Your procedures still cite 820.30 and 820.198, the next FDA inspection will follow a compliance program that no longer uses QSIT, and IT wants to know what "validated" means for the tools in your quality system. Ten months in, those are the questions the new rule keeps raising.
The QMSR (Quality Management System Regulation) is the FDA's revised 21 CFR Part 820, in effect since February 2, 2026. It incorporates ISO 13485:2016 by reference, keeps a short set of FDA-specific sections (820.1, 820.3, 820.7, 820.10, 820.35 and 820.45), and replaces the QSIT inspection technique with Compliance Program 7382.850. The FDA calls the old and new rules substantially similar, but the structure, the vocabulary and access to some records changed.
Below: what the FDA kept and added, a QSR-to-QMSR table for the clauses people ask about most, the inspection change, and software validation under the FDA's February 2026 computer software assurance guidance.
What the QMSR is and when it took effect
The QMSR is the final rule published on February 2, 2024 (89 FR 7496), which took effect two years later, on February 2, 2026. Section 820.7 incorporates ISO 13485:2016 and Clause 3 of ISO 9000:2015 (terms and definitions) by reference.
Where an ISO 13485 clause conflicts with the FD&C Act or the FDA's regulations, the Act and regulations control (820.1). The FD&C Act's definitions of "device" and "labeling" take precedence over the ISO terms. ISO 14971 is referenced inside ISO 13485 but isn't incorporated by reference into the QMSR, although ISO 13485 itself requires risk-based processes throughout.
A technical amendments rule on December 4, 2025 updated 179 sections across 18 parts of 21 CFR to point at the new section numbers. For example, 162 Class I exemptions now cite 820.35 instead of 820.180 and 820.198. The FDA said it added no new requirements.
What the FDA kept, added and removed
The FDA-specific text is short. The rest of the requirements come from ISO 13485.
- 820.10(b) ties the quality system to other FDA rules: UDI (Part 830), device tracking (Part 821), medical device reporting (Part 803) and corrections and removals (Part 806, explained in our recall classes guide).
- 820.10(c) applies ISO 13485 Clause 7.3 (design and development) to Class II and III devices, to Class I devices automated with computer software, and to five listed Class I devices.
- 820.35 adds record content for complaints, servicing and UDI, and lets you mark records as confidential.
- 820.45 requires labeling and packaging to be examined for accuracy before release, with documented release and controls against mix-ups.
What went away: the terms design history file, device master record and device history record, which ISO 13485 doesn't use. And the old 820.180(c) exception is gone, so the FDA can now inspect management review, quality audit and supplier audit reports (QMSR FAQ, Q8). Investigators may also review records created before February 2, 2026.
QSR vs QMSR: where the most-asked clauses went
This table maps the old sections to the ISO 13485:2016 clauses and QMSR sections the FDA now cites. We built it from the final rule's preamble and Attachment A of Compliance Program 7382.850, which lists the clauses investigators evaluate. It isn't an official FDA crosswalk; the FDA's QMSR page lists AAMI TIR102 among its mapping resources.
| QSR section (before Feb 2, 2026) | Where it lives under the QMSR |
|---|---|
| 820.20 Management responsibility | ISO 13485 Clause 5; management review is 5.6.1 to 5.6.3 |
| 820.22 Quality audit | Clause 8.2.4, internal audits |
| 820.25 Personnel | Clause 6.2, human resources |
| 820.30 Design controls | 820.10(c), applying Clause 7.3 (7.3.1 to 7.3.10) |
| Design history file (820.30) | Design and development file, Clause 7.3.10 |
| 820.40 Document controls | Clause 4.2.4 |
| 820.50 Purchasing controls | Clauses 7.4.1 to 7.4.3; outsourcing in 4.1.5 |
| 820.70(i) Automated processes | Clauses 4.1.6 (quality system software), 7.5.6 and 7.6 |
| 820.90 Nonconforming product | Clauses 8.3.1 to 8.3.4; concessions in 8.3.2 |
| 820.100 Corrective and preventive action | Clauses 8.5.2 and 8.5.3 |
| 820.120 Device labeling | 820.45, with Clause 7.5.1 |
| 820.181 Device master record | Medical device file, Clause 4.2.3 |
| 820.184 Device history record | Records under Clause 7.5.1 |
| 820.198 Complaint files | 820.35(a) and Clause 8.2.2; reporting through 820.10(b)(3) |
| 820.200 Servicing | 820.35(b) and Clause 7.5.4 |
How FDA inspections changed
On February 2, 2026, the FDA stopped using QSIT and began inspecting under Compliance Program 7382.850, Inspection of Medical Device Manufacturers. It also retired programs 7382.845 and 7383.001.
The program groups requirements into six QMS Areas: Change Control; Design and Development; Management Oversight; Measurement, Analysis, and Improvement; Outsourcing and Purchasing; and Production and Service Provision. Four other areas cover medical device reporting, corrections and removals, tracking and UDI. Investigators pick elements based on product risk. Under Inspection Model 1 they cover at least one element per area; Model 2, used for baseline surveillance and PMA preapproval inspections, sets specific minimum elements.
An ISO 13485 certificate doesn't exempt a site from FDA inspection, and the FDA won't require or issue certificates (FAQ Q13). For sites enrolled in MDSAP, the FDA uses MDSAP audit reports in place of routine surveillance inspections. Observations still go on Form FDA 483.
Software validation under ISO 13485 4.1.6 and computer software assurance
ISO 13485 Clauses 4.1.6, 7.5.6 and 7.6 require you to validate software used in production or the quality system for its intended use, in proportion to the risk. The FDA's final guidance, Computer Software Assurance for Production and Quality Management System Software, issued February 3, 2026, describes a risk-based way to do that. It replaces the September 24, 2025 version and Section 6 of the General Principles of Software Validation guidance.
The guidance's framework has four steps:
- Identify the intended use. Decide whether the software is used directly in production or the quality system, or supports them.
- Decide the risk-based approach. List reasonably foreseeable failures and decide whether each is high process risk, meaning a quality problem that could foreseeably compromise safety.
- Choose assurance activities. Scripted testing for high-risk functions, alone or with unscripted testing; unscripted testing (scenario, error guessing, exploratory) for lower-risk ones. The developer's practices, validation work and records can be a starting point, and for lower-risk software they may be enough.
- Record it. Intended use, risk analysis, testing, issues found, the conclusion, and who did the work and when. The guidance prefers digital records, such as system logs and audit trails, to screenshots.
Its examples are worth reading if you use data tools in post-market work. It lists automated surveillance or trending of data essential to device safety as high process risk. In its business intelligence example, connections to external data sources are high process risk because a failure could hide a quality trend, while reporting functions that don't inform quality decisions are not.
If your team uses DeviceVigil in post-market searches or predicate research, our page for IT and quality teams sets out what it does, what it doesn't do, how we test it and how to check any figure against openFDA and the FDA's own pages.
See it beside the FDA record
DeviceVigil is a Chrome and Edge extension for regulatory, quality and post-market teams. Open a 510(k), PMA, classification or recall page and see that device type’s clearances, recalls and MAUDE reports in one panel, and get told when something changes.
Frequently asked questions
What is the QMSR?
The Quality Management System Regulation is the FDA's current 21 CFR Part 820 for medical device manufacturers, in effect since February 2, 2026. It incorporates ISO 13485:2016 by reference and adds FDA-specific requirements in 820.10, 820.35 and 820.45. It replaced the Quality System Regulation (QSR).
What is the difference between the QMSR and the QSR?
The QSR wrote out its own requirements in sections such as 820.30 (design controls) and 820.198 (complaint files). The QMSR takes them from ISO 13485:2016 and keeps a few FDA additions. It drops terms like design history file and device master record, and management review, audit and supplier audit reports are no longer exempt from FDA inspection.
Does the FDA require ISO 13485 certification under the QMSR?
No. The FDA's QMSR FAQ says it won't require or issue ISO 13485 certificates, and a certificate doesn't exempt a manufacturer from FDA inspection. The FDA inspects against Part 820, including the ISO 13485 clauses it incorporates.
Is ISO 14971 part of the QMSR?
ISO 14971 isn't incorporated by reference. ISO 13485 refers to it and requires a risk-based approach, and the final rule's preamble lists where risk management sits in ISO 13485, including Clauses 4.1, 7.1 and 7.3.
What replaced QSIT?
Compliance Program 7382.850, Inspection of Medical Device Manufacturers, effective February 2, 2026. It organizes inspections into six QMS Areas plus four areas for other FDA requirements, with elements chosen by product risk.
Sources
- Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, February 2, 2024
- Medical Devices; Quality Management System Regulation Technical Amendments, December 4, 2025
- Quality Management System Regulation (QMSR), FDA
- Quality Management System Regulation: Frequently Asked Questions, FDA
- Compliance Program 7382.850, Inspection of Medical Device Manufacturers, FDA
- Computer Software Assurance for Production and Quality Management System Software, FDA guidance, February 2026
- 21 CFR Part 820, Quality System Regulation (revised as of April 1, 2025), govinfo