FDA product code
FET: Endoscopic Video Imaging System/Component, Gastroenterology-Urology (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.1500
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 1013 in the last 12 months
- Median review time
- 110 daysthe latest 25 decisions
- Recall records
- 80 in the last 12 months
510(k) and De Novo decisions
Decisions for product code FET per year, by decision date, 101 since 1981. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 3 |
| 2018 | 3 |
| 2019 | 3 |
| 2020 | 5 |
| 2021 | 1 |
| 2022 | 2 |
| 2023 | 1 |
| 2024 | 2 |
| 2025 | 6 |
| 2026 (so far) | 3 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262079 | 4K Camera Head OLYMPUS CH-S700-08-LBOlympus Medical Systems Corp. | Sep 10, 202680 days |
| K261880 | FUJIFILM Processor EP-8000; Balloon Controller PB-30; Endoscopy Support Program EW10-VM01Fujifilm Corporation | Jul 2, 202627 days |
| K261260 | HD Camera Head OLYMPUS CH-S200-08-LBOlympus Medical Systems Corporation | Jun 24, 202669 days |
| K251734 | ProVee Video Processing Unit (PV-003)Proverum Limited | Sep 24, 2025110 days |
| K250462 | UL UHD-Clear View 4K Camera System (UL-3Chip 4K)Univlabs Technologies Private Limited | Aug 11, 2025174 days |
| K250204 | Endoscopic Camera SystemShenzhen Sophway Technology Co., Ltd. | Jul 1, 2025158 days |
| K243260 | FUJIFILM Processor EP-8000; FUJIFILM Endoscope Model EG-860RFujifilm Corporation | Apr 11, 2025178 days |
| K242067 | VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700; VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700; 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EAOlympus Corporations of the Americas | Apr 2, 2025261 days |
| K243321 | Endoscopic Video Image Processor (RP-IPD-V1000F)Guangzhou Red Pine Medical Instrument Co., Ltd. | Feb 7, 2025107 days |
| K241532 | Flexible Video-Choledo-Cysto-Ureteroscope System (PL-2100)Shanghai SeeGen Photoelectric Technology Co., Ltd. | Oct 22, 2024145 days |
Recalls
Recall records for product code FET per year, by the date the recall began. Each recalled product is one record; these 8 records belong to 5 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 2 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 2 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0181-2025Class II | Olympus Corporation of the AmericasIt was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specif… | Aug 21, 2024 |
| Z-0329-2024Class II | Aizu Olympus Co., Ltd.Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in th… | Sep 25, 2023 |
| Z-0346-2024Class II | Aizu Olympus Co., Ltd.Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in th… | Sep 25, 2023 |
| Z-2269-2018Class II | Olympus Corporation of the AmericasIssuance of validated, new reprocessing procedures. This corrective action is being undertaken as a result of ongoing manufacturer and FDA activities relating… | May 11, 2018 |
| Z-2270-2018Class II | Olympus Corporation of the AmericasIssuance of validated, new reprocessing procedures. The new cleaning procedure requires additional recess flushing and forceps elevator raising/lowering steps… | May 11, 2018 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code FET. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 66 |
| 2018 | 67 |
| 2019 | 72 |
| 2020 | 3,043 |
| 2021 | 5,300 |
| 2022 | 6,156 |
| 2023 | 8,199 |
| 2024 | 13,153 |
| 2025 | 14,682 |
| 2026 (so far) | 14,883 |
Last 12 months by event type
This section didn’t load from openFDA. It’s retried within the hour.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 369in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 284registered establishments that list this product code
Other codes under 21 CFR 876.1500
Product codes classified under the same regulation.
Watch FET in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get FET every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in FET. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:FETRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:FETRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:FET · count decision_dateRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:FETRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:FET · count event_date_initiatedRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:FET · count res_event_numberRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0181-2025" "Z-0329-2024" "Z-0346-2024" "Z-2269-2018" "Z-2270-2018")Retrieved Nov 28, 2026, 11:27 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FET AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:12 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:FETRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:FETRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:876.1500Retrieved Nov 29, 2026, 8:27 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.