FDA product code
OKI: Emergency Response Safety Kit (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 878.4040
- Review panel
- General Hospital
- Premarket route
- Not stated
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 00 in the last 12 months
- Median review time
- Not availablethe latest 0 decisions
- Recall records
- 65 in the last 12 months
- Adverse event reports
- 10last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code OKI per year, by decision date. Review time is the days from receipt to decision.
No decisions on record for this product code.
Recalls
Recall records for product code OKI per year, by the date the recall began. Each recalled product is one record; these 6 records belong to 3 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 5 |
| 2026 (so far) | 1 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0222-2027Class I | Cardinal Health 200, LLCKits contain infusion sets which were recalled for performance issues. | Sep 24, 2026 |
| Z-1114-2026Class I | SunMed Holdings, LLCProduct contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine. | Dec 15, 2025 |
| Z-1115-2026Class I | SunMed Holdings, LLCProduct contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine. | Dec 15, 2025 |
| Z-1116-2026Class I | SunMed Holdings, LLCProduct contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine. | Dec 15, 2025 |
| Z-1117-2026Class I | SunMed Holdings, LLCProduct contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine. | Dec 15, 2025 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code OKI. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 3 |
| 2024 | 0 |
| 2025 | 17 |
| 2026 (so far) | 3 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions8
- Other or not stated1
Most reported device problems, last 12 months
- Application Program Problem: Dose Calculation Error4
- Application Program Problem: Medication Error1
- Device Displays Incorrect Message1
- Device Markings/Labelling Problem1
- Material Integrity Problem1
In plain English: what the newest reports describe Written by AI
The flumazenil dose in the reference section is printed as 0.1 mg/kg instead of 0.01 mg/kg, which would be a tenfold overdose if used. Several reports say the colored sections list the correct dose.
18 reports: MW5182946, MW5182596, 23837390, MW5180289, MW5180294, MW5180164, MW5179692, MW5179701, MW5179413, MW5179345, MW5179090, MW5178966, MW5178981, MW5178930, MW5178933, MW5178745, MW5178608, MW5179088
The vecuronium dose in the calculation or reference section of the dosing tape is printed as 0.1 mg/mL, a drug concentration, instead of 0.1 mg/kg, a weight-based dose. Several reports say the correct dose appears in the colored sections.
16 reports: MW5182946, MW5182596, MW5180289, MW5180294, MW5180164, MW5179692, MW5179701, MW5179413, MW5179345, MW5179090, MW5178966, MW5178981, MW5178930, MW5178933, MW5178608, MW5179088
Several reports say the manufacturer has not responded, has no plans to correct the errors, or has not told users about them. Some users are blacking out the wrong doses, covering them, or adding corrected stickers while waiting for new cards.
10 reports: MW5180289, MW5180294, MW5178930, MW5178933, MW5179090, MW5178966, MW5178608, MW5179088, 23837390, MW5182946
Two reports say the ketamine dose for pain relief is listed as 1 mg/kg, which is the sedation dose, instead of 0.1 mg/kg. One report warns this tenfold difference could lead to a sedation dose being given when only pain relief was intended.
2 reports: MW5185100, MW5182596
A report describes an error in the mannitol section for raised brain pressure: 26.5 grams at 20 percent strength should be 132.5 mL, but the card lists 131.3 mL. The reporter says both volumes fall within the dosing range for these patient weights.
Report: MW5180704
Written by AI from the narratives of the 20 newest reports that have one, received Nov 7, 2025 to Mar 12, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 107in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 13registered establishments that list this product code
Other codes under 21 CFR 878.4040
Product codes classified under the same regulation.
Watch OKI in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get OKI every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OKI. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:OKIRetrieved Nov 29, 2026, 12:25 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OKIRetrieved Nov 29, 2026, 6:45 PM UTC.
- device/510k · search product_code:OKI · count decision_dateRetrieved Nov 29, 2026, 12:25 AM UTC.
- device/recall · search product_code:OKIRetrieved Nov 29, 2026, 6:45 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OKI · count event_date_initiatedRetrieved Nov 29, 2026, 12:25 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OKI · count res_event_numberRetrieved Nov 29, 2026, 12:25 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0222-2027" "Z-1114-2026" "Z-1115-2026" "Z-1116-2026" "Z-1117-2026")Retrieved Nov 29, 2026, 6:45 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OKI AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:45 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OKI AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:45 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OKI AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 12:25 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:OKIRetrieved Nov 29, 2026, 12:25 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:OKIRetrieved Nov 29, 2026, 12:25 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:878.4040Retrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.