Product codes

FDA product code

BWN: Table And Attachments, Operating-Room (FDA product code)

Follow BWN: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class I
Regulation
21 CFR 878.4950
Review panel
General and Plastic Surgery
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
200 in the last 12 months
Median review time
32 daysthe latest 20 decisions
Recall records
751 in the last 12 months
Adverse event reports
15last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code BWN per year, by decision date, 20 since 1979. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K942699 AIRFRAME SURGICAL DRAPE SUPPORTAlpha Surgical, Inc. Jul 13, 199435 days
K935441 INFINITECH BREATH EZInfinitech, Inc. Feb 7, 199487 days
K925005 ISI SELF-SUPPORTING ARMInstrumentation Specialties, Inc. Apr 29, 1993209 days
K922288 HOLDER, PEDIATRIC POSITIONYamashita and Hemsley Business Trust Feb 10, 1993271 days
K910783 SHAMPAINE MODEL 3800 OR MURANAKA MODEL OR-604MSmith & Nephew Equipment Group Apr 2, 199136 days
K902865 BAUMER SURGICAL TABLESds Dental, Inc. Sep 10, 199073 days
K884558 IOP ADJUSTABLE STIRRUPMend Technologies, Inc. Nov 17, 198817 days
K875291 SPINAL SURGERY FRAMET. Korossurgical Instruments Corp. Jan 22, 198823 days
K873272 ENDOSCOPY TABLE 1532 ENDOSTARStierlen-Maquet AG Sep 8, 198722 days
K860177 ALLEN UNIVERSAL STIRRUPEdgewater Medical Equipment Systems Feb 11, 198625 days

Recalls

Recall records for product code BWN per year, by the date the recall began. Each recalled product is one record; these 75 records belong to 4 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20211
20220
20230
20240
20250
2026 (so far)1

Latest recalls

Recall and classFirm and reasonBegan
Z-0076-2027Class II QUANTUM OPS IncDue to a manufacturing issue (inadequate weld) for patient head positioning system. Jul 24, 2026
Z-0072-2022Class II Innovative Orthopedic Technologies, LLCDue to the holes not being bored to the specific diameter, the locking pins are unable to engage under a load. Sep 13, 2021
Z-1958-2017Class II The Anspach Effort, Inc.Supplied Directions for Use (DFU) did not contain a recommended service interval. Oct 6, 2016
Z-1959-2017Class II The Anspach Effort, Inc.Supplied Directions for Use (DFU) did not contain a recommended service interval. Oct 6, 2016
Z-1960-2017Class II The Anspach Effort, Inc.Supplied Directions for Use (DFU) did not contain a recommended service interval. Oct 6, 2016

Adverse event reports (MAUDE)

Medical device reports received per year that name product code BWN. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20181
20197
20209
202116
20226
202350
202451
202515
2026 (so far)13

Last 12 months by event type

  • Deaths0
  • Injuries4
  • Malfunctions11
  • Other or not stated0

Most reported device problems, last 12 months

  1. Unintended Movement5
  2. Break3
  3. Fracture3
  4. Detachment of Device or Device Component2
  5. Insufficient Device Problem Information1
  6. Material Separation1

In plain English: what the newest reports describe Written by AI

  • Three reports describe the beach chair's welded joints failing. In one, the base separated while the patient was being placed on the chair. In another, the headrest top section broke off during assembly before any clinical use. In the third, the top piece that supports the head broke off while the anesthetized patient was on the table, delaying surgery. No injuries were reported in any of them.

    3 reports: 3019856155-2026-00003, 3019385615-2026-00001, 3019856155-2026-00002

  • Two reports describe a patient's leg slipping out of its holder during surgery. In one, a traction boot came loose during a hip replacement, and the surgeon found a fracture of the thigh bone that had to be wired before the operation could finish, adding about 30 minutes. In the other, staff repositioned the leg and the procedure was completed with no injury reported.

    2 reports: 8010652-2026-0000053, 3009611746-2026-00001

  • Two reports describe arm or leg support parts breaking during procedures. One reports an armboard with a gravity latch breaking off and falling. The other reports a piece of a split leg positioner breaking, and the procedure was finished after a short delay using a different accessory. No injuries were reported.

    2 reports: 1043572-2026-00071, 1043572-2026-00013

  • One report describes a loose plate on the table moving while staff removed a head attachment during positioning for a collarbone repair. The patient's upper body dropped rapidly backward, and the surgery was postponed to the next day. The patient was taken for imaging.

    Report: 8010652-2026-0000054

  • One report describes a carbon-fibre shoulder accessory breaking at the end of surgery while the patient was still sedated, and the accessory fell to the floor. The fall pulled out the breathing tube attached to it, which is called extubation. The surgeon and anesthesiologist assessed the patient and reported no injuries.

    Report: 8010652-2025-0000030

Written by AI from the narratives of the 17 newest reports that have one, received Aug 1, 2025 to Oct 9, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
1,061in the FDA’s Global Unique Device Identification Database
Establishment listings
156registered establishments that list this product code

Watch BWN in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get BWN every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in BWN. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.