FDA product code
MLN: Electrode, Electrocardiograph, Multi-Function (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 870.2360
- Review panel
- Cardiovascular
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 00 in the last 12 months
- Median review time
- Not availablethe latest 0 decisions
- Recall records
- 90 in the last 12 months
- Adverse event reports
- 3last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code MLN per year, by decision date. Review time is the days from receipt to decision.
No decisions on record for this product code.
Recalls
Recall records for product code MLN per year, by the date the recall began. Each recalled product is one record; these 9 records belong to 6 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0856-2012 | ConMed CorporationThe specific lot was assembled without the rivet cover. | Mar 31, 2011 |
| Z-1371-2011 | Heart Sync, IncThe seal on the electrode pouch may be open, exposing the electrode to the outside environment. The package does have a caution statement that states, " Do not… | Dec 17, 2010 |
| Z-1372-2011 | Heart Sync, IncThe seal on the electrode pouch may be open, exposing the electrode to the outside environment. The package does have a caution statement that states, " Do not… | Dec 17, 2010 |
| Z-1950-2011 | Heart Sync, IncThe "leads out" connector had bent pins which would not allow connection to the Medtronic Physio Control Defibrillator cable during a product demonstration. | Dec 5, 2008 |
| Z-1951-2011 | Heart Sync, IncThe "leads out" connector had bent pins which would not allow connection to the Medtronic Physio Control Defibrillator cable during a product demonstration. | Dec 5, 2008 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code MLN. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 19 |
| 2018 | 14 |
| 2019 | 11 |
| 2020 | 1 |
| 2021 | 7 |
| 2022 | 8 |
| 2023 | 8 |
| 2024 | 1 |
| 2025 | 2 |
| 2026 (so far) | 3 |
Last 12 months by event type
- Deaths0
- Injuries2
- Malfunctions1
- Other or not stated0
Most reported device problems, last 12 months
- Connection Problem1
In plain English: what the newest reports describe Written by AI
During defibrillation shocks, pads sparked or arced, causing flames or burns on chest hair, skin, a beard, and in one case the face. One report says the patient died and that user error was suspected.
6 reports: MW5190953, MW5114788, 15958419, 0003015876-2023-00331, 13884544, 1417592-2023-00358
Several reports say the defibrillator could not deliver a shock, or delivered it late, because it would not hold a charge, its paddles peeled off, or it showed a pad-check message. One report describes a significant delay in therapy.
6 reports: 0003015876-2023-01536, 0003015876-2023-00291, 0003015876-2022-02741, 0003015876-2022-02721, 0003015876-2022-01613, 0003015876-2022-02096
Two reports describe defibrillator pads that became hot, melted, or puffed up after shocks, with a burning smell or smoke. In one, the defective pad was thrown away during resuscitation, so it could not be examined.
2 reports: MW5114911, 17450388
Two reports describe skin reactions under the adhesive stickers on heart monitoring electrodes, including bruising, rash, blistering, and burns. One report says a red warning light meant the monitor could not get a good reading from the skin.
2 reports: 20291695, MW5187940
A report says electrical arcing between the oxygen sensor cable and the ECG leads made a loud pop and a burning smell during cardioversion, though the patient's heart rhythm did not change.
Report: MW5166627
Written by AI from the narratives of the 18 newest reports that have one, received Mar 24, 2022 to Jul 22, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 97in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 23registered establishments that list this product code
Other codes under 21 CFR 870.2360
Product codes classified under the same regulation.
Watch MLN in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get MLN every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in MLN. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:MLNRetrieved Nov 29, 2026, 6:53 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MLNRetrieved Nov 29, 2026, 6:53 PM UTC.
- device/510k · search product_code:MLN · count decision_dateRetrieved Nov 29, 2026, 6:53 PM UTC.
- device/recall · search product_code:MLNRetrieved Nov 29, 2026, 6:53 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MLN · count event_date_initiatedRetrieved Nov 29, 2026, 6:53 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MLN · count res_event_numberRetrieved Nov 29, 2026, 6:53 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0856-2012" "Z-1371-2011" "Z-1372-2011" "Z-1950-2011" "Z-1951-2011")Retrieved Nov 29, 2026, 6:53 PM UTC.
- device/event · search device.device_report_product_code:MLN AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:53 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MLN AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:53 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MLN AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 6:53 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:MLNRetrieved Nov 29, 2026, 6:53 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:MLNRetrieved Nov 29, 2026, 6:53 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:870.2360Retrieved Nov 28, 2026, 8:40 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil