FDA product code
PIF: Gastrointestinal Tubes With Enteral Specific Connectors (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.5980
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 523 in the last 12 months
- Median review time
- 200 daysthe latest 25 decisions
- Recall records
- 1121 in the last 12 months
- Adverse event reports
- 175last 12 months, 12 deaths
510(k) and De Novo decisions
Decisions for product code PIF per year, by decision date, 52 since 2014. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 5 |
| 2018 | 2 |
| 2019 | 7 |
| 2020 | 4 |
| 2021 | 4 |
| 2022 | 4 |
| 2023 | 4 |
| 2024 | 2 |
| 2025 | 2 |
| 2026 (so far) | 3 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260110 | danumed Extension Set for Gastrostomy ButtonDanumed Medizintechnik GmbH | Sep 10, 2026239 days |
| K253715 | Edi Catheter ENFit 6Fr/48 cm (6893166); Edi Catheter ENFit 6Fr/49cm (6893273); Edi Catheter ENFit 6Fr/50cm (6893274); Edi Catheter ENFit 8Fr/50cm (6893275); Edi Catheter ENFit 8Fr/100cm (6893276); Edi Catheter ENFit 8Fr/125cm (6893277); Edi Catheter ENFit 12Fr/125cm (6893278); Edi Catheter ENFit 16Fr/125cm (6893279)Maquet Critical Care AB | Aug 26, 2026275 days |
| K254170 | Percutaneous Endoscopic Gastrostomy (PEG) KitDegania Silicone , Ltd. | Mar 20, 202687 days |
| K242917 | Enteral Drainage System, Enteral Medicine strawHMC Premedical S.p.A. | Jun 18, 2025267 days |
| K250481 | Extension Feeding Set with ENFit™ ConnectorsSKY Medical, a.s. | Apr 18, 202558 days |
| K240052 | Disposable Enteral Feeding SetsBeijing L&Z Medical Technology Development Co., Ltd. | Jul 3, 2024177 days |
| K233034 | Mobility+ Enteral Feeding System OTCRockfield Medical | Apr 12, 2024200 days |
| K230326 | ENFit to ENFit Extension SetsVesco Medical | Nov 2, 2023269 days |
| K232046 | Kangaroo Skin Level Balloon Gastrostomy Kit, Kangaroo Bolus Feed Set with ENFIT Connection, Kangaroo Continuous Feed Set with ENFIT ConnectionXeridiem Medical Devices, A Spectrum Plastics Group Company | Oct 6, 202388 days |
| K222773 | Feeding TubeAnhui Tiankang Medical Technology Co., Ltd. | Aug 17, 2023337 days |
Recalls
Recall records for product code PIF per year, by the date the recall began. Each recalled product is one record; these 112 records belong to 7 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 3 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1551-2026Class II | Boston Scientific CorporationRecall of kits which utilized ChloraPrep Triple Swabsticks from BD. Open or compromised seals on the sterile swabstick packaging, which could result in loss of… | Feb 10, 2026 |
| Z-0819-2025Class II | Applied Medical Technology IncThe extension sets subject to this recall have ENFit connections, although they should have non- ENFit connections. ENFit connectors will only fit with other E… | Nov 26, 2024 |
| Z-1513-2024Class II | Boston Scientific CorporationInability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure… | Feb 20, 2024 |
| Z-1514-2024Class II | Boston Scientific CorporationInability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure… | Feb 20, 2024 |
| Z-1153-2024Class II | Boston Scientific CorporationAffected batches of the kit contain Stoma Measuring Devices sized to measure for an 18/24F replacement button, but are labeled as being for a 28F. Use of defec… | Dec 19, 2023 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PIF. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 17 |
| 2018 | 17 |
| 2019 | 28 |
| 2020 | 70 |
| 2021 | 89 |
| 2022 | 77 |
| 2023 | 96 |
| 2024 | 110 |
| 2025 | 84 |
| 2026 (so far) | 166 |
Last 12 months by event type
- Deaths12
- Injuries46
- Malfunctions112
- Other or not stated5
Most reported device problems, last 12 months
- Break53
- Crack24
- Fluid/Blood Leak21
- Detachment of Device or Device Component13
- Insufficient Device Problem Information12
- Device Dislodged or Dislocated10
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 1,031in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 146registered establishments that list this product code
Other codes under 21 CFR 876.5980
Product codes classified under the same regulation.
Watch PIF in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PIF every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PIF. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PIFRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PIFRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PIF · count decision_dateRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PIFRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PIF · count event_date_initiatedRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PIF · count res_event_numberRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1551-2026" "Z-0819-2025" "Z-1513-2024" "Z-1514-2024" "Z-1153-2024")Retrieved Nov 29, 2026, 4:14 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PIF AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PIF AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PIF AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:14 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PIFRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PIFRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:876.5980Retrieved Nov 28, 2026, 9:33 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil