FDA product code
OBJ: Catheter, Ultrasound, Intravascular (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 870.1200
- Review panel
- Cardiovascular
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 804 in the last 12 months
- Median review time
- 63 daysthe latest 25 decisions
- Recall records
- 231 in the last 12 months
- Adverse event reports
- 1,230last 12 months, 14 deaths
510(k) and De Novo decisions
Decisions for product code OBJ per year, by decision date, 80 since 2003. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 6 |
| 2018 | 2 |
| 2019 | 1 |
| 2020 | 4 |
| 2021 | 3 |
| 2022 | 3 |
| 2023 | 3 |
| 2024 | 3 |
| 2025 | 7 |
| 2026 (so far) | 4 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261696 | ACUNAV Luxera 4D ICE CatheterSiemens Medical Solutions USA, Inc. | Sep 10, 2026111 days |
| K261849 | VERAFEYE Imaging Catheter (VF-VIC-002)Luma Vision Limited | Jul 24, 202651 days |
| K252945 | Novasight Hybrid SystemConavi Medical, Inc. | Apr 17, 2026214 days |
| K260816 | OptiCross Coronary Imaging Catheter (H749518110); OptiCross 6 Coronary Imaging Catheter (H7495181160); OptiCross HD Coronary Imaging Catheter (H74939352040); OptiCross 6 HD Coronary Imaging Catheter (H74939354080); OptiCross Coronary Imaging Catheter (Bagless) (H749518130); OptiCross 6 Coronary Imaging Catheter (Bagless) (H7495181360); OptiCross HD Coronary Imaging Catheter (Bagless) (H74939352050); OptiCross 6 HD Coronary Imaging Catheter (Bagless) (H74939354090)Boston Scientific | Apr 8, 202626 days |
| K253399 | Visions® PV .014P RX Digital IVUS Catheter; Visions® PV .018 Digital IVUS CatheterVolcano Corporation (Dba Philips Image Guided Therapy Device | Oct 28, 202528 days |
| K250955 | XC11 ICE System, USAYorlabs, Inc. | Oct 22, 2025205 days |
| K250913 | TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound SystemJiangsu Tingsn Technology Co., Ltd. | Jul 19, 2025114 days |
| K251211 | ViewFlex™ Xtra ICE Catheter; ViewFlex™ Eco Reprocessed ICE Catheter; Advisor™ HD Grid Mapping Catheter, Sensor Enabled™; Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™; Agilis™ NxT Steerable Introducer; Agilis™ NxT Steerable Introducer Dual-Reach™ABBOTT MEDICAL | May 23, 202535 days |
| K251231 | ViewFlex™ X ICE Catheter, Sensor Enabled™ABBOTT MEDICAL | May 20, 202528 days |
| K251103 | VeriSight Intracardiac Echocardiography Catheter (VSICE2D); VeriSight Pro Intracardiac Echocardiography Catheter (VSICE3D)Philips Image Guided Therapy Corporation | May 9, 202528 days |
Recalls
Recall records for product code OBJ per year, by the date the recall began. Each recalled product is one record; these 23 records belong to 20 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 5 |
| 2018 | 3 |
| 2019 | 0 |
| 2020 | 2 |
| 2021 | 1 |
| 2022 | 2 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 2 |
| 2026 (so far) | 1 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2379-2026Class I | ARROW INTERNATIONAL, LLCLidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality is… | Apr 21, 2026 |
| Z-2221-2025Class II | Volcano CorpReports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide cathet… | Jun 13, 2025 |
| Z-1585-2025Class I | Conavi Medical Inc.Due to manufacturing issues there is a potential for the catheter sheath to detach. | Mar 12, 2025 |
| Z-1396-2024Class II | Boston Scientific CorporationCertain batches of Opticross 40MHz Coronary Imaging Catheters manufactured using an incorrect voltage setting, which may have resulted in decreased image brigh… | Mar 5, 2024 |
| Z-1301-2022Class III | Volcano CorpDigital IVUS catheter instructions for use ( IFU) do not contain contrainidications | May 13, 2022 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code OBJ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 524 |
| 2018 | 542 |
| 2019 | 237 |
| 2020 | 288 |
| 2021 | 353 |
| 2022 | 473 |
| 2023 | 697 |
| 2024 | 1,154 |
| 2025 | 797 |
| 2026 (so far) | 1,070 |
Last 12 months by event type
- Deaths14
- Injuries120
- Malfunctions1,091
- Other or not stated5
Most reported device problems, last 12 months
- Air/Gas in Device490
- No Display/Image318
- Nonstandard Device227
- Difficult to Flush191
- Entrapment of Device185
- Poor Quality Image140
In plain English: what the newest reports describe Written by AI
Several reports describe air that could not be removed from the catheter during preparation, or air getting in while flushing. Flushing is the step of pushing saline through the device to clear air, and it did not clear the air in these cases.
6 reports: 2124215-2026-57264, 2124215-2026-57296, 2124215-2026-55556, 2124215-2026-55483, 2124215-2026-55570, 2124215-2026-57525
In several reports the imaging catheter's picture went black, flickered, grey or disappeared during use. In some of these, device analysis found the catheter twisted, with the imaging window ripped or rippled.
5 reports: 2124215-2026-56060, 2124215-2026-55954, 2124215-2026-55955, 2124215-2026-57525, MW6004389
In a few reports the catheter could not be used or had to be replaced before the procedure. One catheter would not turn on and the procedure was cancelled.
2 reports: 2124215-2026-56577, 2124215-2026-56319
In one report, a catheter was kinked about 15 mm from its tip when removed. In another, the catheter got stuck in a blockage and had to be cut and retrieved with a guide extension catheter.
2 reports: 2124215-2026-57507, 2124215-2026-57140
In one report the tip of the catheter detached inside the patient and was removed with a snare, a loop device used to grab objects in blood vessels. In another, a catheter snagged on a guidewire and split in half, and the far end was retrieved with a snare.
2 reports: 2124215-2026-57123, 3008363989-2026-00168
Written by AI from the narratives of the 19 newest reports that have one, received Oct 20, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 226in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 382registered establishments that list this product code
Other codes under 21 CFR 870.1200
Product codes classified under the same regulation.
Watch OBJ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get OBJ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OBJ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:OBJRetrieved Nov 28, 2026, 7:18 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OBJRetrieved Nov 29, 2026, 7:47 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OBJ · count decision_dateRetrieved Nov 28, 2026, 7:18 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:OBJRetrieved Nov 29, 2026, 7:47 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OBJ · count event_date_initiatedRetrieved Nov 28, 2026, 7:18 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OBJ · count res_event_numberRetrieved Nov 28, 2026, 7:18 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2379-2026" "Z-2221-2025" "Z-1585-2025" "Z-1396-2024" "Z-1301-2022")Retrieved Nov 29, 2026, 7:47 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OBJ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:47 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OBJ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:47 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OBJ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:47 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:OBJRetrieved Nov 28, 2026, 7:18 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:OBJRetrieved Nov 28, 2026, 7:18 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:870.1200Retrieved Nov 28, 2026, 7:18 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.