FDA product code
KEY: Stainer, Tissue, Automated (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 864.3800
- Review panel
- Pathology
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 10 in the last 12 months
- Median review time
- 23 daysthe latest 1 decisions
- Recall records
- 50 in the last 12 months
510(k) and De Novo decisions
Decisions for product code KEY per year, by decision date, 1 since 1977. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K771714 | SLIDE STAINERS GLS 360 & 360-1Honeywell, Inc. | Oct 5, 197723 days |
Recalls
Recall records for product code KEY per year, by the date the recall began. Each recalled product is one record; these 5 records belong to 5 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1016-2010 | Biogenex LaboratoriesUnder certain conditions, product system failure has the potential to cause operator injury due to splash of Hemo-De solution. | Jan 23, 2009 |
| Z-1929-2009 | Abbott MolecularThermal runaway conditions, in that the target temperature was exceeded and the temperature display could not track the actual temperature within the heated sl… | Sep 15, 2008 |
| Z-1471-2008 | Biogenex LaboratoriesBulk reagent dispensing mechanism may break resulting in reagent spill and possible exposure to chemicals. | Jan 24, 2008 |
| Z-0553-2008 | Ventana Medical Systems IncFluid Leaks-The T fitting located in the Symphony Clear system of the Symphony High Volume Slide Stainer needs to be larger to securely hold the tubing in plac… | Oct 30, 2007 |
| Z-0300-2008 | Ventana Medical Systems IncShock hazard: Devices released prior to full testing; the configuration of the Symphony instrument during Hi pot testing procedures did not completely evaluate… | Sep 6, 2007 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code KEY. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
This section didn’t load from openFDA. It’s retried within the hour.
Last 12 months by event type
This section didn’t load from openFDA. It’s retried within the hour.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 19in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 12registered establishments that list this product code
Other codes under 21 CFR 864.3800
Product codes classified under the same regulation.
Watch KEY in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get KEY every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in KEY. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:KEYRetrieved Nov 28, 2026, 2:50 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:KEYRetrieved Nov 28, 2026, 2:50 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:KEY · count decision_dateRetrieved Nov 28, 2026, 2:50 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:KEYRetrieved Nov 28, 2026, 2:50 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:KEY · count event_date_initiatedRetrieved Nov 28, 2026, 2:50 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:KEY · count res_event_numberRetrieved Nov 28, 2026, 2:50 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1016-2010" "Z-1929-2009" "Z-1471-2008" "Z-0553-2008" "Z-0300-2008")Retrieved Nov 28, 2026, 2:50 PM UTC.
- device/udi · search product_codes.code:KEYRetrieved Nov 28, 2026, 2:50 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:KEYRetrieved Nov 28, 2026, 2:50 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:864.3800Retrieved Nov 29, 2026, 6:59 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.