FDA device news · Product code BSZ
FDA 510(k) clearance
Atlan (A100, A100 XL, A350, A350 XL) (K252746): Drägerwerk AG Co. Kgaa 510(k) clearance
Lübeck, DE
- Decision
- Substantially Equivalent
- Decision date
- May 21, 2026
- Received
- Aug 29, 2025
- 510(k) type
- Traditional
- Product code
- BSZGas-Machine, Anesthesia
- Review panel
- Anesthesiology
- Device class
- Class II
- Regulation
- 21 CFR 868.5160
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 265 daysreceipt to decision, calendar days
- Median in BSZ
- 204 daysthe latest 24 other decisions
- BSZ decisions
- 2in the last 12 months
- Applicant’s 510(k)s
- 5in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The Atlan anesthesia workstation is a machine that receives medical gases, mixes them, and delivers the mixture to a patient's breathing circuit, along with volatile anesthetic agents and ventilation support.
- What it’s cleared for
- The device is cleared for anesthetizing adults, pediatric patients, and neonates, and for mechanical, manual, pressure-supported, and spontaneous breathing. It provides basic ventilation and device monitoring, and optional patient-gas measurement and O2 insufflation.
- How it differs
- The summary says the proposed device adds a mechanically controlled gas mixer with flow tubes for the A100/A100 XL, and that the submission includes modifications made since K230931. The proposed device runs software version SW 2.11.00, while the predicate runs SW 2.00.01.
- Testing the summary lists
- Electrical safety and electromagnetic compatibility (IEC 60601-1, IEC 60601-1-2, IEC TS 60601-4-2)
- Anesthetic workstation and respiratory gas monitor particular requirements (ISO 80601-2-13, ISO 80601-2-55)
- Usability (IEC 62366-1, IEC 60601-1-6)
- Alarm systems (IEC 60601-1-8)
- Software lifecycle (IEC 62304)
- Biocompatibility of breathing gas pathways (ISO 10993-1, ISO 18562-1, ISO 18562-4)
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K252746 took 265 calendar days from receipt to decision. Of the latest 24 other decisions in product code BSZ, 8 took longer, 16 were faster; their median was 204 days.
Latest decisions in BSZ
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K253330 | A5 Anesthesia System (A5); A7 Anesthesia System (A7)Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Traditional | Jun 25, 2026268 days |
| K252746 | Atlan (A100, A100 XL, A350, A350 XL)Drägerwerk AG Co. Kgaa · Traditional | May 21, 2026265 days |
| K251352 | Codonics Safe Labeling System-Point of Care Station (PCS) (SLS-700i and related 610i, 620i, 630i series enabled to retrofit RFID features, SLS-1, SLS XXX, Multiple Models Pending Configuration)Codonics.Incorporated · Traditional | Jul 11, 202572 days |
| K230931 | AtlanDrägerwerk AG Co. Kgaa · Traditional | Jul 23, 2023111 days |
| K213867 | Carestation 750/750cDatex-Ohmeda, Inc. · Traditional | Mar 28, 2023473 days |
| K201957 | A8, A9 Anesthesia SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. · Traditional | Mar 26, 2021255 days |
| K191027 | Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia SystemMaquet Critical Care AB · Traditional | Mar 18, 2020335 days |
| K171292 | A7 Anesthesia SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. · Traditional | Jan 12, 2018255 days |
| K171311 | A5 Anesthesia SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. · Traditional | Jan 12, 2018255 days |
| K172045 | Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900, Aespire View, Avance, Avance CS2, Aisys, Aisys CS2Datex-Ohmeda, Inc. · Traditional | Nov 3, 2017120 days |
Other 510(k)s from Drägerwerk AG Co. Kgaa
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K251619 | Babyleo TN500Drägerwerk AG & Co. KGaA · Traditional | Feb 13, 2026262 days |
| K252827 | IBP cable Becton Dickinson (2606488); IBP cable Edwards (2606489); IBP cable Abbott, Medex (2606490); IBP cable Utah (2606491); IBP Adapter Cable Draeger 7pin (2607558); IBP Adapter Cable Draeger 10pin (2607559)Drägerwerk AG & Co. KGaA · Traditional | Dec 1, 202587 days |
| K230931 | AtlanDrägerwerk AG Co. Kgaa · Traditional | Jul 23, 2023111 days |
| K221836 | Filter CareStar Plus, Filter SafeStar Plus, Filter/HME TwinStar PlusDrägerwerk AG & Co. KGaA · Traditional | Dec 7, 2022167 days |
Watch BSZ in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get BSZ every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in BSZ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K252746Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:BSZRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:BSZRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:BSZ AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Drägerwerk AG Co. Kgaa"Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil