FDA product code
OAR: Injector, Vertebroplasty (Does Not Contain Cement) (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 888.4200
- Review panel
- Orthopedic
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 10 in the last 12 months
- Median review time
- 63 daysthe latest 1 decisions
- Recall records
- 260 in the last 12 months
- Adverse event reports
- 20last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code OAR per year, by decision date, 1 since 2005. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K051820 | PARALLAX EZFLOW CEMENT DELIVERY SYSTEMArthrocare Corp. | Sep 6, 200563 days |
Recalls
Recall records for product code OAR per year, by the date the recall began. Each recalled product is one record; these 26 records belong to 9 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 4 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 2 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 2 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0073-2025Class II | Merit Medical Systems, Inc.Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may result in exposure to pathogens/pyrogens which cou… | Aug 19, 2024 |
| Z-2260-2024Class II | Stryker CorporationExpired Products distributed to customers | Jun 3, 2024 |
| Z-2067-2020Class II | Merit Medical Systems, Inc.Vertebral hydraulic assemblies has incorrect expiration date printed on the labels.The product's shelf-life is two (2) years, but the expiration date printed o… | Feb 13, 2020 |
| Z-2068-2020Class II | Merit Medical Systems, Inc.Vertebral hydraulic assemblies has incorrect expiration date printed on the labels.The product's shelf-life is two (2) years, but the expiration date printed o… | Feb 13, 2020 |
| Z-0352-2018Class II | Medtronic Sofamor Danek USA IncMedtronic has discovered the directional arrow at the proximal end of the instrument may not correctly align with the cut out opening on the distal end of the… | Dec 6, 2017 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code OAR. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 22 |
| 2018 | 11 |
| 2019 | 14 |
| 2020 | 5 |
| 2021 | 13 |
| 2022 | 11 |
| 2023 | 14 |
| 2024 | 17 |
| 2025 | 12 |
| 2026 (so far) | 19 |
Last 12 months by event type
- Deaths0
- Injuries5
- Malfunctions15
- Other or not stated0
Most reported device problems, last 12 months
- Fracture15
- Break2
- Difficult to Remove1
- Failure to Deliver1
In plain English: what the newest reports describe Written by AI
The cannula or needle broke while being removed from the spine, and a piece was left in the bone or pedicle. In some cases the case was stopped and the piece was removed later, and in others the provider judged removal unsafe and left it in place. Reports say the piece was seen on X-ray or CT, and the patient or family was told.
9 reports: MW6001542, MW6001471, 3015967359-2026-00465, 3015967359-2026-00438, 3015967359-2026-00327, 3015967359-2026-00136, 24303673, 23777868, 3015967359-2026-00432
Fracture events are listed in manufacturer summary records. These summaries state no health consequences, but they give no details about how the fractures happened.
5 reports: 3015967359-2026-99357, 3015967359-2026-99204, 3015967359-2026-99035, 3015967359-2025-99270, 3015967359-2025-99275
A spring came loose from the cannula handle and caught around the needle shaft, so the device could not be removed. The device then broke, leaving about 4 cm of needle in the pedicle. One report says anesthesia was prolonged by 1 to 1.5 hours, the procedure could not be completed, and the patient had more pain and spinal collapse.
Report: 3015967359-2026-00432
One report says the blue handle broke away from the silver access cannula twice during a lumbar fusion procedure.
Report: 24673360
One report says the surgeon saw cement in the spinal canal on imaging during surgery. The reporter believes the cement gun was pushed with too much pressure, and the patient had canal compression and was monitored. The manufacturer's note says the product did not malfunction during the case.
Report: 3003272754-2026-00001
Written by AI from the narratives of the 16 newest reports that have one, received Oct 23, 2025 to Aug 26, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 185in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 72registered establishments that list this product code
Other codes under 21 CFR 888.4200
Product codes classified under the same regulation.
Watch OAR in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get OAR every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OAR. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:OARRetrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OARRetrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OAR · count decision_dateRetrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:OARRetrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OAR · count event_date_initiatedRetrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OAR · count res_event_numberRetrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0073-2025" "Z-2260-2024" "Z-2067-2020" "Z-2068-2020" "Z-0352-2018")Retrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OAR AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OAR AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OAR AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:OARRetrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:OARRetrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:888.4200Retrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.