Product codes

FDA product code

EYC: Catheter, Upper Urinary Tract (FDA product code)

Follow EYC: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 876.5130
Review panel
Gastroenterology and Urology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
50 in the last 12 months
Median review time
182 daysthe latest 5 decisions
Recall records
00 in the last 12 months
Adverse event reports
9last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code EYC per year, by decision date, 5 since 1989. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20201
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K200757 Foley Catheter with Temperature SensorDeroyal Industries, Inc. Oct 15, 2020205 days
K061918 FOLEY CATHETERS WITH TEMPERATURE PROBEGE Healthcare Aug 18, 200643 days
K041416 EXAC-TEMP FOLEY CATHETER AND CLINI-TEMP FOLEY CATHETERDeroyal Jul 2, 200436 days
K925790 EXAC-TEMP & CLINI-TEMP FOLEY CATHETER'SDeroyal Industries, Inc. May 17, 1993182 days
K890310 STERILE FOLEY CATHETERKosan Far East Ent. , Ltd. Nov 30, 1989311 days

Recalls

Recall records for product code EYC per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code EYC. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20171
20183
20193
20206
20216
20220
20230
202426
202515
2026 (so far)8

Last 12 months by event type

  • Deaths0
  • Injuries4
  • Malfunctions5
  • Other or not stated0

Most reported device problems, last 12 months

  1. Blocked Connection3
  2. Material Separation2
  3. Device Dislodged or Dislocated1
  4. Difficult to Remove1
  5. Entrapment of Device1
  6. Insufficient Device Problem Information1

In plain English: what the newest reports describe Written by AI

  • The temperature sensor built into some of these catheters has failed in more than one way. Reports describe a wire left exposed, a probe that read 11.4 after 37.0 and then stopped working, and a probe that would not read at all once connected.

    6 reports: 2320762-2025-00014, 22736678, 2320762-2025-00015, 22838264, 2320762-2025-00009, 22344851

  • The balloon that holds the catheter in the bladder has deflated, burst or leaked in several reports, so the catheter can slide out or sit out of place. In one case a burst balloon was found after surgery, with the catheter no longer in the patient.

    5 reports: 2320762-2025-00010, 2320762-2025-00017, 2320762-2025-00011, 2320762-2025-00012, 2320762-2025-00021

  • Catheter ends have come apart. One report describes the catheter splitting at the joint where the balloon and temperature probe attach, leaving the lower part still in the patient; another says the ends keep coming apart.

    2 reports: 2320762-2026-00009, 2320762-2025-00013

  • Some catheters became blocked, reducing urine output. A distributing facility reported three patients whose catheters were all found blocked, and another report describes no urine for six hours until the catheter was changed and 1100 mL was drained.

    2 reports: 2320762-2026-00006, 2320762-2025-00008

  • One report describes a second-degree burn to a patient's penis and scrotum, found after surgery. The report says a temperature-probe catheter was in place during the operation but does not state how the burn happened.

    Report: 2320762-2026-00008

Written by AI from the narratives of the 18 newest reports that have one, received Jun 27, 2025 to Sep 21, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
29in the FDA’s Global Unique Device Identification Database
Establishment listings
21registered establishments that list this product code

Watch EYC in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get EYC every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in EYC. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.