FDA product code
PBY: Temporary Tissue Marker (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 878.4300
- Review panel
- General and Plastic Surgery
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 70 in the last 12 months
- Median review time
- 135 daysthe latest 7 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 1last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code PBY per year, by decision date, 7 since 2013. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 1 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K223682 | Sirius Pintuition Seed, Sirius Pintuition Probe, Sirius Pintuition Base UnitSirius Medical Systems B.V. | Nov 9, 2023336 days |
| K222643 | Sirius Pintuition Seed, Sirius Pintuition Probe, Sirius Pintuition Base UnitSirius Medical Systems B.V. | Oct 19, 202248 days |
| K200734 | Sirius Pintuition Seed, Sirius Pintuition Probe, Sirius Pintuition Base UnitSirius Medical Systems B.V. | Feb 24, 2021341 days |
| K163667 | RFID Localization System- Localizer Reader, RFID Localization System- Localizer Surgical Probe, RFID Localization System- Localizer Surgical Probe (5 pack), RFID Localization System- Tag Applicator, RFID Localization System -Tag Applicator (10 pack)Health Beacons, Inc. | Apr 28, 2017122 days |
| K163541 | Magseed Magnetic Marker SystemEndomagnetics Ltd., | Jan 13, 201728 days |
| K153044 | Sentimag System, Sentimark Magnetic Marker SystermEndomagnetics Ltd., | Mar 2, 2016135 days |
| K120804 | CIANNA MEDICAL TISSUE MARKER AND DELIVERY SYSTEMCianna Medical | Feb 4, 2013325 days |
Recalls
Recall records for product code PBY per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PBY. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 1 |
| 2025 | 1 |
| 2026 (so far) | 1 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions0
- Other or not stated0
Most reported device problems, last 12 months
- Material Fragmentation1
In plain English: what the newest reports describe Written by AI
One report says a marker's glass casing broke as it was released from its applicator in a dense breast, leaving glass fragments behind. The marker then stopped sending its signal. It was removed with the tissue, and a follow-up mammogram was clear.
Report: 3013649990-2019-00001
One report says white threads at the tip of a probe came off during a procedure and went into the patient. The threads were removed before the incision was closed. No further information was available.
Report: 1222780-2026-00157
One report describes a surgical probe that gave unreliable readings, jumping between low and high numbers when placed near a marker, so the surgeon said it was not reading correctly. Switching to a different device resolved the problem.
Report: 22152472
One report from the UK says metal surgical forceps caused a false signal in the detector used to find a seed during lymph node removal, meaning the detector reacted to the forceps. Plastic retractors were used instead. The report says there was no patient harm.
Report: 3020733550-2023-00001
One report says a reflector tumor marker failed after it was removed from a patient. A company representative said the reflector system is deactivated when touched by a bovie, an electrosurgical cutting tool. The manufacturer wants to examine the device.
Report: MW5073437
Written by AI from the narratives of the 6 newest reports that have one, received Nov 20, 2017 to Apr 8, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 18in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 6registered establishments that list this product code
Other codes under 21 CFR 878.4300
Product codes classified under the same regulation.
Watch PBY in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PBY every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PBY. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PBYRetrieved Nov 29, 2026, 12:52 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PBYRetrieved Nov 29, 2026, 4:24 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PBY · count decision_dateRetrieved Nov 29, 2026, 12:52 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PBYRetrieved Nov 29, 2026, 4:24 PM UTC.
- device/recall · search product_code:PBY · count event_date_initiatedRetrieved Nov 29, 2026, 12:52 AM UTC.
- device/recall · search product_code:PBY · count res_event_numberRetrieved Nov 29, 2026, 12:52 AM UTC.
- device/event · search device.device_report_product_code:PBY AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:24 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PBY AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:24 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PBY AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 12:52 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PBYRetrieved Nov 29, 2026, 12:52 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PBYRetrieved Nov 29, 2026, 12:52 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:878.4300Retrieved Nov 28, 2026, 11:00 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil