FDA product code
FAJ: Cystoscope And Accessories, Flexible/Rigid (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.1500
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 947 in the last 12 months
- Median review time
- 130 daysthe latest 25 decisions
- Recall records
- 220 in the last 12 months
- Adverse event reports
- 5,915last 12 months, 1 death
510(k) and De Novo decisions
Decisions for product code FAJ per year, by decision date, 94 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 2 |
| 2019 | 3 |
| 2020 | 2 |
| 2021 | 3 |
| 2022 | 5 |
| 2023 | 5 |
| 2024 | 9 |
| 2025 | 8 |
| 2026 (so far) | 7 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261779 | Single-use Cystoscope (PCS12-1, PCS12-2, PCS12-3, PCS22-1, PCS22-2, PCS22-3, PCS22-4, PCS22-5, PCS22-6, PCS24-1, PCS24-2, PCS24-3, PCS24-4, PCS24-5, PCS24-6, PCS12-1S, PCS12-2S, PCS12-3S, PCS22-1S, PCS22-2S, PCS22-3S, PCS22-4S, PCS22-5S, PCS22-6S, PCS24-1S, PCS24-2S, PCS24-3S, PCS24-4S, PCS24-5S, PCS24-6S)Hunan Endoso Life Technology Co., Ltd. | Nov 10, 2026165 days |
| K263075 | Strauss Surgical Cystoscope/Hysteroscope and Accessories InstrumentsStrauss Surgical USA | Oct 30, 202658 days |
| K260649 | OES ELITE Cystoscope and Accessories; Obturator, 17 Fr. (WA2C17BO); Obturator, 19.8 Fr. (WA2C19BO); Obturator, 21 Fr. (WA2C21BO); Obturator, 21 Fr., for sheath with ramp (WA2C21RO); Obturator, 21 Fr., long (WA2C21LO); Obturator, 21 Fr., for straight sheath (WA2C21SO); Obturator, 22.5 Fr. (WA2C22BO); Obturator, 25 Fr. (WA2C25BO); Obturator, optical, 19.8 Fr. (WA2C19BV); Obturator, optical, 21 Fr. (WA2C21BV); Obturator, optical, 21 Fr., for straight sheath (WA2C21SV); Obturator, optical,Olympus Winter & Ibe GmbH | May 21, 202683 days |
| K253696 | CFlex Cystoscope - Standard Deflection (0570-0455); CFlex Cystoscope - Reverse Deflection (0570-0456); CystoView Monitor (0570-0465)Verathon Medical (Canada) Ulc | May 11, 2026171 days |
| K253892 | Strauss Surgical Cystoscopes & Hysteroscopes (STS-2040-302ASA HD-endoscope, 4 x 302 mm, view. Dir. 0°, standard, autoclavable, STS-2040-302AA HD-endoscope, 4 x 302 mm, view. Dir. 0°, wide angle, autoclavable, STS-2040-302BA HD-endoscope, 4 x 302 mm, view. Dir. 30°, wide angle, autoclavable); Strauss Surgical Cystoscopes & Hysteroscopes (STS-2040-302BSA HD-endoscope, 4 x 302 mm, view. Dir. 30°,Strauss Surgical USA | Apr 16, 2026133 days |
| K253905 | PrimeSight UltraView SystemCogentix Medical, Inc. | Apr 7, 2026123 days |
| K260271 | Strauss Surgical cystoscope, hysteroscope and accessories InstrumentsAmerican Medical Endoscopy, Inc. | Mar 27, 202658 days |
| K253226 | UV5000W HandleUroviu Corporation | Oct 23, 202524 days |
| K251140 | Zenflow Spring ScopeZenflow, Inc. | Sep 23, 2025162 days |
| K252598 | Flexible Cystoscope (CY50H-20EU, CY50H-20US, CY55H-24EU, CY55H-24US)Shanghai AnQing Medical Instrument Co., Ltd. | Sep 15, 202528 days |
Recalls
Recall records for product code FAJ per year, by the date the recall began. Each recalled product is one record; these 22 records belong to 10 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 7 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 3 |
| 2023 | 7 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0684-2024Class II | Shirakawa Olympus Co., Ltd.Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible. | Dec 4, 2023 |
| Z-0685-2024Class II | Shirakawa Olympus Co., Ltd.Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible. | Dec 4, 2023 |
| Z-0689-2024Class II | Shirakawa Olympus Co., Ltd.Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible. | Dec 4, 2023 |
| Z-0690-2024Class II | Shirakawa Olympus Co., Ltd.Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible. | Dec 4, 2023 |
| Z-0691-2024Class II | Shirakawa Olympus Co., Ltd.Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible. | Dec 4, 2023 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code FAJ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 79 |
| 2018 | 79 |
| 2019 | 102 |
| 2020 | 339 |
| 2021 | 494 |
| 2022 | 744 |
| 2023 | 920 |
| 2024 | 893 |
| 2025 | 2,706 |
| 2026 (so far) | 5,633 |
Last 12 months by event type
- Deaths1
- Injuries64
- Malfunctions5,846
- Other or not stated4
Most reported device problems, last 12 months
- Detachment of Device or Device Component2,368
- Corroded794
- Loss of or Failure to Bond784
- Failure to Clean Adequately439
- Device Reprocessing Problem308
- Material Separation241
In plain English: what the newest reports describe Written by AI
Corrosion (rust-like damage to metal) was found on parts of the scope in several reports, including around the objective lens (the camera lens at the tip), the biopsy or forceps port, and the suction cylinder. Reports describe corrosion at the objective lens in more than one report.
5 reports: 3002808148-2026-37440, 3002808148-2026-37380, 3002808148-2026-37291, 3002808148-2026-37452, 3002808148-2026-37394
Adhesive or glue holding the bending section (the flexible part that bends) and the rubber covers was reported chipped, missing or loose in several reports. Some reports describe the bending section glue or rubber bonding area as missing or detached.
5 reports: 3002808148-2026-37596, 3002808148-2026-37505, 3002808148-2026-37303, 3002808148-2026-37378, 3002808148-2026-37151
The rubber cover or packing over the forceps channel port (the opening for passing instruments) was reported detached or lifted in more than one report, and one report also describes the cover being dented.
3 reports: 3002808148-2026-37479, 3002808148-2026-37451, 3002808148-2026-37442
Foreign objects or debris were found inside the scope in a couple of reports, at the distal end (the tip) or under the cover glass (the clear window over the lens). One report describes the distal end as having external detachment, which is a separate problem.
2 reports: 3002808148-2026-37660, 3002808148-2026-37524
In one report, the channel mount inside the videoscope was found deformed (bent out of shape) during testing, after the scope was returned with a blurry image. No patient involvement was reported in any of these reports.
Report: 3002808148-2026-37391
Written by AI from the narratives of the 19 newest reports that have one, received Oct 27, 2026 to Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 1,449in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 284registered establishments that list this product code
Other codes under 21 CFR 876.1500
Product codes classified under the same regulation.
Watch FAJ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get FAJ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in FAJ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:FAJRetrieved Nov 29, 2026, 5:09 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:FAJRetrieved Nov 29, 2026, 5:09 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:FAJ · count decision_dateRetrieved Nov 29, 2026, 5:09 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:FAJRetrieved Nov 29, 2026, 5:09 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:FAJ · count event_date_initiatedRetrieved Nov 29, 2026, 5:09 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:FAJ · count res_event_numberRetrieved Nov 29, 2026, 5:09 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0684-2024" "Z-0685-2024" "Z-0689-2024" "Z-0690-2024" "Z-0691-2024")Retrieved Nov 29, 2026, 5:09 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FAJ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:09 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FAJ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:09 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FAJ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:09 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:FAJRetrieved Nov 29, 2026, 5:09 AM UTC. FDA data updated Nov 2, 2026.