one device and its technology from another, that is the split pattern the 2014 guidance rejects, and a reason to look harder at that candidate.
- Its methods. Check whether the standards it was tested to are still recognized, in the FDA's Recognized Consensus Standards database.
- Its market status. Confirm it wasn't removed at the FDA's initiative or found misbranded or adulterated by a court (807.100(b)(3)).
DeviceVigil puts most of this beside the FDA's pages. Open a 510(k) record and the side panel shows the product code's class and regulation, its clearance history, its recalls with class, firm and reason, and its MAUDE reports, which you can filter by the submission number they cite. On FDA search results, hovering over a K number previews that record, so you can triage a list without opening each one. Pro adds a printable predicate dossier you can save as PDF and file with your rationale.
See it beside the FDA record
DeviceVigil is a Chrome and Edge extension for regulatory, quality and post-market teams. Open a 510(k), PMA, classification or recall page and see that device type’s clearances, recalls and MAUDE reports in one panel, and get told when something changes.
Frequently asked questions
What is a predicate device in a 510(k)?
It is the legally marketed device you claim your device is substantially equivalent to. Under 21 CFR 807.92(a)(3), it can be a device marketed before May 28, 1976, a device reclassified from Class III to Class II or I, or a device found substantially equivalent through a 510(k). Most 510(k)s cite a recently cleared device.
Can you use more than one predicate device?
Yes, in certain cases, such as combining features of devices with the same intended use or seeking more than one indication. The FDA's 2014 guidance recommends naming a primary predicate. A reference device can support test methods but isn't a predicate.
What is a split predicate?
It is a comparison that takes intended use from one device and technological characteristics from another device with a different intended use. The FDA's 2014 substantial equivalence guidance says this is inconsistent with the 510(k) standard. It says that, generally, the intended use and technology comparisons should be made against one predicate.
Can a recalled device be a predicate?
807.100(b)(3) rules out a predicate removed from the market at the FDA Commissioner's initiative or found misbranded or adulterated by a court. A recall alone isn't named there. The FDA's 2023 draft guidance recommends avoiding predicates with a design-related recall, while treating manufacturing or labeling recalls differently.
Does substantial equivalence mean the devices are identical?
No. The FDA's predicate page says substantial equivalence doesn't require identical devices. Different technological characteristics are allowed if your data show the device is as safe and effective and raises no different questions of safety and effectiveness (807.100(b)(2)).
Sources
- 21 CFR Part 807 (revised as of April 1, 2025), govinfo
- Best Practices for Selecting a Predicate Device to Support a Premarket Notification 510(k) Submission, FDA draft guidance, September 2023
- The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications 510(k), FDA guidance, July 2014
- How to Find and Effectively Use Predicate Devices, FDA
- Product Classification database, FDA
- 510(k) Premarket Notification database, FDA
- Medical Device Recalls database, FDA