Product codes

FDA product code

OBJ: Catheter, Ultrasound, Intravascular (FDA product code)

Follow OBJ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 870.1200
Review panel
Cardiovascular
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
804 in the last 12 months
Median review time
63 daysthe latest 25 decisions
Recall records
231 in the last 12 months
Adverse event reports
1,230last 12 months, 14 deaths

510(k) and De Novo decisions

Decisions for product code OBJ per year, by decision date, 80 since 2003. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20176
20182
20191
20204
20213
20223
20233
20243
20257
2026 (so far)4

Latest decisions

NumberDevice and applicantDecision and review time
K261696 ACUNAV Luxera 4D ICE CatheterSiemens Medical Solutions USA, Inc. Sep 10, 2026111 days
K261849 VERAFEYE Imaging Catheter (VF-VIC-002)Luma Vision Limited Jul 24, 202651 days
K252945 Novasight Hybrid SystemConavi Medical, Inc. Apr 17, 2026214 days
K260816 OptiCross Coronary Imaging Catheter (H749518110); OptiCross 6 Coronary Imaging Catheter (H7495181160); OptiCross HD Coronary Imaging Catheter (H74939352040); OptiCross 6 HD Coronary Imaging Catheter (H74939354080); OptiCross Coronary Imaging Catheter (Bagless) (H749518130); OptiCross 6 Coronary Imaging Catheter (Bagless) (H7495181360); OptiCross HD Coronary Imaging Catheter (Bagless) (H74939352050); OptiCross 6 HD Coronary Imaging Catheter (Bagless) (H74939354090)Boston Scientific Apr 8, 202626 days
K253399 Visions® PV .014P RX Digital IVUS Catheter; Visions® PV .018 Digital IVUS CatheterVolcano Corporation (Dba Philips Image Guided Therapy Device Oct 28, 202528 days
K250955 XC11 ICE System, USAYorlabs, Inc. Oct 22, 2025205 days
K250913 TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound SystemJiangsu Tingsn Technology Co., Ltd. Jul 19, 2025114 days
K251211 ViewFlex™ Xtra ICE Catheter; ViewFlex™ Eco Reprocessed ICE Catheter; Advisor™ HD Grid Mapping Catheter, Sensor Enabled™; Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™; Agilis™ NxT Steerable Introducer; Agilis™ NxT Steerable Introducer Dual-Reach™ABBOTT MEDICAL May 23, 202535 days
K251231 ViewFlex™ X ICE Catheter, Sensor Enabled™ABBOTT MEDICAL May 20, 202528 days
K251103 VeriSight Intracardiac Echocardiography Catheter (VSICE2D); VeriSight Pro Intracardiac Echocardiography Catheter (VSICE3D)Philips Image Guided Therapy Corporation May 9, 202528 days

Recalls

Recall records for product code OBJ per year, by the date the recall began. Each recalled product is one record; these 23 records belong to 20 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20175
20183
20190
20202
20211
20222
20230
20241
20252
2026 (so far)1

Latest recalls

Recall and classFirm and reasonBegan
Z-2379-2026Class I ARROW INTERNATIONAL, LLCLidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality is… Apr 21, 2026
Z-2221-2025Class II Volcano CorpReports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide cathet… Jun 13, 2025
Z-1585-2025Class I Conavi Medical Inc.Due to manufacturing issues there is a potential for the catheter sheath to detach. Mar 12, 2025
Z-1396-2024Class II Boston Scientific CorporationCertain batches of Opticross 40MHz Coronary Imaging Catheters manufactured using an incorrect voltage setting, which may have resulted in decreased image brigh… Mar 5, 2024
Z-1301-2022Class III Volcano CorpDigital IVUS catheter instructions for use ( IFU) do not contain contrainidications May 13, 2022

Adverse event reports (MAUDE)

Medical device reports received per year that name product code OBJ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017524
2018542
2019237
2020288
2021353
2022473
2023697
20241,154
2025797
2026 (so far)1,070

Last 12 months by event type

  • Deaths14
  • Injuries120
  • Malfunctions1,091
  • Other or not stated5

Most reported device problems, last 12 months

  1. Air/Gas in Device490
  2. No Display/Image318
  3. Nonstandard Device227
  4. Difficult to Flush191
  5. Entrapment of Device185
  6. Poor Quality Image140

In plain English: what the newest reports describe Written by AI

  • Several reports describe air that could not be removed from the catheter during preparation, or air getting in while flushing. Flushing is the step of pushing saline through the device to clear air, and it did not clear the air in these cases.

    6 reports: 2124215-2026-57264, 2124215-2026-57296, 2124215-2026-55556, 2124215-2026-55483, 2124215-2026-55570, 2124215-2026-57525

  • In several reports the imaging catheter's picture went black, flickered, grey or disappeared during use. In some of these, device analysis found the catheter twisted, with the imaging window ripped or rippled.

    5 reports: 2124215-2026-56060, 2124215-2026-55954, 2124215-2026-55955, 2124215-2026-57525, MW6004389

  • In a few reports the catheter could not be used or had to be replaced before the procedure. One catheter would not turn on and the procedure was cancelled.

    2 reports: 2124215-2026-56577, 2124215-2026-56319

  • In one report, a catheter was kinked about 15 mm from its tip when removed. In another, the catheter got stuck in a blockage and had to be cut and retrieved with a guide extension catheter.

    2 reports: 2124215-2026-57507, 2124215-2026-57140

  • In one report the tip of the catheter detached inside the patient and was removed with a snare, a loop device used to grab objects in blood vessels. In another, a catheter snagged on a guidewire and split in half, and the far end was retrieved with a snare.

    2 reports: 2124215-2026-57123, 3008363989-2026-00168

Written by AI from the narratives of the 19 newest reports that have one, received Oct 20, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
226in the FDA’s Global Unique Device Identification Database
Establishment listings
382registered establishments that list this product code

Watch OBJ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get OBJ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OBJ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.