Follow LGZ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 880.5725
Review panel
General Hospital
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
762 in the last 12 months
Median review time
155 daysthe latest 25 decisions
Recall records
331 in the last 12 months
Adverse event reports
85last 12 months, 1 death

510(k) and De Novo decisions

Decisions for product code LGZ per year, by decision date, 76 since 1981. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20172
20181
20193
20200
20210
20223
20231
20241
20252
2026 (so far)2

Latest decisions

NumberDevice and applicantDecision and review time
K262532 NORMOFLO™ Irrigation Fluid Warming Sets and Accessories (IR-601, IR-501, IR-61, PL-61)ICU Medical Sep 18, 202658 days
K260988 °M Warmer System with °M StationMequ A/S Apr 24, 202630 days
K252636 Fluido Compact SystemThe Surgical Company International B.V. Nov 18, 202590 days
K251733 Fluido® AirGuard SystemThe Surgical Company International BV (As Tsc Life) Sep 4, 202590 days
K232107 °M Warmer SystemMequ A/S May 31, 2024322 days
K211618 Fluido Compact Blood and Fluid Warming SystemThe Surgical Company International B.V. Jun 5, 2023740 days
K211800 Warrior Blood and Fluid Warmer, Warrior EXTREME Blood and Fluid Warmer, Warrior Lite Blood and Fluid WarmerQuality IN Flow , Ltd. Dec 8, 2022546 days
K202461 ThermaCor 1200 Disposable Sets for the ThermaCor 1200 Rapid Thermal Infusion SystemSmisson-Cartledge Biomedical, LLC Aug 12, 2022715 days
K213191 S-LineBarkey GmbH & Co. KG May 24, 2022237 days
K192325 Quantum Blood and Fluid Warming SystemLife Warmer, Inc. Oct 18, 201952 days

Recalls

Recall records for product code LGZ per year, by the date the recall began. Each recalled product is one record; these 33 records belong to 19 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20192
20200
20218
20220
20230
20240
20255
2026 (so far)1

Latest recalls

Recall and classFirm and reasonBegan
Z-2759-2026Class II ICU Medical Inc.ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent applic… Jun 24, 2026
Z-2231-2025Class II 3M CompanyNotice was issued to clarify flow rates related to the inlet fluid temperature. Apr 22, 2025
Z-1674-2025Class II Belmont Instrument LLCPotential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid… Mar 31, 2025
Z-1675-2025Class II Belmont Instrument LLCPotential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid… Mar 31, 2025
Z-1676-2025Class II Belmont Instrument LLCPotential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid… Mar 31, 2025

Adverse event reports (MAUDE)

Medical device reports received per year that name product code LGZ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201758
201894
2019340
20201,221
2021885
20221,548
20231,033
2024862
2025380
2026 (so far)78

Last 12 months by event type

  • Deaths1
  • Injuries3
  • Malfunctions78
  • Other or not stated3

Most reported device problems, last 12 months

  1. Fluid/Blood Leak31
  2. Overheating of Device10
  3. Crack4
  4. Electrical /Electronic Property Problem4
  5. Leak/Splash4
  6. Smoking4

In plain English: what the newest reports describe Written by AI

  • Several reports describe fluid or blood leaking from connections, cracks or tubing on the device or its disposable parts. In one case a leak occurred at a Y-junction on tubing sets during setup, and in another a leak came from the tubing end connected to the patient.

    8 reports: 9617604-2026-00907, MW6000036, 3012307300-2026-07138, 3012307300-2026-06149, 3012307300-2026-07001, 3012307300-2026-05600, 3012307300-2026-06790, 3012602431-2026-00049

  • A report describes a warmer that smoked and gave off a bad smell, found during preventive maintenance. Another report describes burn marks on internal resistors, which the reporter said could raise a risk of burns or smoke inhalation. In one case the manufacturer's evaluation of the board found a shorted solder joint that was arcing.

    3 reports: 3012307300-2026-05972, 3012307300-2026-07229, 9616567-2026-00037

  • Several reports describe the warmer displaying a system error or overheating. In one liver transplant case the error read "System Error #102," telling staff to discard the disposable and blood, restart with a new disposable, and service the machine if it persisted. The unit was also unusually hot to the touch.

    2 reports: 1219702-2026-00028, MW6000052

  • Two reports describe electrical problems: one reports a leakage current fault, and another reports that the current closed when a float closed the circuit. The reports do not state when these occurred.

    2 reports: 3012307300-2026-07109, 3012307300-2026-07229

  • One report describes a warmer that would not warm up because of a damaged part. Another describes a unit whose temperature and reprime messages kept appearing during patient use, which the clinical engineers linked to intermittent battery-related issues that could not be reproduced during repeated servicing.

    2 reports: 1417592-2026-47668, 25888846

Written by AI from the narratives of the 19 newest reports that have one, received May 31, 2026 to Sep 25, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
598in the FDA’s Global Unique Device Identification Database
Establishment listings
111registered establishments that list this product code

Watch LGZ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get LGZ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LGZ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.