DRX: Electrode, Electrocardiograph (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 870.2360
- Review panel
- Cardiovascular
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 3290 in the last 12 months
- Median review time
- 93 daysthe latest 25 decisions
- Recall records
- 200 in the last 12 months
- Adverse event reports
- 396last 12 months, 1 death
510(k) and De Novo decisions
Decisions for product code DRX per year, by decision date, 329 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K171244 | PhysiotraceNimbleheart, Inc. | Aug 23, 2017117 days |
| K140096 | HMICRO WIRED C-PATCH ELECTRODE, TRADE NAME SUBJECT TO CHANGEHmicro, Inc. | Jul 24, 2014191 days |
| K123658 | BIOMODULE 3-M1Zephyr Technology Corporation | Apr 24, 2013147 days |
| K101576 | CENTERRIDGE DIAGNOSTIC ECG ELECTRODE, CENTERRIDGE MONITORING ECG ELECTRODESTeam Innovations, Inc. | Jul 22, 201045 days |
| K101685 | IP-SETIntegral Process Sas | Jul 1, 201015 days |
| K100129 | AMBU BLUE SENSOR NEO, AMBU BLUE SENSOR NEO XAmbu A/S | Jun 11, 2010143 days |
| K093327 | INTCO TAB ELECTRODEShanghai Intco Electrode Manufacturing Co., Ltd. | May 10, 2010196 days |
| K100315 | 12 LEAD GLOVEI Needmd, Inc. | Apr 23, 201078 days |
| K092744 | NEONATAL ECG ELECTRODE, M203KENR & D Medical Products, Inc. | Dec 10, 200993 days |
| K092389 | DISPOSAL ECG ELECTRODES, MODEL EASYRODESuzuken Co., Ltd. | Oct 13, 200969 days |
Recalls
Recall records for product code DRX per year, by the date the recall began. Each recalled product is one record; these 20 records belong to 10 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 5 |
| 2021 | 1 |
| 2022 | 6 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0935-2022Class II | Graphic Controls Acquisition CorporationInstructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of… | Mar 15, 2022 |
| Z-0936-2022Class II | Graphic Controls Acquisition CorporationInstructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of… | Mar 15, 2022 |
| Z-0937-2022Class II | Graphic Controls Acquisition CorporationInstructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of… | Mar 15, 2022 |
| Z-0938-2022Class II | Graphic Controls Acquisition CorporationInstructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of… | Mar 15, 2022 |
| Z-0939-2022Class II | Graphic Controls Acquisition CorporationInstructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of… | Mar 15, 2022 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code DRX. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 55 |
| 2018 | 43 |
| 2019 | 50 |
| 2020 | 53 |
| 2021 | 55 |
| 2022 | 61 |
| 2023 | 113 |
| 2024 | 61 |
| 2025 | 250 |
| 2026 (so far) | 358 |
Last 12 months by event type
- Deaths1
- Injuries302
- Malfunctions90
- Other or not stated3
Most reported device problems, last 12 months
- Biocompatibility160
- Adverse Event Without Identified Device or Use Problem136
- Loss of or Failure to Bond36
- Adhesive Too Strong10
- Insufficient Device Problem Information10
- Patient Device Interaction Problem5
In plain English: what the newest reports describe Written by AI
Many reports describe skin irritation under the adhesive electrode patch, with redness, itching, rash, and blisters or welts. Most patients were treated with prescription creams or pills, usually without the medication being named, and many patients stopped or paused using the patch.
19 reports: 2133409-2026-00252, 2133409-2026-00265, 2133409-2026-00245, 2133409-2026-00246, 2133409-2026-00248, 2133409-2026-00250, 2133409-2026-00230, 2133409-2026-00231, 2133409-2026-00229, 2133409-2026-00228, 2133409-2026-00220, 2133409-2026-00221, 2133409-2026-00201, 2133409-2026-00213, 2133409-2026-00238, 2133409-2026-00240, 2133409-2026-00244, 2133409-2026-00242, 1317188-2026-00113
Several patients said they had a known sensitivity or allergy to adhesives, and some said they had problems with other tapes or plastics. Other patients said they had no such history, so the link between allergy history and the reaction is unclear.
10 reports: 2133409-2026-00252, 2133409-2026-00265, 2133409-2026-00248, 2133409-2026-00250, 2133409-2026-00230, 2133409-2026-00213, 2133409-2026-00220, 2133409-2026-00201, 2133409-2026-00244, 2133409-2026-00240
Several reports describe open sores, oozing, or wounds on the skin, and one says the sores were infected. These reports involve the skin breaking down, not just redness.
4 reports: 2133409-2026-00201, 2133409-2026-00240, 2133409-2026-00244, 2133409-2026-00242
In one report, the patient had a whole-body rash with small red bumps and pus-filled spots after using the patch, and went to an emergency room. In another report, the patient said the rash on the left breast looked like chigger bites and did not improve after prescribed treatments.
2 reports: 2133409-2026-00245, 2133409-2026-00246
One report says the monitors that display heart waveforms began flickering during function checks before use. The manufacturer planned to replace the systems. This report concerns a monitor rather than the electrode patch, and the reporter's description of the cause is unclear.
Report: MW6004353
Written by AI from the narratives of the 20 newest reports that have one, received Sep 30, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 2,297in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 312registered establishments that list this product code
Other codes under 21 CFR 870.2360
Product codes classified under the same regulation.
Watch DRX in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get DRX every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in DRX. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:DRXRetrieved Nov 28, 2026, 8:40 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:DRXRetrieved Nov 29, 2026, 6:43 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:DRX · count decision_dateRetrieved Nov 28, 2026, 8:40 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:DRXRetrieved Nov 29, 2026, 6:43 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:DRX · count event_date_initiatedRetrieved Nov 28, 2026, 8:40 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:DRX · count res_event_numberRetrieved Nov 28, 2026, 8:40 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0935-2022" "Z-0936-2022" "Z-0937-2022" "Z-0938-2022" "Z-0939-2022")Retrieved Nov 29, 2026, 6:43 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DRX AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:43 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DRX AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:43 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DRX AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 6:43 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:DRXRetrieved Nov 28, 2026, 8:40 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:DRXRetrieved Nov 28, 2026, 8:40 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:870.2360Retrieved Nov 28, 2026, 8:40 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil