FDA product code
DHA: System, Test, Human Chorionic Gonadotropin (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 862.1155
- Review panel
- Clinical Chemistry
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 720 in the last 12 months
- Median review time
- 144 daysthe latest 25 decisions
- Recall records
- 140 in the last 12 months
- Adverse event reports
- 220last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code DHA per year, by decision date, 72 since 1986. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 1 |
| 2021 | 3 |
| 2022 | 1 |
| 2023 | 3 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K233581 | VITROS Immunodiagnostic Products Total ß-hCG II Reagent PackOrtho Clinical Diagnostics | May 8, 2024183 days |
| K223690 | iFlash-HCG; Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C)Shenzhen Yhlo Biotech Co., Ltd. | Dec 11, 2023367 days |
| K230937 | Alinity i Total ß-hCG Reagent Kit, GLP systems TrackAbbott Laboratories | Jun 5, 202363 days |
| K230790 | Alinity i Total B-hCG Reagent Kit, Alinity c Glucose Reagent Kit, Alinity c ICT Sample Diluent, Alinity ci-seriesAbbott Laboratories | May 19, 202358 days |
| K221990 | Access Total ßhCG (5th IS)Beckman Coulter, Inc. | Dec 27, 2022174 days |
| K212221 | Diazyme DZ-Lite iFlash Total beta-hCG Assay, Diazyme DZ-Lite iFlash 1800 Chemiluminescence Immunoassay AnalyzerDiazyme Laboratories, Inc. | Dec 13, 2021150 days |
| K203227 | Elecsys HCG STATRoche Diagnostics | Aug 18, 2021289 days |
| K200210 | ADVIA Centaur® Total hCG assaySiemens Healthcare Diagnostics, Inc. | Jul 13, 2021532 days |
| K192547 | MAGLUMI 2000 HCG/ß-HCGShenzhen New Industries Biomedical Engineering Co., Ltd. | Jan 17, 2020122 days |
| K192790 | Atellica IM Total hCG (ThCG)Siemens Healthcare Diagnostics, Inc. | Oct 30, 201930 days |
Recalls
Recall records for product code DHA per year, by the date the recall began. Each recalled product is one record; these 14 records belong to 10 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 2 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0958-2023Class III | Abbott LaboratoriesReaction vessels may contain a potential contaminant that could affect their optical performance. | Nov 24, 2022 |
| Z-1015-2022Class II | Siemens Healthcare Diagnostics, Inc.Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors. | Feb 22, 2022 |
| Z-0666-2019Class II | Abbott Ireland Diagnostics DivisionAbbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appen… | Nov 21, 2018 |
| Z-2084-2018Class II | Tosoh Bioscience IncAsfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, dependi… | Mar 5, 2018 |
| Z-1700-2016Class II | Ortho-Clinical DiagnosticsWhen testing was conducted using certain lots, VITROS Systems generated results within the measuring range for samples known to not contain measurable hCG. The… | Feb 18, 2016 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code DHA. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 101 |
| 2018 | 103 |
| 2019 | 152 |
| 2020 | 115 |
| 2021 | 210 |
| 2022 | 185 |
| 2023 | 194 |
| 2024 | 261 |
| 2025 | 238 |
| 2026 (so far) | 198 |
Last 12 months by event type
- Deaths0
- Injuries16
- Malfunctions204
- Other or not stated0
Most reported device problems, last 12 months
- False Positive Result101
- High Test Results52
- Non Reproducible Results48
- Low Test Results46
- False Negative Result12
- Incorrect Measurement8
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 77in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 81registered establishments that list this product code
Other codes under 21 CFR 862.1155
Product codes classified under the same regulation.
Watch DHA in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get DHA every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in DHA. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:DHARetrieved Nov 28, 2026, 8:38 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:DHARetrieved Nov 29, 2026, 6:10 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:DHA · count decision_dateRetrieved Nov 28, 2026, 8:38 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:DHARetrieved Nov 29, 2026, 6:10 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:DHA · count event_date_initiatedRetrieved Nov 28, 2026, 8:38 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:DHA · count res_event_numberRetrieved Nov 28, 2026, 8:38 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0958-2023" "Z-1015-2022" "Z-0666-2019" "Z-2084-2018" "Z-1700-2016")Retrieved Nov 29, 2026, 6:10 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DHA AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:10 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DHA AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:10 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DHA AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 6:10 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:DHARetrieved Nov 28, 2026, 8:38 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:DHARetrieved Nov 28, 2026, 8:38 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:862.1155Retrieved Nov 29, 2026, 5:00 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil