Product codes

FDA product code

KIH: Dispenser, Cement (FDA product code)

Follow KIH: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class I
Regulation
21 CFR 888.4200
Review panel
Orthopedic
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
110 in the last 12 months
Median review time
34 daysthe latest 11 decisions
Recall records
50 in the last 12 months
Adverse event reports
28last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code KIH per year, by decision date, 11 since 1978. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K042591 SMARTMIX PRE-FILLED MIXING SYSTEMDePuy Orthopaedics, Inc. Oct 22, 200429 days
K943906 P.F.C. CEMENT PRESSURIZING SYSTEMJohnson & Johnson Professionals, Inc. Dec 21, 1994132 days
K897134 CMW CEMENT PRESSURISATION SYSTEMDentsply Intl. Mar 2, 199065 days
K851881 ORTHOPEADIC CEMENT DELIVERY SYSTEMDentsply Intl. Jun 13, 198545 days
K840046 PRESSURIZING/SCAVENGING ACRYLIC CEMENTPrecision Medical Devices, Inc. Feb 9, 198434 days
K833675 BRIGHAM TIBIAL CEMENT PRESSURIZERHowmedica Corp. Dec 12, 198355 days
K811317 HOWMEDICA FEMORAL PRESSURING SEALHowmedica Corp. May 29, 198117 days
K801388 OH-HARRISJohnson & Johnson Professionals, Inc. Jul 28, 198045 days
K790275 GUN, EXETER CEMENTHowmedica Corp. Feb 21, 197912 days
K781955 GUN, CEMENTDepuy, Inc. Dec 4, 197813 days

Recalls

Recall records for product code KIH per year, by the date the recall began. Each recalled product is one record; these 5 records belong to 5 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20191
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1081-2019Class II Zimmer Biomet, Inc.Lack of an adequate sterilization validation Feb 28, 2019
Z-2772-2015Class II Orthovita, Inc., dBA Stryker Orthobiologics.Two isolated instances of holes in the outer header bag component of the double barrier sterile packaging configuration Aug 20, 2015
Z-1744-2015Class II BIOMET FRANCE S.A.R.L.The pin which maintains the knob button, may disconnect and become lost. This may result in a delay of surgery to obtain a new Optigun and detect/retrieve the… May 14, 2015
Z-1198-2012 Stryker Instruments Div. of Stryker CorporationThe firm identified two complaints have been reported indicating that the product inside the package did not match with the label information. The product was… Jan 13, 2012
Z-3188-2011 Orthovita, Inc.Recalling firm has discovered through internal testing that the Cortoss(TM) Delivery Gun may abrade its packaging material under certain transportation and han… Mar 31, 2011

Adverse event reports (MAUDE)

Medical device reports received per year that name product code KIH. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201714
201815
201930
202023
202130
202226
202330
202419
202593
2026 (so far)24

Last 12 months by event type

  • Deaths0
  • Injuries1
  • Malfunctions27
  • Other or not stated0

Most reported device problems, last 12 months

  1. Break7
  2. Leak/Splash5
  3. Mechanical Jam4
  4. Device Contaminated During Manufacture or Shipping3
  5. Mechanical Problem3
  6. Delivered as Unsterile Product2

In plain English: what the newest reports describe Written by AI

  • The tip of a cement delivery tool detached and became stuck inside the screw head, in several reports. In these cases the tip was hard to pull out, and surgeons had to widen the wound, break off screw towers, or use specialized tools, adding surgery time. One report noted the tip was not visible on X-ray after removal.

    3 reports: 1824199-2026-00097, 1824199-2026-00091, 1824199-2026-00017

  • The tips of the cannula, a hollow tube used to inject cement, broke during insertion in several reports. In some cases a fragment stayed in the screw and had to be removed, which caused a surgical delay of about 30 minutes in one report.

    2 reports: 1526439-2026-00699, 1526439-2026-00275

  • Material fragmentation and disintegration were reported in summary records, with two events of fragmentation found before clinical use. The reports involve no patient involvement.

    2 reports: 3015967359-2026-99587, 3015967359-2026-99420

  • A hair was found inside a sterile pack in one report, and a foreign substance was found inside a sterile package in another. In both cases the packaging was reported as not broken or no patient was involved.

    2 reports: 3006946279-2026-00041, 3006946279-2026-00014

  • A funnel in a cement mixing system was reported broken in one report, with no patient harm. A cement gun handle spring was reported broken in another, with no further detail.

    2 reports: 1818910-2026-16513, 1818910-2026-06536

Written by AI from the narratives of the 18 newest reports that have one, received Feb 10, 2026 to Oct 27, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
237in the FDA’s Global Unique Device Identification Database
Establishment listings
131registered establishments that list this product code

Product codes classified under the same regulation.

Watch KIH in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get KIH every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in KIH. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.