Product codes

FDA product code

LTK: Test, Epithelial Ovarian Tumor-Associated Antigen (Ca125) (FDA product code)

Follow LTK: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 866.6010
Review panel
Immunology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
260 in the last 12 months
Median review time
163 daysthe latest 25 decisions
Recall records
80 in the last 12 months
Adverse event reports
7last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code LTK per year, by decision date, 26 since 1997. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20201
20210
20221
20231
20241
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K240479 Access OV MonitorBeckman Coulter, Inc. May 10, 202480 days
K213510 IMMULITE/IMMULITE 1000 OM-MA, IMMULITE 2000 OM-MASiemens Healthcare Diagnostics Products, Ltd. Sep 8, 2023675 days
K221355 VITROS Immuodiagnostic Products CA 125 II Reagent PackOrtho Clinical Diagnostics Dec 12, 2022216 days
K200199 ADVIA Centaur CA 125IISiemens Healthcare Diagnostics, Inc. Apr 6, 202070 days
K143534 Elecsys CA-125 II assayRoche Professional Diagnostics Aug 6, 2015237 days
K142895 LUMIPULSE G1200 System, LUMIPULSE G CA 125II Immunoreaction Cartridges, LUMIPULSE G CA 125II CalibratorsFujirebio Diagnostics,Inc. May 21, 2015230 days
K100321 DIMENSION VISTA LOCI CA 125 FLEX REAGENT CARTRIDGE,AND DIMENSIONS VISTA LOCI 6 CAL, MODEL#'S K6455, AND KC604Siemens Healthcare Diagnostics Apr 12, 2011432 days
K080469 VIDAS CA 15-3 ASSAYbioMerieux, Inc. Jun 22, 2009487 days
K080561 VIDAS CA 125 II ASSAYbioMerieux, Inc. Apr 10, 2009407 days
K072794 IMMULITE 2500 OM-MA, MODEL L5KOPSiemens Medical Solutions Diagnostics Nov 5, 200735 days

Recalls

Recall records for product code LTK per year, by the date the recall began. Each recalled product is one record; these 8 records belong to 6 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20183
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1486-2018Class II Siemens Healthcare Diagnostics, IncUpon dilution, some patient samples may exhibit over-recovery outside the representative data provided in the assay instructions for use. Feb 21, 2018
Z-1682-2018Class II Siemens Healthcare Diagnostics, Inc.Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista… Jan 2, 2018
Z-1683-2018Class II Siemens Healthcare Diagnostics, Inc.Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista… Jan 2, 2018
Z-0525-2014Class II Ortho-Clinical DiagnosticsOrtho Clinical Diagnostics issued a Product Correction Notification for VITROS Immunodiagnostic Products CA 125 II Reagent Pack and Calibrators due to a positi… Oct 9, 2013
Z-1628-2011 BioCheck IncProduct labels indicate that the products are for IVD use and they do not have an approved 510(k). Dec 22, 2010

Adverse event reports (MAUDE)

Medical device reports received per year that name product code LTK. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201715
20188
201915
20205
20212
20222
20236
20246
20254
2026 (so far)7

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions7
  • Other or not stated0

Most reported device problems, last 12 months

  1. Low Test Results4
  2. High Test Results2
  3. Incorrect, Inadequate or Imprecise Result or Readings1

In plain English: what the newest reports describe Written by AI

  • Some samples gave a CA 125 result that was much lower than expected, a blood marker used in ovarian cancer care. Repeat tests on the same sample gave much higher values, and the reports did not identify a cause.

    7 reports: 3002809144-2026-00079, 3002809144-2026-00069, 1823260-2025-04791, 1823260-2025-01673, 1823260-2023-00770, 3002809144-2023-00353, 3002809144-2023-00033

  • Some samples gave a CA 125 result higher than expected, and physicians questioned it. Repeat testing gave lower values, which the reporters considered correct, and corrected reports were issued in some cases.

    6 reports: 1219913-2026-00061, 1219913-2026-00058, 1219913-2025-00039, 1823260-2024-03493, 1823260-2024-02778, 1823260-2023-00860

  • Some reports name specific reagent lots (batches of test chemicals) linked to discrepant results. In one report, a bias between two lots was found during a lot change while quality control stayed in range.

    6 reports: 1219913-2025-00121, 1219913-2025-00039, 1219913-2026-00061, 1219913-2026-00058, 1219913-2024-00475, 1219913-2024-00381

  • In some reports, a sample tested undiluted gave a different result from the same sample after dilution (mixing it with fluid to lower its concentration). The diluted result was higher, and one report says the manufacturer judged the undiluted result correct.

    3 reports: 1219913-2024-00475, 1219913-2024-00381, 1823260-2024-03771

  • A report describes a 69-year-old patient with no prior cancer diagnosis who had scans and procedures, including biopsies and a colonoscopy, after falsely elevated CA 125 results. No cancer was found during that testing.

    Report: 3002809144-2024-00127

Written by AI from the narratives of the 19 newest reports that have one, received Jan 15, 2023 to Mar 31, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
41in the FDA’s Global Unique Device Identification Database
Establishment listings
44registered establishments that list this product code

Watch LTK in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get LTK every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LTK. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.