FDA product code
OJU: Skin Prep Tray (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 878.4800
- Review panel
- General and Plastic Surgery
- Premarket route
- Not stated
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 10 in the last 12 months
- Median review time
- 16 daysthe latest 1 decisions
- Recall records
- 213 in the last 12 months
- Adverse event reports
- 44last 12 months, 1 death
510(k) and De Novo decisions
Decisions for product code OJU per year, by decision date, 1 since 1978. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K781797 | PREP TRAY, PREOPERATIVE SKINPharmaseal Div., Baxter Healthcare Corp. | Nov 8, 197816 days |
Recalls
Recall records for product code OJU per year, by the date the recall began. Each recalled product is one record; these 21 records belong to 17 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 3 |
| 2020 | 3 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 1 |
| 2024 | 3 |
| 2025 | 2 |
| 2026 (so far) | 3 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-3074-2026Class II | Medline Industries, LPDownstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on th… | Jul 20, 2026 |
| Z-3058-2026Class II | Medline Industries, LPThe kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in… | Jun 12, 2026 |
| Z-2044-2026Class II | Medline Industries, LPMedline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal… | Mar 16, 2026 |
| Z-1057-2026Class II | Medline Industries, LPMedline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check val… | Nov 26, 2025 |
| Z-1606-2025Class II | MEDLINE INDUSTRIES, LP - NorthfieldThere is a potential for open/weak seals on purchased breather pouches used for packaging of select sterile convenience kits. | Mar 3, 2025 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code OJU. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 6 |
| 2018 | 2 |
| 2019 | 1 |
| 2020 | 9 |
| 2021 | 88 |
| 2022 | 41 |
| 2023 | 4 |
| 2024 | 4 |
| 2025 | 2 |
| 2026 (so far) | 44 |
Last 12 months by event type
- Deaths1
- Injuries0
- Malfunctions43
- Other or not stated0
Most reported device problems, last 12 months
- Unsealed Device Packaging13
- Tear, Rip or Hole in Device Packaging12
- Fluid/Blood Leak10
- Adhesive Too Strong4
- Break2
- Adverse Event Without Identified Device or Use Problem1
In plain English: what the newest reports describe Written by AI
Several reports describe kit packaging that was not sealed or was already open, so the contents may not be sterile. In one, a seam split when the tray was picked up. In another, the bag's perforation made it easy to open, and staff found bags open on the supply cart.
5 reports: 1417592-2026-47425, 1417592-2026-44631, 1417592-2026-15439, 1417592-2026-54078, 1417592-2026-48190
Three reports describe leaking or burst packaging. One pack had liquid leak out, two packs had fluid bags burst, and a table cover had holes. One report states the breather pouch had no holes.
3 reports: 1417592-2026-55363, 1417592-2026-36408, 1417592-2026-56576
Two reports describe blue lint from the towels in the kit. In one, lint was seen on the patient's prepped skin in the operating room, and the patient was re-prepped to remove it. The other says the towels lint when surgeons dry off with them and when they are placed on prepped skin.
2 reports: 1417592-2026-53690, 1417592-2026-15094
Two reports describe skin tears where an IV was placed or anchored. One says the tape caused the tear, but no further details are available. The other says no medical treatment was needed and the patient is stable.
2 reports: 3004122598-2026-00026, 3004122598-2026-00030
One report says the antiseptic applicators in the sterile central line dressing kits had dried out.
Report: 23605200
Written by AI from the narratives of the 18 newest reports that have one, received Nov 20, 2025 to Oct 19, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 1,336in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 41registered establishments that list this product code
Other codes under 21 CFR 878.4800
Product codes classified under the same regulation.
Watch OJU in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get OJU every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OJU. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:OJURetrieved Nov 29, 2026, 9:48 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OJURetrieved Nov 29, 2026, 9:48 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OJU · count decision_dateRetrieved Nov 29, 2026, 9:48 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:OJURetrieved Nov 29, 2026, 9:48 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OJU · count event_date_initiatedRetrieved Nov 29, 2026, 9:48 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OJU · count res_event_numberRetrieved Nov 29, 2026, 9:48 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-3074-2026" "Z-3058-2026" "Z-2044-2026" "Z-1057-2026" "Z-1606-2025")Retrieved Nov 29, 2026, 9:48 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OJU AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 9:48 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OJU AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 9:48 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OJU AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 9:48 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:OJURetrieved Nov 29, 2026, 9:48 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:OJURetrieved Nov 29, 2026, 9:48 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:878.4800Retrieved Nov 29, 2026, 6:19 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil