FDA product code
GCE: Adaptor, Catheter (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 878.4200
- Review panel
- General and Plastic Surgery
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 40 in the last 12 months
- Median review time
- 28 daysthe latest 4 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 5last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code GCE per year, by decision date, 4 since 1980. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K901716 | KLEIN CONNECTOR(TM)Jeff Klein Surgical, Inc. | May 11, 199028 days |
| K901715 | KLEIN CORK(TM)Jeff Klein Surgical, Inc. | May 11, 199028 days |
| K855039 | MEDIFLEX INJECTION CAPMediflex Intl. | Jan 24, 198638 days |
| K801680 | BETA-CAP II ADAPTERQuinton, Inc. | Aug 4, 198013 days |
Recalls
Recall records for product code GCE per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code GCE. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 1 |
| 2018 | 5 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 3 |
| 2025 | 3 |
| 2026 (so far) | 5 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions5
- Other or not stated0
Most reported device problems, last 12 months
- Deformation Due to Compressive Stress1
- Difficult to Remove1
- Disconnection1
In plain English: what the newest reports describe Written by AI
Several reports describe a gel medication that would not come out of the catheter after the syringe was attached and used as instructed. In one investigation, a blue connection port on the device was found cracked, which kept the product from dispensing.
3 reports: 3004837686-2025-00003, 3004837686-2024-00016, 21034612
Several reports describe ports on a drain connector coming open or loose, and a cap that would not stay on or seal tightly, which can let fluid or air through. One report states that air entered the chest cavity through unplugged ports, with no injury reported but a risk of breathing problems noted.
3 reports: 1018233-2024-04413, 1018233-2019-07741, 1018233-2018-05449
Several reports describe a kink in the catheter before it was opened, so the staff could not push the medication through properly and it leaked. The manufacturer's findings are not stated.
2 reports: 3004837686-2026-00002, 24505660
One report describes a crack where the hub meets the tubing of a central line's white lumen, which required an additional procedure. A separate report describes blood leaking from a connection on a catheter, with about 5 mL of blood loss and no adverse effects reported.
2 reports: 6099095, 9615050-2008-00372
One report describes a catheter tip breaking off inside a patient's leg during withdrawal. The doctor could not retrieve it, and the patient needed surgery to bypass the artery.
Report: MW5067200
Written by AI from the narratives of the 18 newest reports that have one, received Dec 22, 2008 to Apr 23, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 27in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 40registered establishments that list this product code
Other codes under 21 CFR 878.4200
Product codes classified under the same regulation.
Watch GCE in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get GCE every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in GCE. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:GCERetrieved Nov 28, 2026, 8:53 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:GCERetrieved Nov 29, 2026, 6:44 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:GCE · count decision_dateRetrieved Nov 28, 2026, 8:53 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:GCERetrieved Nov 29, 2026, 6:44 PM UTC.
- device/recall · search product_code:GCE · count event_date_initiatedRetrieved Nov 28, 2026, 8:53 PM UTC.
- device/recall · search product_code:GCE · count res_event_numberRetrieved Nov 28, 2026, 8:53 PM UTC.
- device/event · search device.device_report_product_code:GCE AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:44 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:GCE AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:44 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:GCE AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 6:44 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:GCERetrieved Nov 28, 2026, 8:53 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:GCERetrieved Nov 28, 2026, 8:53 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:878.4200Retrieved Nov 28, 2026, 7:43 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil