Product codes

FDA product code

FHP: Needle, Pneumoperitoneum, Simple (FDA product code)

Follow FHP: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 876.1500
Review panel
Gastroenterology and Urology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
20 in the last 12 months
Median review time
41 daysthe latest 2 decisions
Recall records
00 in the last 12 months
Adverse event reports
1last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code FHP per year, by decision date, 2 since 1986. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K910855 ENDOPATH VERESS NEEDLESymbiosis Corp. Apr 18, 199149 days
K860102 MOD. ENDONEEDLE SINGLE-USE PNEUMOPERITONEUM NEEDLEEndotherapeutics Feb 12, 198633 days

Recalls

Recall records for product code FHP per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code FHP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20180
20190
20201
20210
20220
20230
20240
20250
2026 (so far)1

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions1
  • Other or not stated0

Most reported device problems, last 12 months

  1. Mechanical Problem1

In plain English: what the newest reports describe Written by AI

  • Two reports describe the needle penetrating the bowel during insertion. In one a small bowel was punctured, requiring additional surgery.

    5 reports: 537412, MW1005777, MW1005778, MW1005779, MW1005780

  • Several reports describe the Veress needle (a spring-loaded needle used to create space in the abdomen before keyhole surgery) being inserted into or near blood vessels, most often in gallbladder removal.

    3 reports: 1527736-1998-03105, 193090, 174711

  • Several reports describe the needle breaking, with the shaft or sheath coming off inside the patient. In one case a broken shaft was pushed further in before open surgery; in another a broken shaft was removed from the patient after the patient returned with abdominal pain.

    2 reports: 2020550-1998-00015, 53186

  • In a report of a diagnostic laparoscopy, the needle was placed three times while pressure readings showed it was not inside the abdominal cavity; the sheath did not fully retract over the blunt edge, and a bleeding area behind the abdominal lining and a defect in the lining were later found. The patient was monitored in hospital and discharged in stable condition.

    Report: 797318

  • In one report of a tubal ligation, the needle was too short for an obese patient, so insufflation (filling the abdomen with gas) took about ten minutes and the needle eventually passed through the abdominal lining. No consequence to the patient was reported.

    Report: 1218017-1997-00013

Written by AI from the narratives of the 18 newest reports that have one, received Jan 31, 1995 to Sep 24, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
2in the FDA’s Global Unique Device Identification Database
Establishment listings
1registered establishments that list this product code

Watch FHP in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get FHP every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in FHP. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.