Product codes

FDA product code

JEB: Table, Surgical With Orthopedic Accessories, Manual (FDA product code)

Follow JEB: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class I
Regulation
21 CFR 878.4950
Review panel
General and Plastic Surgery
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
130 in the last 12 months
Median review time
31 daysthe latest 13 decisions
Recall records
110 in the last 12 months
Adverse event reports
1last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code JEB per year, by decision date, 13 since 1977. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K905816 ORTHOVISION TABLEAmerican Sterilizer Co. Mar 6, 199168 days
K900405 SCHOLTEN FRAMEScholten Surgical Instruments, Inc. Mar 1, 199031 days
K891917 ORTHO-J, INC. BEK-JS-004Ortho-J, Inc. May 26, 198960 days
K891387 WILLIAMS BELT & BOOTArthro-Medic, Inc. May 23, 198970 days
K881101 K ORTHOPEDIC TABLEInstrumentation Specialties, Inc. Apr 6, 198822 days
K880263 OSI RIZZOLI ORTHOPEDIC TABLE/MANUAL VERSIONOrthopedic Systems, Inc. Mar 15, 198853 days
K873273 ORTHOPAEDIC TABLE 1420Stierlen-Maquet AG Sep 17, 198731 days
K873276 EXTENSION DEVICE 1416Stierlen-Maquet AG Sep 17, 198731 days
K873274 PLASTER TABLE 1418Stierlen-Maquet AG Sep 17, 198731 days
K873275 EXTENSION TABLE 1417Stierlen-Maquet AG Sep 17, 198731 days

Recalls

Recall records for product code JEB per year, by the date the recall began. Each recalled product is one record; these 11 records belong to 7 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20173
20180
20191
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1494-2019Class II Zimmer Surgical IncForeign Material in the Alvarado Knee Holder Base Plate Assembly and Foot Piece and Alvarado II Base Plate and Foot Piece due to potentially inadequate cleanin… Apr 16, 2019
Z-0641-2018Class II Allen Medical SystemsThe base of the frame has the potential to crack when the patient is initially transferred to the Bow Frame. Oct 6, 2017
Z-0642-2018Class II Allen Medical SystemsThe base of the frame has the potential to crack when the patient is initially transferred to the Bow Frame. Oct 6, 2017
Z-0803-2018Class II Zimmer Biomet, Inc.Zimmer Biomet is conducting a medical device recall for the 2nd Assistant Knee Positioner due to the potential for the sterile packaging barrier to be breached… Sep 7, 2017
Z-0371-2017Class II Schaerer Medical USA IncThe floor locking cylinders contained in the base of the Schaerer 7300 model surgical tables were manufactured with the incorrect "O" ring which may cause the… Sep 27, 2016

Adverse event reports (MAUDE)

Medical device reports received per year that name product code JEB. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20173
20183
20191
20201
20211
20220
20232
20240
20250
2026 (so far)1

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions1
  • Other or not stated0

Most reported device problems, last 12 months

  1. Misassembled1

In plain English: what the newest reports describe Written by AI

  • Several reports describe a table part, such as a boot, stirrup, or adaptor, that loosened or came apart, so the patient's leg or the attached part moved or dropped. In one report the patient's leg fell from the boot and the surgeon fell backward; in another a leg dropped during a transfer, and the procedure was delayed or completed after adjustment.

    6 reports: 1043572-2018-00050, 1043572-2016-00021, 1043572-2014-00071, 6303585, MW5039955, 1526439-2023-01945

  • Several reports describe the operating table or bed moving unexpectedly. In some cases the table slowly drifted down or tilted while a patient was on it, and one report says a bed disengaged while a patient was still on it.

    5 reports: 7074353, MW5074597, 8586190, 1043572-2015-00113, 4371782

  • Two reports describe a mounting bracket or caster wheel breaking. In one report the bracket broke twice, and in another a wheel fell off and one side of the table dropped about 4 inches, tilting the patient, who was not harmed. The manufacturer's inspection of the wheel found older damage to the stem flange that routine maintenance does not check.

    2 reports: 12266780, 4371782

  • One report describes a table's height or positioning controls failing. In one case the table could not be raised using the remote or buttons, and in another two buttons did not work for positioning, so the optimal position was not reached.

    2 reports: 9859825, 6053619

  • One report describes a screw that was tightened so the rod was not locked in place. During placement, a part was found inserted about 1 to 1.5 inches into the patient's rectum, causing a tear on the outside only, with no internal damage found.

    Report: 3006460162-2026-00029

Written by AI from the narratives of the 19 newest reports that have one, received Aug 13, 2014 to Jun 19, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
133in the FDA’s Global Unique Device Identification Database
Establishment listings
44registered establishments that list this product code

Watch JEB in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get JEB every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in JEB. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.