Product codes

FDA product code

OQY: Intrauterine Tamponade Balloon (FDA product code)

Follow OQY: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 884.4530
Review panel
Obstetrics and Gynecology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
510(k) and De Novo decisions
111 in the last 12 months
Median review time
116 daysthe latest 11 decisions
Recall records
20 in the last 12 months
Adverse event reports
95last 12 months, 3 deaths

510(k) and De Novo decisions

Decisions for product code OQY per year, by decision date, 11 since 2010. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20171
20180
20191
20201
20211
20221
20231
20240
20250
2026 (so far)1

Latest decisions

NumberDevice and applicantDecision and review time
K253642 Jada System (Jada-2002)Organon, LLC May 15, 2026177 days
K230849 ELLAVI UBTSinapi Biomedical (Pty) , Ltd. Nov 13, 2023230 days
K223098 Bakri Essential Postpartum BalloonCook Incorporated Oct 28, 202228 days
K212757 Jada SystemAlydia Health Sep 30, 202130 days
K201199 Jada SystemAlydia Health Aug 28, 2020116 days
K191264 Every Second Matters-Uterine Balloon Tamponade (“ESM-UBT”)Ujenzi Charitable Trust Oct 30, 2019173 days
K170622 Bakri Postpartum Balloon, Bakri Postpartum Balloon with Rapid Instillation ComponentCook Incorporated Mar 31, 201730 days
K153591 ebb Complete Tamponade SystemClinical Innovations, LLC Jan 14, 201629 days
K150573 ebb Complete Tamponade SystemClinical Innovations, LLC Mar 25, 201519 days
K101535 BT-CATHUtah Medical Products, Inc. Jan 11, 2011222 days

Recalls

Recall records for product code OQY per year, by the date the recall began. Each recalled product is one record; these 2 records belong to 2 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0123-2016Class II Clinical Innovations, LLCClinical Innovations, LLC is voluntarily recalling the Ebb Complete Tamponade System (CTS-1000) Lot Number 1214-F-403.) to address the potential for a balloon… Jan 5, 2015
Z-1014-2015Class II Clinical Innovations, LLCPotential for a balloon leak in all lots of (CTS-1000) manufactured since August 2014. Dec 30, 2014

Adverse event reports (MAUDE)

Medical device reports received per year that name product code OQY. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20173
201855
201990
202090
2021113
2022274
2023228
2024215
2025180
2026 (so far)85

Last 12 months by event type

  • Deaths3
  • Injuries22
  • Malfunctions70
  • Other or not stated0

Most reported device problems, last 12 months

  1. Fluid/Blood Leak53
  2. Material Rupture10
  3. Adverse Event Without Identified Device or Use Problem8
  4. Component Missing7
  5. Inflation Problem4
  6. Improper or Incorrect Procedure or Method3

In plain English: what the newest reports describe Written by AI

  • Balloons leaked fluid after being filled, so the device could not stop the bleeding and a new balloon was placed. One report found pinhole defects on the balloon's surface, and another found a small hole.

    6 reports: 1820334-2026-01097, 1820334-2026-01081, 1820334-2026-00822, 1820334-2026-00812, 1820334-2026-00740, 1820334-2026-00733

  • Two reports describe leaks at the inflation valve or bulb, the part used to fill the balloon. The balloon could not be inflated properly, and a new device was used.

    2 reports: 1820334-2026-01069, 1820334-2026-01011

  • Parts of the device were missing or broke. One report says a connector piece was absent, which delayed care while another device was opened, and another says a syringe tip broke off at the valve.

    2 reports: 1820334-2026-00882, 1820334-2026-00945

  • Two reports describe serious outcomes in cases where the device was used for postpartum bleeding. One describes a cervical tear found when the device was removed and a patient who died during admission, and the other describes a hysterectomy after a device placement attempt failed, with the patient recovering.

    2 reports: 3017448360-2026-00004, 3017448360-2026-00007

  • One report describes fluid appearing in the collection bag about an hour after surgery. The balloon was intact when removed and no defect was found on testing, but the reporter suggested a possible internal defect let fluid pass into the drainage tube.

    Report: 51023849

Written by AI from the narratives of the 17 newest reports that have one, received Jul 2, 2026 to Oct 16, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
18in the FDA’s Global Unique Device Identification Database
Establishment listings
15registered establishments that list this product code

Watch OQY in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get OQY every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OQY. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.