FDA device news · Product code DSA

FDA 510(k) clearance

IBP cable Becton Dickinson (2606488); IBP cable Edwards (2606489); IBP cable Abbott, Medex (2606490); IBP cable Utah (2606491); IBP Adapter Cable Draeger 7pin (2607558); IBP Adapter Cable Draeger 10pin (2607559) (K252827): Drägerwerk AG & Co. KGaA 510(k) clearance

Luebeck, DE

Follow DSA: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Dec 1, 2025
Received
Sep 5, 2025
510(k) type
Traditional
Product code
DSACable, Transducer And Electrode, Patient, (Including Connector)
Review panel
Cardiovascular
Device class
Class II
Regulation
21 CFR 870.2900
Third-party review
No
Summary or statement
Summary
Review time
87 daysreceipt to decision, calendar days
Median in DSA
133 daysthe latest 24 other decisions
DSA decisions
3in the last 12 months
Applicant’s 510(k)s
5in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
IBP cables and IBP adapter cables are passive electrical cables that carry the electrical signal from an invasive blood pressure (IBP) transducer to the Draeger Vista 300 Patient Monitor. Six versions are included, each with a different connector, and they do not change the signal as it passes through.
What it’s cleared for
The cables are intended to forward electrical signals from connected IBP pressure transducer sets to the patient monitor. The summary states the device is for intrahospital use only.
Compared with (the predicate)
  • K162768, IBP transducers interface cables (INTEGRAL PROCESS SAS)
How it differs
The summary says the subject cables have a 3.8 m length, while the predicate comes in various specified standard lengths. The connector overmolding material also differs: the subject uses TPU and the predicate uses PA.
Testing the summary lists
  • Biocompatibility (ISO 10993-1; cytotoxicity ISO 10993-5; sensitization and irritation ISO 10993-10)
  • Electrical safety and EMC (IEC 60601-1; IEC 60601-1-2)
  • Invasive blood pressure monitoring equipment safety (IEC 60601-2-34)
  • Electromagnetic immunity (TS 60601-4-2)
  • Reprocessing and cleaning (ISO 17664-2)
  • Mechanical and electrical function testing; shipping packaging (ISTA 3A)

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K252827 took 87 calendar days from receipt to decision. Of the latest 24 other decisions in product code DSA, 20 took longer, 4 were faster; their median was 133 days.

Latest decisions in DSA

NumberDevice and applicantDecision and review time
K262816 EXG PulseLinQ Cables (EXGPL)Cb Innovations, LLC · Traditional Oct 30, 202684 days
K254114 Baylis Connector CableBaylis Medical Technologies, Inc. · Special Jan 18, 202630 days
K252827 IBP cable Becton Dickinson (2606488); IBP cable Edwards (2606489); IBP cable Abbott, Medex (2606490); IBP cable Utah (2606491); IBP Adapter Cable Draeger 7pin (2607558); IBP Adapter Cable Draeger 10pin (2607559)Drägerwerk AG & Co. KGaA · Traditional Dec 1, 202587 days
K251669 IBP transducer adapter cable, Argon/BD/Ohmeda, 2m; IBP transducer adapter cable, Draeger 10-pin, 2m; IBP transducer adapter cable, Draeger 7-pin, 2m; IBP transd. adapter cbl, Abbott/Medex, Smiths/TranStar 2m; IBP transducer adapter cable, Baxter/Edwards, 0.3m; IBP transducer adapter cable, Baxter/Edwards, 2m; IBP transducer adapter cable, Argon/BD/Ohmeda, 0.3m; IBP transducer adapter cable, Draeger 10-pin, 0.3m; IBP transducer adapter cable, Draeger 7-pin, 0.3m; IBP transducer adapterShanghai Draeger Medical Instrument Co., Ltd. · Traditional Sep 30, 2025123 days
K242986 SpO2 Extension CableBeijing Rongrui-Century Science & Technology Co., Ltd. · Special May 2, 2025218 days
K243545 Philips Reusable ECG Lead Sets and Trunk Cables : Model Number,Model Name: 989803170171, OR 3-Lead ECG Trunk Cable, AAMI/IEC; 989803170181, OR 5-Lead ECG Trunk Cable, AAMI/IEC; M1667A, 6 Lead ECG Trunk, AAMI/IEC 2.7m; M1668A,5 Lead ECG Trunk, AAMI/IEC 2.7m; M1669A,3 Lead ECG Trunk, AAMI/IEC 2.7m; M1949A,5+5 ECG Trunk cable AAMI, IEC 2.7m; M1663A,10 Lead ECG Trunk AAMI/IEC 2m; M1665A, 6+4 Lead ECG Trunk AAMI/IEC 2.7m; M1671A, 3 Leadset, Grabber, AAMI, ICU; M1673A, 3 Leadset, SnPhilips Medizin Systeme Böblingen GmbH · Traditional Jan 17, 202563 days
K240087 RT Carbon ECG LeadsIvy Biomedical Systems, Inc. · Traditional Sep 13, 2024246 days
K240377 LIFEPAK® ECG CablesStryker Physio-Control · Traditional May 7, 202490 days
K240156 LIFEPAK® Invasive Pressure Adapter Cable (Interfaces with ICU Medical Transpack IV Invasive Pressure Transducers (or equivalent)) (11230-000021); LIFEPAK® Invasive Pressure Adapter Cable (interfaces with Edwards Lifesciences TruWave Invasive Pressure Transducers (or equivalent)) (11230-000024)Physio-Control, Inc. · Traditional Apr 30, 2024102 days
K222712 Fogg System Patient Monitoring CablesFogg System Company, Inc. · Traditional Nov 22, 2023440 days

Other 510(k)s from Drägerwerk AG & Co. KGaA

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K252746 Atlan (A100, A100 XL, A350, A350 XL)Drägerwerk AG Co. Kgaa · Traditional May 21, 2026265 days
K251619 Babyleo TN500Drägerwerk AG & Co. KGaA · Traditional Feb 13, 2026262 days
K230931 AtlanDrägerwerk AG Co. Kgaa · Traditional Jul 23, 2023111 days
K221836 Filter CareStar Plus, Filter SafeStar Plus, Filter/HME TwinStar PlusDrägerwerk AG & Co. KGaA · Traditional Dec 7, 2022167 days

Watch DSA in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get DSA every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in DSA. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.