Product codes

FDA product code

MQR: Stent, Colonic, Metallic, Expandable (FDA product code)

Follow MQR: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 878.3610
Review panel
Gastroenterology and Urology
Premarket route
510(k)
Implant
Yes
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
160 in the last 12 months
Median review time
97 daysthe latest 16 decisions
Recall records
20 in the last 12 months
Adverse event reports
34last 12 months, 2 deaths

510(k) and De Novo decisions

Decisions for product code MQR per year, by decision date, 16 since 1996. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20181
20192
20202
20210
20220
20230
20241
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K234046 WallFlex Colonic Stent System with Anchor Lock Delivery System; WallFlex Duodenal Stent System with Anchor Lock Delivery System; WallFlex Colonic Soft Stent System with Anchor Lock Delivery System; WallFlex Duodenal Soft Stent System with Anchor Lock Delivery SystemBoston Scientific Corporation Mar 8, 202478 days
K201159 WallFlex Colonic Stent System with Anchor Lock Delivery SystemBoston Scientific Corporation Aug 17, 2020109 days
K200257 WallFlex Colonic Soft Stent System with Anchor Lock Delivery System, WallFlex Duodenal Soft Stent System with Anchor Lock Delivery SystemBoston Scientific Corporation Apr 17, 202074 days
K190141 HANAROSTENT LowAx Colon/Rectum (NNN)M.I.Tech Co., Ltd. Mar 7, 201937 days
K183616 HANAROSTENT LowAx Colon/Rectum (NNN), HANAROSTENT LowAx Duodenum/Pylorus (NNN)M.I.Tech Co., Ltd. Jan 10, 201915 days
K180180 HANAROSTENT LowAxTM Colon/Rectum (NNN), HANAROSTENT LowAxTM Duodenum/Pylorus (NNN)M.I.Tech Co., Ltd. Nov 2, 2018283 days
K113510 EVOLUTION COLONIC STENT SYSTEMCook Ireland, Ltd. May 17, 2012171 days
K061877 WALLFLEX ENTERAL COLONIC STENT WITH ANCHOR LOCK DELIVERY SYSTEMBoston Scientific Corp Sep 15, 200674 days
K060323 ULTRAFLEX PRECISION COLONIC STENT, MODELS M00557330, MO00557360, M00557380Boston Scientific Corporation May 12, 200692 days
K042065 WALLFLEX ENTERNAL COLONIC STENT AND ANCHOR LOCK DELIVERY SYSTEMBoston Scientific Corp Sep 30, 200459 days

Recalls

Recall records for product code MQR per year, by the date the recall began. Each recalled product is one record; these 2 records belong to 2 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20181
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-3038-2018Class II SPS Sterilization, IncThe expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for th… Mar 7, 2018
Z-2331-2010 Boston Scientific CorporationUnits have excessive fraying of the deployment suture thread, as a result, these devices may experience a suture break during deployment. Jul 15, 2010

Adverse event reports (MAUDE)

Medical device reports received per year that name product code MQR. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201749
201844
201942
202046
202156
202235
202338
202470
202529
2026 (so far)31

Last 12 months by event type

  • Deaths2
  • Injuries11
  • Malfunctions21
  • Other or not stated0

Most reported device problems, last 12 months

  1. Positioning Problem11
  2. Activation Failure9
  3. Adverse Event Without Identified Device or Use Problem9
  4. Break7
  5. Activation, Positioning or Separation Problem6
  6. Material Deformation5

In plain English: what the newest reports describe Written by AI

  • In several reports the stent would not release from its delivery system. It stayed inside the outer sheath, separated from the handle, or stayed attached to the inner catheter and moved out of place, so it was pulled out and usually replaced with a second stent.

    7 reports: 3005099803-2026-03657, 3001845648-2026-00225, 3001845648-2026-00123, 3001845648-2026-00375, 3005099803-2026-03885, 3005099803-2026-03148, 3005099803-2026-03150

  • In several reports the stent's wire was broken, split or bent out of shape. Examples include a broken wire at the tip, a partly fractured wire, and wires splayed apart on one side of the stent.

    4 reports: 3005099803-2026-04103, 3005099803-2026-02942, 3003902943-2026-00011, 3005099803-2026-03885

  • In a few reports the stent opened early, before the doctor intended, as the delivery system was being used. In one, the delivery system then broke and the stent was released in the wrong position, and in another the catheter broke during the procedure.

    3 reports: 3005099803-2026-03930, 3003902943-2026-00013, 3005099803-2026-01020

  • In two reports a hole (perforation) in the bowel was found near the stent, and in one the stent could not be removed so surgeons had to repair the bowel. The reports do not agree on whether the hole happened during placement or before it. One report also describes a patient who died, but the manufacturer says no link to the device has been established.

    3 reports: 3003902943-2026-00015, 3003902943-2026-00011, 3005099803-2026-02644

  • In two reports the stent or its sheath was kinked, and in one of them the doctor met strong resistance while moving the device to the narrowing. The doctor stopped using the device and a new stent completed the procedure.

    2 reports: 3001845648-2026-00172, 3005099803-2026-02288

Written by AI from the narratives of the 19 newest reports that have one, received Mar 19, 2026 to Oct 12, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
76in the FDA’s Global Unique Device Identification Database
Establishment listings
27registered establishments that list this product code

Product codes classified under the same regulation.

Watch MQR in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get MQR every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in MQR. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.