Product codes

FDA product code

HFX: Clamp, Circumcision (FDA product code)

Follow HFX: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 884.4530
Review panel
Obstetrics and Gynecology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
271 in the last 12 months
Median review time
148 daysthe latest 25 decisions
Recall records
140 in the last 12 months
Adverse event reports
112last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code HFX per year, by decision date, 27 since 1988. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20221
20231
20240
20250
2026 (so far)1

Latest decisions

NumberDevice and applicantDecision and review time
K251687 Konig Bell Circumcision ClampMedline Industries, LP Feb 25, 2026268 days
K221356 Wee BellWee Medical Feb 2, 2023268 days
K212911 Konig Mogen ClampMedline Industries, Inc. Nov 18, 2022431 days
K151095 Circumplast Circumcision DeviceNovadien Healthcare Sep 18, 2015148 days
K142163 MALE CIRCUMCISION KITS:RAPIDECLAMPMedicfit Technology Sdn Bhd Apr 20, 2015257 days
K131373 SHANGRINGWuhu Snnda Medical Treatment Appliance Technology Jun 25, 201343 days
K121789 SHANGRINGWuhu Snna Medical Treatment Appliance Technology C Aug 3, 201245 days
K103695 PREPEX SYSTEMCirc Medtech Jan 10, 2012389 days
K100916 MEDICON MOGEN CLAMPMedicon EG Dec 6, 2010248 days
K061539 GOMCO CIRCUMCISION CLAMP, MOGEN CLAMP, PLASTIBELL CIRCUMCISION DEVICEClinical Innovations, LLC Sep 20, 2007472 days

Recalls

Recall records for product code HFX per year, by the date the recall began. Each recalled product is one record; these 14 records belong to 10 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20192
20200
20210
20220
20232
20241
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0133-2025Class II MEDLINE INDUSTRIES, LP - NorthfieldProduct has been identified as having holes in the packaging. Sep 3, 2024
Z-2649-2023Class II MEDLINE INDUSTRIES, LP - NorthfieldThe kits incorrectly contain CirClamp subassembly 345CRSAK which results in the kits containing an incorrect bell and base size. May 23, 2023
Z-1407-2023Class II MEDLINE INDUSTRIES, LP - NorthfieldThe CirClamp subassembly found in the kit was the incorrect size. Mar 2, 2023
Z-0120-2020Class II Centurion Medical Products CorporationA mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel. Aug 28, 2019
Z-0121-2020Class II Centurion Medical Products CorporationA mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel. Aug 28, 2019

Adverse event reports (MAUDE)

Medical device reports received per year that name product code HFX. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201725
201813
201911
202022
20215
20223
202333
2024121
2025122
2026 (so far)105

Last 12 months by event type

  • Deaths0
  • Injuries38
  • Malfunctions73
  • Other or not stated1

Most reported device problems, last 12 months

  1. Insufficient Device Problem Information19
  2. Mechanical Problem18
  3. Sharp Edges12
  4. Nonstandard Device9
  5. Difficult to Open or Close8
  6. Appropriate Device Problem Term/Code Not Available5

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
130in the FDA’s Global Unique Device Identification Database
Establishment listings
55registered establishments that list this product code

Watch HFX in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get HFX every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HFX. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.