Product codes

FDA product code

MUM: Stent, Metallic, Expandable, Duodenal (FDA product code)

Follow MUM: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 878.3610
Review panel
Gastroenterology and Urology
Premarket route
510(k)
Implant
Yes
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
510(k) and De Novo decisions
60 in the last 12 months
Median review time
112 daysthe latest 6 decisions
Recall records
00 in the last 12 months
Adverse event reports
19last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code MUM per year, by decision date, 6 since 1998. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20171
20180
20190
20200
20210
20220
20231
20240
20251
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K250663 Niti-S Duodenal Comfort Stent, Niti-S Colonic Comfort StentTaewoong Medical Co., Ltd. Mar 31, 202526 days
K223067 Niti-S Duodenal Stent; Niti-S Colonic Comfort StentTaewoong Medical Co., Ltd. Jun 14, 2023264 days
K163468 Evolution Duodenal Stent System - Uncovered, Evolution Colonic Stent System - UncoveredCook Ireland, Ltd. May 4, 2017143 days
K101530 EVOLUTION DUODENAL STENT SYSTEM, MODEL EVO-22-27-6-D, EVO-22-27-9-D, EVP-22-27-12-DCook Ireland, Ltd. Mar 29, 2011299 days
K062750 WALLFLEX ENTERAL DUODENAL STENT WITH ANCHOR LOCK DELIVERY SYSTEM, MODELS M00565010, M00565020, M00565030Boston Scientific Corp Dec 4, 200681 days
K980113 WALLSTENT ENTERAL ENDOPROSTHESISBoston Scientific Scimed, Inc. Apr 3, 199880 days

Recalls

Recall records for product code MUM per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code MUM. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201785
201849
201935
202032
202170
202228
202353
202434
202532
2026 (so far)15

Last 12 months by event type

  • Deaths0
  • Injuries1
  • Malfunctions18
  • Other or not stated0

Most reported device problems, last 12 months

  1. Break10
  2. Fracture6
  3. Activation, Positioning or Separation Problem3
  4. Activation Failure2
  5. Material Deformation2
  6. Positioning Problem2

In plain English: what the newest reports describe Written by AI

  • Seven reports describe a stent that broke. Breaks were found after placement, often in the middle or the end of the stent, and in one case a broken piece was pulled from the body. In some cases another stent was then placed, and one report described a stent that moved into the small intestine.

    7 reports: 3003902943-2026-00029, 3003902943-2026-00009, 3003902943-2026-00008, 3003902943-2025-00024, 3003902943-2025-00022, 3008146331-2025-00007, 3001845648-2026-00166

  • In four reports, the stent could not be released smoothly from its delivery tool. In one, it stuck and was forced out, and in others the stent could not be deployed at all, so the procedure was finished with another stent.

    4 reports: 3005099803-2026-03065, 3001845648-2025-00653, 3001845648-2025-00626, 3001845648-2025-00563

  • In one report, a stent came off its delivery catheter before the doctor confirmed it was in the right place, and it had to be partly pulled back. Another report says a stent fell off early and was not placed accurately, so the treatment did not work.

    2 reports: 3001845648-2025-00607, 3005099803-2026-02912

  • Two reports describe stents with rough surfaces or damage. One found small burrs on the end of a new stent, and the other found a stent that was breaking apart with its internal wires showing, so the doctor placed two more stents below it.

    2 reports: 3005099803-2026-01676, 3005099803-2025-05130

  • In two reports, the stent did not fit through the endoscope channel or could not pass a narrowed area. One stent was deployed inside the scope channel, and in the other the tip of the delivery catheter was found kinked.

    2 reports: 3005099803-2026-03834, 3001845648-2026-00282

Written by AI from the narratives of the 20 newest reports that have one, received Sep 17, 2025 to Oct 20, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
51in the FDA’s Global Unique Device Identification Database
Establishment listings
22registered establishments that list this product code

Product codes classified under the same regulation.

Watch MUM in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get MUM every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in MUM. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.