Product codes

FDA product code

KPA: Slide Stainer, Automated (FDA product code)

Follow KPA: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class I
Regulation
21 CFR 864.3800
Review panel
Pathology
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
210 in the last 12 months
Median review time
35 daysthe latest 21 decisions
Recall records
300 in the last 12 months

510(k) and De Novo decisions

Decisions for product code KPA per year, by decision date, 21 since 1977. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K933996 PROFIBLOTTecan U.S., Ltd. Jan 28, 1994164 days
K884258 VERSABLOTAmerican Bionetics, Inc. Dec 14, 198864 days
K881592 WESCOR MODEL 7300 AEROSPRAY GRAM STAINERWescor, Inc. May 13, 198831 days
K880745 LX-100 SLIDE STAINERInnovative Medical Systems, Inc. Apr 27, 198863 days
K874887 WESBLOT PROCESSORAmerican Bionetics, Inc. Mar 24, 1988114 days
K874897 SAKURA DRS-60 AUTOMATIC SLIDE STAINERSakura Finetek U.S.A., Inc. Jan 5, 198835 days
K872916 HISTOMATIC SLIDE STAINER, CODE-ON VERSIONFisher Scientific Co., LLC Sep 14, 198752 days
K853210 MST-1050-II X-RAY UNITRadiological Engineering, Inc. Oct 10, 198571 days
K850333 PUMP TUBE SETS FOR HEMA-TEK I & II SLIDE STAINERSShaban Mfg. Co. Feb 27, 198530 days
K850157 CANNULAS FOR HEMA-TEK I & KK SLIDE STAINERSShaban Mfg. Co. Feb 27, 198543 days

Recalls

Recall records for product code KPA per year, by the date the recall began. Each recalled product is one record; these 30 records belong to 23 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20183
20192
20202
20212
20222
20230
20241
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1919-2024Class II Agilent TechnologiesThere is the potential for the front plexiglass door of the automated slide processing system to break causing potential injury. Apr 2, 2024
Z-1374-2022Class II Ventana Medical Systems IncThere is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire Jun 1, 2022
Z-0894-2022Class II Ventana Medical Systems IncPotential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that c… Jan 20, 2022
Z-1835-2021Class II Hardy DiagnosticsInstruments were assembled and serviced using a defective lot of tubing, which is part of the pump mechanism within the unit. Apr 16, 2021
Z-1333-2021Class II Biocare Medical, LLCTheir is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range of m… Feb 25, 2021

Adverse event reports (MAUDE)

Medical device reports received per year that name product code KPA. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

This section didn’t load from openFDA. It’s retried within the hour.

Last 12 months by event type

This section didn’t load from openFDA. It’s retried within the hour.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
284in the FDA’s Global Unique Device Identification Database
Establishment listings
97registered establishments that list this product code

Product codes classified under the same regulation.

Watch KPA in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get KPA every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in KPA. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.