FDA device news · Product code PZP

FDA 510(k) clearance

HYDROS Robotic System (HY1000); HYDROS Handpiece (HH1000); HYDROS TRUS Probe (HU1000) (K260723): Procept Biorobotics 510(k) clearance

San Jose, CA

Follow PZP: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Jul 9, 2026
Received
Mar 5, 2026
510(k) type
Traditional
Product code
PZPFluid Jet Removal System
Review panel
General, Plastic Surgery
Device class
Class II
Regulation
21 CFR 876.4350
Third-party review
No
Summary or statement
Summary
Review time
126 daysreceipt to decision, calendar days
Median in PZP
173 daysthe latest 8 other decisions
PZP decisions
2in the last 12 months
Applicant’s 510(k)s
9in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The HYDROS Robotic System is a robotic system with a TRUS probe and a single-use handpiece for removing prostate tissue. The handpiece goes through the urethra and uses a jet of water to cut and remove tissue, guided by ultrasound imaging from the transrectal ultrasound (TRUS) probe, which is an ultrasound probe placed in the rectum. The handpiece has a built-in camera to view the surgical site.
What it’s cleared for
The device is cleared to resect and remove prostate tissue in adult males with lower urinary tract symptoms caused by benign prostatic hyperplasia (an enlarged, non-cancerous prostate). It is intended for use by a urologist supported by operating room staff who is trained in endoscopic prostate surgery and in TRUS imaging.
Compared with (the predicate)
How it differs
The summary says the differences from the predicate are limited to changes to various waterjet parameters and expanded automated registration features associated with FirstAssist AI. It states the device has the same indications for use and principles of operation as the predicate.
Testing the summary lists
  • Simulated use testing
  • Cadaver testing
  • EMC testing (IEC 60601-1-2:2020)
  • Software testing
  • Cybersecurity testing

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K260723 took 126 calendar days from receipt to decision. Of the latest 8 other decisions in product code PZP, 6 took longer, 2 were faster; their median was 173 days.

Latest decisions in PZP

NumberDevice and applicantDecision and review time
K260723 HYDROS Robotic System (HY1000); HYDROS Handpiece (HH1000); HYDROS TRUS Probe (HU1000)Procept Biorobotics · Traditional Jul 9, 2026126 days
K253211 AQUABEAM Robotic SystemProcept Biorobotics · Traditional Jun 23, 2026267 days
K251082 HYDROS Robotic System (HY1000); HYDROS Handpiece (HH1000); HYDROS TRUS Probe (HU1000)Procept Biorobotics · Traditional Oct 10, 2025184 days
K241952 AQUABEAM Robotic System (AB2000)Procept Biorobotics · Special Sep 30, 202489 days
K240200 HYDROS Robotic System; HYDROS TRUS Probe; HYDROS HandpieceProcept Biorobotics · Traditional Aug 20, 2024208 days
K231024 AquaBeam Robotic SystemProcept Biorobotics, Corporation · Traditional Aug 30, 2023141 days
K212835 AquaBeam Robotic SystemProcept Biorobotics, Corporation · Special Oct 6, 202129 days
K202961 AquaBeam Robotic SystemProcept Biorobotics, Corporation · Traditional Mar 11, 2021162 days
DEN170024 AQUABEAM SystemProcept Biorobotics, Corporation · Direct Dec 21, 2017248 days

Other 510(k)s from Procept Biorobotics

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K253211 AQUABEAM Robotic SystemProcept Biorobotics · Traditional Jun 23, 2026267 days
K251082 HYDROS Robotic System (HY1000); HYDROS Handpiece (HH1000); HYDROS TRUS Probe (HU1000)Procept Biorobotics · Traditional Oct 10, 2025184 days
K241952 AQUABEAM Robotic System (AB2000)Procept Biorobotics · Special Sep 30, 202489 days
K240200 HYDROS Robotic System; HYDROS TRUS Probe; HYDROS HandpieceProcept Biorobotics · Traditional Aug 20, 2024208 days
K231024 AquaBeam Robotic SystemProcept Biorobotics, Corporation · Traditional Aug 30, 2023141 days
K212835 AquaBeam Robotic SystemProcept Biorobotics, Corporation · Special Oct 6, 202129 days
K202961 AquaBeam Robotic SystemProcept Biorobotics, Corporation · Traditional Mar 11, 2021162 days
DEN170024 AQUABEAM SystemProcept Biorobotics, Corporation · Direct Dec 21, 2017248 days

Watch PZP in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get PZP every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in PZP. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.