Product codes

FDA product code

ODR: Electrosurgical Patient Return Electrode (FDA product code)

Follow ODR: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 878.4400
Review panel
General and Plastic Surgery
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
30 in the last 12 months
Median review time
192 daysthe latest 3 decisions
Recall records
00 in the last 12 months
Adverse event reports
3last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code ODR per year, by decision date, 3 since 2008. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K160366 Electrosurgical disposable grounding padsTop-Rank Health Care Co., Ltd. Aug 19, 2016192 days
K121268 WANDY DISPERSVIE ELECTRODEBio-Med USA, Inc. Dec 20, 2012238 days
K082204 SERIES SY DISPOSABLE ELECTROSURGICAL PADVentlab Corp. Aug 13, 20088 days

Recalls

Recall records for product code ODR per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code ODR. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20171
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)3

Last 12 months by event type

  • Deaths0
  • Injuries2
  • Malfunctions1
  • Other or not stated0

Most reported device problems, last 12 months

  1. Energy Output Problem1

In plain English: what the newest reports describe Written by AI

  • Several reports describe patients getting burns where the return electrode pad (the flat sticky pad that carries the current back to the generator) touched the skin. In one 2026 report, the pad was loose with moisture trapped between the pad and skin, and the burn was about 8 by 4.5 inches; the patient had mild discomfort the next day and no signs of infection.

    7 reports: MW5189901, 1717344-2014-00173, 1717344-2013-00838, 1717344-2013-00781, 1717344-2013-00370, 1717344-2012-01209, 1717344-2012-00440

  • In a 2014 report, a patient moved during surgery and was left resting on the pad without anyone checking it. The burned tissue had to be cleaned from the wound and the patient stayed in intensive care.

    Report: 1717344-2014-00903

  • In a 2016 report, the unit did not warn that the patient pad was not properly attached. During a demonstration, it delivered an electrical charge to a pad that was still stuck to the top of the unit, which should not have happened.

    Report: 1216677-2016-00049

  • One report describes a problem with the electrode monitor, the safety feature that checks whether the pad is attached well. One 2026 report says the monitor was not working properly, and another report of a 2013 surgery says the pad was loose on the thigh while the device was in use, but the reports do not explain the cause.

    Report: MW5188279

  • A 2012 report describes the generator being unable to power on using its foot switch, which the manufacturer received for service. The report says no patient complications were associated with the event.

    Report: 2182208-2012-04034

Written by AI from the narratives of the 20 newest reports that have one, received Mar 28, 2012 to Sep 28, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
11in the FDA’s Global Unique Device Identification Database
Establishment listings
13registered establishments that list this product code

Watch ODR in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get ODR every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in ODR. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.