FDA product code
PPB: Foley Catheter Kit (Excludes Hiv Testing) (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.5130
- Review panel
- Gastroenterology and Urology
- Premarket route
- Not stated
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 00 in the last 12 months
- Median review time
- Not availablethe latest 0 decisions
- Recall records
- 143 in the last 12 months
- Adverse event reports
- 29last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code PPB per year, by decision date. Review time is the days from receipt to decision.
No decisions on record for this product code.
Recalls
Recall records for product code PPB per year, by the date the recall began. Each recalled product is one record; these 14 records belong to 14 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 3 |
| 2024 | 5 |
| 2025 | 2 |
| 2026 (so far) | 3 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2024-2026Class II | Medline Industries, LPDuring an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market… | Mar 19, 2026 |
| Z-1396-2026Class II | Medline Industries, LPMedline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated ste… | Jan 7, 2026 |
| Z-2117-2026Class II | Medline Industries, LPMedline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated ste… | Jan 7, 2026 |
| Z-1599-2025Class II | MEDLINE INDUSTRIES, LP - NorthfieldThere is a potential for open/weak seals on purchased breather pouches used for packaging of select sterile convenience kits. | Mar 3, 2025 |
| Z-1443-2025Class II | MEDLINE INDUSTRIES, LP - NorthfieldMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have bee… | Feb 11, 2025 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PPB. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 17 |
| 2024 | 50 |
| 2025 | 13 |
| 2026 (so far) | 27 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions26
- Other or not stated2
Most reported device problems, last 12 months
- Fluid/Blood Leak6
- Deflation Problem3
- Detachment of Device or Device Component3
- Device Dislodged or Dislocated3
- Difficult to Remove2
- Adverse Event Without Identified Device or Use Problem1
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 358in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 25registered establishments that list this product code
Other codes under 21 CFR 876.5130
Product codes classified under the same regulation.
Watch PPB in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PPB every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PPB. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PPBRetrieved Nov 29, 2026, 1:13 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PPBRetrieved Nov 29, 2026, 4:28 PM UTC.
- device/510k · search product_code:PPB · count decision_dateRetrieved Nov 29, 2026, 1:13 AM UTC.
- device/recall · search product_code:PPBRetrieved Nov 29, 2026, 4:28 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PPB · count event_date_initiatedRetrieved Nov 29, 2026, 1:13 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PPB · count res_event_numberRetrieved Nov 29, 2026, 1:13 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2024-2026" "Z-1396-2026" "Z-2117-2026" "Z-1599-2025" "Z-1443-2025")Retrieved Nov 29, 2026, 4:28 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PPB AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:28 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PPB AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:28 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PPB AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 1:13 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PPBRetrieved Nov 29, 2026, 1:13 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PPBRetrieved Nov 29, 2026, 1:13 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:876.5130Retrieved Nov 29, 2026, 5:58 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.