FDA product code
OCW: Endoscopic Tissue Approximation Device (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.1500
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 712 in the last 12 months
- Median review time
- 119 daysthe latest 25 decisions
- Recall records
- 43 in the last 12 months
- Adverse event reports
- 532last 12 months, 1 death
510(k) and De Novo decisions
Decisions for product code OCW per year, by decision date, 71 since 1994. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 2 |
| 2019 | 1 |
| 2020 | 1 |
| 2021 | 3 |
| 2022 | 0 |
| 2023 | 3 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 2 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262044 | Mediclose Cartridge SystemHealth Value Creation B.V. (trading as Corporis Medical) | Oct 14, 2026119 days |
| K252400 | Endomina EZFuse systemEndo Tools Therapeutics S.A. | Feb 13, 2026197 days |
| K243750 | SimpleStitch Suturing SystemEnvision Endoscopy | Apr 17, 2025133 days |
| K231240 | MaxTack™ Motorized Fixation DeviceCovidien | Sep 8, 2023133 days |
| K213271 | MedicloseTM SystemHealth Value Creation B.V., Trading AS Corporis Medical | Aug 31, 2023700 days |
| K231553 | OverStitch NXT Endoscopic Suturing SystemApollo Endosurgery, Inc. | Jun 29, 202330 days |
| K211309 | endomina systemEndo Tools Therapeutics S.A. | Dec 7, 2021221 days |
| K201744 | Su2ura Approximation DeviceAnchora Medical, Ltd. | Jun 17, 2021357 days |
| K210266 | OverStitch Sx Endoscopic Suturing SystemApollo Endosurgery, Inc. | Mar 2, 202129 days |
| K193652 | AbClose - Port Site Closure DeviceMedeon Biodesign, Inc. | Mar 26, 202087 days |
Recalls
Recall records for product code OCW per year, by the date the recall began. Each recalled product is one record; these 4 records belong to 3 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 3 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2961-2026Class II | Boston Scientific CorporationDevice may fail to deploy the cinch implant and/or fail to cut the suture as intended. | Jul 14, 2026 |
| Z-2962-2026Class II | Boston Scientific CorporationDevice may fail to deploy the cinch implant and/or fail to cut the suture as intended. | Jul 14, 2026 |
| Z-2281-2026Class II | Covidien, LPThe presence of holes in the pouch may compromise the package s ability to maintain the sterile barrier and protect the product from environmental exposure. | Apr 15, 2026 |
| Z-0986-2024Class II | Covidien LPPackaging of the device may cause a low seal strength in the side seal of the foil pouch potentially resulting in a sterile/moisture barrier breach, compromisi… | Nov 16, 2023 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code OCW. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 420 |
| 2018 | 413 |
| 2019 | 291 |
| 2020 | 183 |
| 2021 | 198 |
| 2022 | 235 |
| 2023 | 227 |
| 2024 | 211 |
| 2025 | 329 |
| 2026 (so far) | 499 |
Last 12 months by event type
- Deaths1
- Injuries180
- Malfunctions350
- Other or not stated1
Most reported device problems, last 12 months
- Break179
- Activation Failure121
- Detachment of Device or Device Component86
- Adverse Event Without Identified Device or Use Problem84
- Failure to Cut60
- Failure to Fire45
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 108in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 106registered establishments that list this product code
Other codes under 21 CFR 876.1500
Product codes classified under the same regulation.
Watch OCW in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get OCW every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OCW. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:OCWRetrieved Nov 29, 2026, 12:08 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OCWRetrieved Nov 29, 2026, 12:08 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OCW · count decision_dateRetrieved Nov 29, 2026, 12:08 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:OCWRetrieved Nov 29, 2026, 12:08 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OCW · count event_date_initiatedRetrieved Nov 29, 2026, 12:08 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OCW · count res_event_numberRetrieved Nov 29, 2026, 12:08 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2961-2026" "Z-2962-2026" "Z-2281-2026" "Z-0986-2024")Retrieved Nov 29, 2026, 12:08 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OCW AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 12:08 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OCW AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 12:08 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OCW AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 12:08 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:OCWRetrieved Nov 29, 2026, 12:08 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:OCWRetrieved Nov 29, 2026, 12:08 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:876.1500Retrieved Nov 29, 2026, 8:27 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil