FDA device news · Product code DXN
FDA 510(k) clearance
Automatic Electronic Blood Pressure Monitor (U90B series models including ( U90B,U90B Pro,U90B Plus,U90B Ultra,U90C,U90C Pro,U90C Plus,U90C Ultra).) (K260273): Shenzhen Urion Technology Co., Ltd. 510(k) clearance
Shenzhen, CN
- Decision
- Substantially Equivalent
- Decision date
- Mar 26, 2026
- Received
- Jan 29, 2026
- 510(k) type
- Traditional
- Product code
- DXNSystem, Measurement, Blood-Pressure, Non-Invasive
- Review panel
- Cardiovascular
- Device class
- Class II
- Regulation
- 21 CFR 870.1130
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 56 daysreceipt to decision, calendar days
- Median in DXN
- 148 daysthe latest 24 other decisions
- DXN decisions
- 31in the last 12 months
- Applicant’s 510(k)s
- 8in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- An automatic upper arm blood pressure monitor (U90B series, including U90B, U90B Pro, U90B Plus, U90B Ultra, U90C, U90C Pro, U90C Plus and U90C Ultra). An inflatable cuff wraps around the upper arm and an internal pump inflates it.
- What it’s cleared for
- The device is intended to measure systolic blood pressure, diastolic blood pressure and pulse rate through the upper arm. It is expected to be used in the home or in a hospital, for people over 12 years old. The summary says it is not intended to be a diagnostic device.
- Compared with (the predicate)
- K243115, Upper Arm Electronic Blood Pressure Monitor
- How it differs
- The summary says the device differs from the predicate in cuff size (220 to 400 mm for the subject device versus 220 to 420 mm for the predicate), power supply (some modes use AAA batteries), and operating environment (a lower maximum humidity of 93% versus 85%, as listed in the table; the table lists the subject's humidity range as 15% to 93%).
- Testing the summary lists
- Electrical safety and electromagnetic compatibility (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-30)
- Biocompatibility (ISO 10993-1; cytotoxicity, sensitization and irritation)
- Cuff size distribution testing across the arm circumference range in the clinical study
- Blood pressure range distribution testing in the clinical study
- Software and performance validation for the larger arm circumference range (IEC 80601-2-30 and ISO 81060-2)
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K260273 took 56 calendar days from receipt to decision. Of the latest 24 other decisions in product code DXN, 23 took longer, 1 was faster; their median was 148 days.
Latest decisions in DXN
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260670 | YUWELL® Electronic Blood Pressure MonitorJiangsu Yuyue Medical Equipment & Supply Co., Ltd. · Traditional | Oct 28, 2026240 days |
| K260470 | Electronic Blood Pressure Monitor (BP-204C, BP-204A, BP-215, BP-203C, BP-205A, BP-201C, BP-217, BP-202C, BP-202 BP-205, BP-205B,BP-206)Shenzhen Brav Electronic Technologies Co., Ltd. · Traditional | Oct 1, 2026231 days |
| K261929 | Upper Arm Blood pressure monitor (BP7155D)OMRON Healthcare, Inc. · Special | Sep 10, 202693 days |
| K254000 | Happy Health Blood Pressure MonitorHappy Health, Inc. · Traditional | Sep 3, 2026262 days |
| K262658 | Automatic Digital Blood Pressure monitor (LD-526)Honsun (Nantong) Co., Ltd. · Special | Aug 31, 202632 days |
| K261394 | Arm-type Fully Automatic Digital Blood Pressure Monitor (DBP-6186 Plus, DBP-6286B Plus, DBP-61E3 Plus, DBP-61E2 Plus, DBP-62E3B Plus, DBP-62E2B Plus)JOYTECH Healthcare Co., Ltd. · Traditional | Aug 26, 2026119 days |
| K253425 | Arm-type Fully Automatic Digital Blood Pressure Monitor (DBP-6194-P,DBP-6294B-P,DBP-6195P,DBP-6295B-P, DBP-6196P,DBP-6296B-P,DBP-61E1-P, DBP-62E1B-P,DBP-62E3B-P,DBP-61E2-P,DBP-62E2B-P,DBP-61E3-P,DBP-6173-P,DBP-6273B-P,DBP-6179-P,DBP-6279B-P)Joytech Healthcare Co. , Ltd. · Traditional | Aug 13, 2026316 days |
| K261061 | BP Go (BPP-855)Wellvii, Inc. · Traditional | Aug 13, 2026135 days |
| K261082 | Microlife Wrist Blood Pressure Monitor , (Model BPHJG5-G)Microlife Corporation · Traditional | Jul 23, 2026113 days |
| K261136 | Blood Pressure Monitor (M501,M501A,M502,M502A,M502C,M502D)Guangzhou Daxin Health Technology Co., Ltd. · Traditional | Jul 23, 2026107 days |
Other 510(k)s from Shenzhen Urion Technology Co., Ltd.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K254048 | Wrist Watch Electronic Blood Pressure Monitor (U16H, U16L, U16P, U16W, U19M, U19R, U19S, and U19T)Shenzhen Urion Technology Co., Ltd. · Traditional | May 14, 2026148 days |
| K250046 | Air Compression Leg Massager (UM605, UM606, UM607, UM608,UM609,UM610, UM611, UM617)Shenzhen Urion Technology Co., Ltd. · Traditional | Apr 25, 2025105 days |
| K243115 | Upper Arm Electronic Blood Pressure Monitor (U87Y series (models including U80Y, U81Y, U82Y,U83Y,U86Y U80N, U81NH), U86E series (including U82E, U80E, U80EH, U81E, U83E, U85E, U80L, U87E ), U81X series (including U81X, U80X, U82X, U83X, U81D, U82D, U83D, U81RH, U82RH) and U83Z series (including U83Z, U80Z, U81Z, U82Z, U85Z, U86Z and U87Z))Shenzhen Urion Technology Co., Ltd. · Traditional | Feb 21, 2025144 days |
| K241007 | Upper Arm Electronic Blood Pressure Monitor (U86E and TelliBP01)Shenzhen Urion Technology Co., Ltd. · Traditional | Oct 24, 2024195 days |
| K200802 | Infrared ThermometerShenzhen Urion Technology Co., Ltd. · Traditional | Oct 7, 2020194 days |
| K160019 | U80 Series Upper Arm Electronic Blood Pressure MonitorShenzhen Urion Technology Co., Ltd. · Traditional | Apr 22, 2016108 days |
| K141683 | ELECTRONIC BLOOD PRESSURE MONITORShenzhen Urion Technology Co., Ltd. · Traditional | Apr 3, 2015284 days |
Watch DXN in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get DXN every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in DXN. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K260273Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:DXNRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:DXNRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:DXN AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Shenzhen Urion Technology Co., Ltd."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil