FDA device news · Product code DXN
FDA 510(k) clearance
Automatic Digital Blood Pressure monitor (LD-526) (K262658): Honsun (Nantong) Co., Ltd. 510(k) clearance
Nantong, CN
- Decision
- Substantially Equivalent
- Decision date
- Aug 31, 2026
- Received
- Jul 30, 2026
- 510(k) type
- Special
- Product code
- DXNSystem, Measurement, Blood-Pressure, Non-Invasive
- Review panel
- Cardiovascular
- Device class
- Class II
- Regulation
- 21 CFR 870.1130
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 32 daysreceipt to decision, calendar days
- Median in DXN
- 148 daysthe latest 24 other decisions
- DXN decisions
- 31in the last 12 months
- Applicant’s 510(k)s
- 13in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- An automatic digital blood pressure monitor, model LD-526, made by HONSUN (Nantong) Co., Ltd. It uses the oscillometric method: a pump inflates an arm cuff, an electronic valve lets it deflate, and software reads the pressure pulses from the artery to work out systolic and diastolic pressure and pulse rate.
- What it’s cleared for
- It is a non-invasive blood pressure monitor for measuring systolic and diastolic arterial blood pressure and pulse rate in adults.
- Compared with (the predicate)
- K170466, Automatic Digital Blood Pressure Monitor LD-537
- How it differs
- The summary says the LD-526 has a different measuring range presentation (systolic and diastolic ranges listed separately), a different battery type (4 AAA instead of 4 AA) and a Type-C charging port that is for power delivery only, different operating and transport/storage conditions, a smaller and lighter body, and a different cuff range arrangement.
- Testing the summary lists
- Electrical safety (ANSI/AAMI ES 60601-1)
- Electromagnetic compatibility (IEC 60601-1-2)
- Home healthcare environment requirements (IEC 60601-1-11)
- Automated non-invasive sphygmomanometer requirements (ISO 80601-2-30), listed in the standards table
- Software verification and validation (FDA software guidance, May 2005)
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K262658 took 32 calendar days from receipt to decision. Of the latest 24 other decisions in product code DXN, 24 took longer; their median was 148 days.
Latest decisions in DXN
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260670 | YUWELL® Electronic Blood Pressure MonitorJiangsu Yuyue Medical Equipment & Supply Co., Ltd. · Traditional | Oct 28, 2026240 days |
| K260470 | Electronic Blood Pressure Monitor (BP-204C, BP-204A, BP-215, BP-203C, BP-205A, BP-201C, BP-217, BP-202C, BP-202 BP-205, BP-205B,BP-206)Shenzhen Brav Electronic Technologies Co., Ltd. · Traditional | Oct 1, 2026231 days |
| K261929 | Upper Arm Blood pressure monitor (BP7155D)OMRON Healthcare, Inc. · Special | Sep 10, 202693 days |
| K254000 | Happy Health Blood Pressure MonitorHappy Health, Inc. · Traditional | Sep 3, 2026262 days |
| K262658 | Automatic Digital Blood Pressure monitor (LD-526)Honsun (Nantong) Co., Ltd. · Special | Aug 31, 202632 days |
| K261394 | Arm-type Fully Automatic Digital Blood Pressure Monitor (DBP-6186 Plus, DBP-6286B Plus, DBP-61E3 Plus, DBP-61E2 Plus, DBP-62E3B Plus, DBP-62E2B Plus)JOYTECH Healthcare Co., Ltd. · Traditional | Aug 26, 2026119 days |
| K253425 | Arm-type Fully Automatic Digital Blood Pressure Monitor (DBP-6194-P,DBP-6294B-P,DBP-6195P,DBP-6295B-P, DBP-6196P,DBP-6296B-P,DBP-61E1-P, DBP-62E1B-P,DBP-62E3B-P,DBP-61E2-P,DBP-62E2B-P,DBP-61E3-P,DBP-6173-P,DBP-6273B-P,DBP-6179-P,DBP-6279B-P)Joytech Healthcare Co. , Ltd. · Traditional | Aug 13, 2026316 days |
| K261061 | BP Go (BPP-855)Wellvii, Inc. · Traditional | Aug 13, 2026135 days |
| K261082 | Microlife Wrist Blood Pressure Monitor , (Model BPHJG5-G)Microlife Corporation · Traditional | Jul 23, 2026113 days |
| K261136 | Blood Pressure Monitor (M501,M501A,M502,M502A,M502C,M502D)Guangzhou Daxin Health Technology Co., Ltd. · Traditional | Jul 23, 2026107 days |
Other 510(k)s from Honsun (Nantong) Co., Ltd.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K252399 | Portable Oxygen ConcentratorHonsun (Nantong) Co., Ltd. · Traditional | Jun 17, 2026321 days |
| K260260 | Automatic Digital Blood Pressure Monitor (LD-587, LD-528, LD-562, LD-581, LD-521, LD-527, LD-572 & LD-569)Honsun (Nantong) Co., Ltd. · Special | Apr 28, 202690 days |
| K251795 | Wrist automatic blood pressure monitor (LD-735, LD-752, LD-753)Honsun (Nantong) Co., Ltd. · Special | Oct 9, 2025119 days |
| K211629 | Infrared forehead thermometer LD-FT-100BHonsun (Nantong) Co., Ltd. · Traditional | Dec 8, 2021195 days |
| K192500 | Aneroid sphygmomanometer with stethoscope, Aneroid sphygmomanometerHonsun (Nantong) Co., Ltd. · Traditional | Apr 6, 2020207 days |
| K170866 | Scian Automatic Digital Blood Pressure Monitor, Model: LD-576, LD-575 & LD-579Honsun (Nantong) Co., Ltd. · Special | Feb 21, 2018335 days |
| K170466 | Automatic Blood Pressure Monitor Model LD-518 & LD-537Honsun (Nantong) Co., Ltd. · Traditional | Oct 27, 2017254 days |
| K170454 | Automatic Blood Pressure Monitor Model LD-562U, LD-588U & LD-533UHonsun (Nantong) Co., Ltd. · Traditional | Oct 25, 2017252 days |
| K132627 | AUTOMATIC DIGITAL BLOOD PRESSURE MONITORHonsun (Nantong) Co., Ltd. · Special | Oct 23, 201362 days |
| K130925 | AUTOMATIC BLOOD PRESSURE MONITORHonsun (Nantong) Co., Ltd. · Traditional | Sep 12, 2013162 days |
| K131463 | WRIST AUTOMATIC BLOOD PRESSURE MONITORHonsun (Nantong) Co., Ltd. · Traditional | Jul 8, 201348 days |
| K092245 | ANEROID SPHYGMOMANOMETER WITH STETHOSCOPE, MODEL LD-100Honsun (Nantong) Co., Ltd. · Traditional | Oct 28, 200992 days |
Watch DXN in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get DXN every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in DXN. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K262658Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:DXNRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:DXNRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:DXN AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Honsun Nantong Co., Ltd."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.