FDA product code
PBQ: Fixation, Non-Absorbable Or Absorbable, For Pelvic Use (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 884.4530
- Review panel
- Obstetrics and Gynecology
- Premarket route
- 510(k)
- Implant
- Yes
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 81 in the last 12 months
- Median review time
- 126 daysthe latest 8 decisions
- Recall records
- 10 in the last 12 months
- Adverse event reports
- 7last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code PBQ per year, by decision date, 8 since 2004. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260380 | MenditEscala Medical, Ltd. | Mar 4, 202627 days |
| K230730 | ApyxEscala Medical | Jul 25, 2023131 days |
| K220420 | Saffron Fixation SystemColoplast A/S | Jun 10, 2022116 days |
| K213783 | ApyxEscala Medical | Apr 5, 2022123 days |
| K160569 | NeuGuidePop Medical Solutions | Jul 6, 2016128 days |
| K130059 | ENDOFAST RELIANT SCP, ENDOFAST RELIANT LAPIbi Israel Biomedical Innovations , Ltd. | Jul 29, 2013200 days |
| K120831 | ANCHORSURENeomedic International S.L. | Oct 12, 2012207 days |
| K042603 | GYNECARE PROLENE FASTENER SYSTEMETHICON, Inc. | Dec 22, 200489 days |
Recalls
Recall records for product code PBQ per year, by the date the recall began. Each recalled product is one record; these 1 records belong to 1 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0556-2022Class II | FEMSelect LtdCertain lots may not have been packaged wth the Channel Tube Limiter. | Dec 13, 2021 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PBQ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 34 |
| 2024 | 16 |
| 2025 | 5 |
| 2026 (so far) | 7 |
Last 12 months by event type
- Deaths0
- Injuries4
- Malfunctions2
- Other or not stated1
Most reported device problems, last 12 months
- Device Dislodged or Dislocated2
- Improper or Incorrect Procedure or Method1
- No Apparent Adverse Event1
- Positioning Problem1
In plain English: what the newest reports describe Written by AI
In several reports, anchors (small devices that hold the sutures in the ligament) pulled out or came loose while the surgeon tightened the sutures. In one report, four of six anchors failed and the vaginal top kept falling, and a pulley method may have contributed.
5 reports: 2125050-2024-01151, 2125050-2024-01152, 2125050-2024-01611, 2125050-2025-01538, 2125050-2025-01290
Surgeons said they could not reliably hear the click that confirms an anchor is seated in the insertion tool. In one report, an anchor would not deploy and became stuck in the tool, and in another, an anchor placed earlier was dislodged during tie-down.
2 reports: 2125050-2025-01290, 2125050-2024-01611
A surgeon reportedly preferred a different fixation system because its anchors can wrap around the ligament, which made the hold easier to feel confident about. One report says the surgeon had several of these anchors fail and ended up performing a sacrocolpopexy, a different surgery to support the vaginal top, and another says pull-out strength was a concern.
2 reports: 2125050-2026-00774, 2125050-2024-01264
One report describes a patient with severe pain on both sides after two anchors were placed. A review found the right anchor in an unexpected position and the left anchor passed through the ligament, deeper and closer to the skin than intended, and the patient underwent an examination under anesthesia.
Report: MW6000264
One report says the patient believes the device has migrated and the prolapse surgery failed, and that a new gynecologist recommended revision surgery and a full hysterectomy. The patient also reports urine leakage, painful sex, urinary infections, and pelvic and back pain.
Report: MW5189081
Written by AI from the narratives of the 15 newest reports that have one, received Jul 29, 2024 to Sep 28, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 37in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 9registered establishments that list this product code
Other codes under 21 CFR 884.4530
Product codes classified under the same regulation.
Watch PBQ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PBQ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PBQ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PBQRetrieved Nov 29, 2026, 12:52 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PBQRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PBQ · count decision_dateRetrieved Nov 29, 2026, 12:52 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PBQRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PBQ · count event_date_initiatedRetrieved Nov 29, 2026, 12:52 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PBQ · count res_event_numberRetrieved Nov 29, 2026, 12:52 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0556-2022")Retrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PBQ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PBQ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PBQ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 12:52 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PBQRetrieved Nov 29, 2026, 12:52 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PBQRetrieved Nov 29, 2026, 12:52 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:884.4530Retrieved Nov 29, 2026, 4:12 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil