Product codes

FDA product code

GBX: Catheter, Irrigation (FDA product code)

Follow GBX: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class I
Regulation
21 CFR 878.4200
Review panel
General and Plastic Surgery
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
770 in the last 12 months
Median review time
78 daysthe latest 25 decisions
Recall records
821 in the last 12 months
Adverse event reports
1,054last 12 months, 2 deaths

510(k) and De Novo decisions

Decisions for product code GBX per year, by decision date, 77 since 1977. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K131659 MOERAE VEIN PREPARATION KITMoerae Matrix, Inc. Mar 3, 2014270 days
K082725 GRAFTGUARD PRESSURE CONTROLLING SYRINGEMaquet Cardiovascular, LLC Dec 17, 200891 days
K052286 DEUTSCH ANTI-BLOCKAGE WOUND DRAINThe Catheter Exchange, Inc. Oct 21, 200560 days
K043515 GRAFTSAVERVescare, Inc. Mar 31, 2005101 days
K013417 MISONIX IRRIGATION SYSTEM MOEDL BC20PMisonix, Inc. Jan 10, 200287 days
K000704 DMC SAPHENOUS VEIN DISTENTION SYSTEM (SVDS)Dmc Medical, Ltd. May 19, 200078 days
K974233 LYSONIX IRRIGATION SYSTEMLysonix, Inc. Feb 3, 199883 days
K973703 JACKSON-PRATT HEMADUCT WOUND DRAIN, JACKSON-PRATT GOLD HEMADUCT WOUND DRAINAllegiance Healthcare Corp. Dec 4, 199769 days
K972635 ULTRACELL SUCTION SPONGEUltracell Medical Technologies, Inc. Oct 1, 199779 days
K965146 FLAT CLOSED WOUND DRAIN WITH TROCARAxiom Medical, Inc. Mar 10, 199777 days

Recalls

Recall records for product code GBX per year, by the date the recall began. Each recalled product is one record; these 82 records belong to 19 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
201711
201811
20198
20202
20212
202229
20230
20245
20256
2026 (so far)1

Latest recalls

Recall and classFirm and reasonBegan
Z-1790-2026Class II Cook IncorporatedProducts from the affected device lots were labeled with expiration dates that exceed the true shelf life. Mar 5, 2026
Z-0950-2026Class II Rocket Medical PlcIFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use. Nov 7, 2025
Z-0951-2026Class II Rocket Medical PlcIFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use. Nov 7, 2025
Z-0952-2026Class II Rocket Medical PlcIFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use. Nov 7, 2025
Z-0953-2026Class II Rocket Medical PlcIFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use. Nov 7, 2025

Adverse event reports (MAUDE)

Medical device reports received per year that name product code GBX. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017176
2018277
2019196
2020198
2021383
2022477
2023628
2024945
20251,193
2026 (so far)944

Last 12 months by event type

  • Deaths2
  • Injuries85
  • Malfunctions964
  • Other or not stated3

Most reported device problems, last 12 months

  1. Fracture276
  2. Material Separation254
  3. Nonstandard Device213
  4. Break180
  5. Fluid/Blood Leak141
  6. Material Integrity Problem63

In plain English: what the newest reports describe Written by AI

  • Several reports describe the plastic connector at the end of the drainage catheter cracking, fracturing or breaking, in some cases with fluid leaking out. In these cases the catheter was often replaced or another device was used to finish the procedure.

    6 reports: 3006260740-2026-07996, 3006260740-2026-07977, 3006260740-2026-07585, 3006260740-2026-07683, 3006260740-2026-07570, 3006260740-2026-07573

  • Several reports describe the drain's trocar, the sharp tool used to puncture the skin, as shorter than the catheter or not sharp enough to pierce skin cleanly. In these cases the procedure was completed with another device.

    3 reports: 3006260740-2026-07574, 3006260740-2026-07483, 1018233-2026-06757

  • Two reports describe the drain tubing coming loose or disconnecting from its connection, including one where the tubing was forced into the ports and the insertion points became loose and did not work correctly.

    2 reports: 1018233-2026-06757, 2210968-2026-10584

  • One report describes a drain that could not be removed after surgery and broke inside the patient, leaving a fragment in the body. The patient needed a further operation to remove the piece, and the surgeons suggested a guide-wire may have pierced the drain, though this was not confirmed.

    Report: 2210968-2026-10739

  • One report describes a catheter that arrived with its outer package damaged and its inner package found unsealed. The procedure was completed using another device.

    Report: 3006260740-2026-07717

Written by AI from the narratives of the 13 newest reports that have one, received Oct 14, 2026 to Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
1,327in the FDA’s Global Unique Device Identification Database
Establishment listings
275registered establishments that list this product code

Watch GBX in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get GBX every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in GBX. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.