Product codes

FDA product code

ODD: Endoscopic Retrograde Cholangiopancreatography (Ercp) Cannula (FDA product code)

Follow ODD: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 876.1500
Review panel
Gastroenterology and Urology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
180 in the last 12 months
Median review time
58 daysthe latest 18 decisions
Recall records
00 in the last 12 months
Adverse event reports
6last 12 months, 1 death

510(k) and De Novo decisions

Decisions for product code ODD per year, by decision date, 18 since 1980. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20181
20190
20201
20210
20220
20230
20240
20251
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K250573 Single Use Cannula V PR-V223Q/V414Q/V416Q/V418Q/V420Q/V427Q/V434Q/V435Q; Single Use 2-Lumen Cannula V PR-V614MOlympus Medical Systems Corp. May 27, 202590 days
K202583 BioShield biopsy valve EUS - LinearSTERIS Corporation Nov 6, 202059 days
K171989 Classic ERCP Catheter and Huibregtse-Katon ERCP Catheter and Huibregtse-Katon Angled, Glo-Tip ERCP Catheter, Glo-Tip II Double Lumen ERCP Catheter & w/ Radiopaque Bands, Fusion ERCP Catheter with DomeTip and Fusion OMNI ERCP Catheter with DomeTip, Haber RAMP Catheter, Soehendra Universal CatheterWilson-Cook Medical, Inc. Mar 20, 2018260 days
K070420 BIOSHIELD-ERCP BIOPSY VALVE, MODEL 00711138United States Endoscopy Group, Inc. Mar 19, 200727 days
K011149 DISPOSABLE BENDING CANNULA PR-233QThe Olympus Optical Co. May 15, 200129 days
K010993 ALIEN RX MICRO CANNULA, MODEL 4530Boston Scientific Corp Apr 30, 200127 days
K973205 WILSON-COOK STRICTURE MEASURING CATHETERWilson-Cook Medical, Inc. Nov 24, 199790 days
K963051 ERCP CANNULAUnited States Endoscopy Group, Inc. Mar 28, 1997234 days
K953967 ENDOSCOPIC CATHETERAct Medical, Inc. Jan 22, 1996153 days
K950700 BARD BILISYSTEM ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) CANNULAC.R. Bard, Inc. Apr 27, 199571 days

Recalls

Recall records for product code ODD per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code ODD. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20172
201813
20195
202011
20218
202211
202315
202412
20255
2026 (so far)5

Last 12 months by event type

  • Deaths1
  • Injuries2
  • Malfunctions3
  • Other or not stated0

Most reported device problems, last 12 months

  1. Adverse Event Without Identified Device or Use Problem4
  2. Break1
  3. Detachment of Device or Device Component1

In plain English: what the newest reports describe Written by AI

  • In three reports, the small marker at the tip, which shows up on X-ray, was hard or impossible to see. In one report the marker ring had come off and lodged in the patient's papilla (the opening where the bile and pancreatic ducts enter the bowel) and was removed; the physician said this may raise the risk of pancreatitis.

    3 reports: 3005099803-2026-01690, 9614641-2025-00901, 1037905-2024-00678

  • In two reports, the sterile packaging was faulty. One report found a hair between the outer packaging and the inner pouch before opening, and the other found the package was not sealed. Neither report describes any patient complications.

    2 reports: 3005099803-2024-01792, 3005099803-2023-07049

  • In one report, a balloon catheter may have been damaged while the physician pulled on it to drag out a stone in the pancreatic duct. Contrast dye then flowed into the duct, and the patient developed pancreatitis (inflammation of the pancreas) and later had part of the pancreas removed.

    Report: 3005099803-2026-04030

  • In one report, the guidewire, a thin wire used to guide devices into place, could not pass through the device, and a second guidewire also failed. The procedure was finished with another device. The report first said the patient was injured, then said later information showed no harm to the patient.

    Report: 3005099803-2024-02814

  • In one report, a patient said they got an infection from the device during a routine check-up. The narrative does not describe the infection or confirm that the device caused it, and it states that the infection was unrelated to medications.

    Report: MW5173136

Written by AI from the narratives of the 16 newest reports that have one, received Jan 5, 2024 to Oct 6, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
104in the FDA’s Global Unique Device Identification Database
Establishment listings
27registered establishments that list this product code

Watch ODD in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get ODD every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in ODD. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.