FDA product code
HFX: Clamp, Circumcision (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 884.4530
- Review panel
- Obstetrics and Gynecology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 271 in the last 12 months
- Median review time
- 148 daysthe latest 25 decisions
- Recall records
- 140 in the last 12 months
- Adverse event reports
- 112last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code HFX per year, by decision date, 27 since 1988. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K251687 | Konig Bell Circumcision ClampMedline Industries, LP | Feb 25, 2026268 days |
| K221356 | Wee BellWee Medical | Feb 2, 2023268 days |
| K212911 | Konig Mogen ClampMedline Industries, Inc. | Nov 18, 2022431 days |
| K151095 | Circumplast Circumcision DeviceNovadien Healthcare | Sep 18, 2015148 days |
| K142163 | MALE CIRCUMCISION KITS:RAPIDECLAMPMedicfit Technology Sdn Bhd | Apr 20, 2015257 days |
| K131373 | SHANGRINGWuhu Snnda Medical Treatment Appliance Technology | Jun 25, 201343 days |
| K121789 | SHANGRINGWuhu Snna Medical Treatment Appliance Technology C | Aug 3, 201245 days |
| K103695 | PREPEX SYSTEMCirc Medtech | Jan 10, 2012389 days |
| K100916 | MEDICON MOGEN CLAMPMedicon EG | Dec 6, 2010248 days |
| K061539 | GOMCO CIRCUMCISION CLAMP, MOGEN CLAMP, PLASTIBELL CIRCUMCISION DEVICEClinical Innovations, LLC | Sep 20, 2007472 days |
Recalls
Recall records for product code HFX per year, by the date the recall began. Each recalled product is one record; these 14 records belong to 10 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 2 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 2 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0133-2025Class II | MEDLINE INDUSTRIES, LP - NorthfieldProduct has been identified as having holes in the packaging. | Sep 3, 2024 |
| Z-2649-2023Class II | MEDLINE INDUSTRIES, LP - NorthfieldThe kits incorrectly contain CirClamp subassembly 345CRSAK which results in the kits containing an incorrect bell and base size. | May 23, 2023 |
| Z-1407-2023Class II | MEDLINE INDUSTRIES, LP - NorthfieldThe CirClamp subassembly found in the kit was the incorrect size. | Mar 2, 2023 |
| Z-0120-2020Class II | Centurion Medical Products CorporationA mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel. | Aug 28, 2019 |
| Z-0121-2020Class II | Centurion Medical Products CorporationA mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel. | Aug 28, 2019 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code HFX. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 25 |
| 2018 | 13 |
| 2019 | 11 |
| 2020 | 22 |
| 2021 | 5 |
| 2022 | 3 |
| 2023 | 33 |
| 2024 | 121 |
| 2025 | 122 |
| 2026 (so far) | 105 |
Last 12 months by event type
- Deaths0
- Injuries38
- Malfunctions73
- Other or not stated1
Most reported device problems, last 12 months
- Insufficient Device Problem Information19
- Mechanical Problem18
- Sharp Edges12
- Nonstandard Device9
- Difficult to Open or Close8
- Appropriate Device Problem Term/Code Not Available5
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 130in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 55registered establishments that list this product code
Other codes under 21 CFR 884.4530
Product codes classified under the same regulation.
Watch HFX in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get HFX every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HFX. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:HFXRetrieved Nov 28, 2026, 8:42 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HFXRetrieved Nov 29, 2026, 4:15 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HFX · count decision_dateRetrieved Nov 28, 2026, 8:42 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:HFXRetrieved Nov 29, 2026, 4:15 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HFX · count event_date_initiatedRetrieved Nov 28, 2026, 8:42 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HFX · count res_event_numberRetrieved Nov 28, 2026, 8:42 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0133-2025" "Z-2649-2023" "Z-1407-2023" "Z-0120-2020" "Z-0121-2020")Retrieved Nov 29, 2026, 4:15 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HFX AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:15 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HFX AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:15 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HFX AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:15 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:HFXRetrieved Nov 28, 2026, 8:42 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:HFXRetrieved Nov 28, 2026, 8:42 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:884.4530Retrieved Nov 29, 2026, 4:12 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil