FDA product code
IZF: System, X-Ray, Tomographic (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 892.1740
- Review panel
- Radiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 421 in the last 12 months
- Median review time
- 122 daysthe latest 25 decisions
- Recall records
- 60 in the last 12 months
- Adverse event reports
- 0last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code IZF per year, by decision date, 42 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 1 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 2 |
| 2024 | 1 |
| 2025 | 2 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K252649 | TAP2DNano-X Imaging , Ltd. | Feb 2, 2026165 days |
| K252133 | Adaptix Ortho350Adaptix Limited | Nov 6, 2025122 days |
| K250850 | Nanox.ARC XNano-X Imaging , Ltd. | Apr 16, 202527 days |
| K242395 | Nanox.ARCNano-X Imaging , Ltd. | Dec 4, 2024113 days |
| K222934 | Nanox.ARCNano-X Imaging , Ltd. | Apr 28, 2023214 days |
| K221949 | Ortho Device, ADAPTIX 3D Orthopedic Imaging SystemPausch Medical GmbH | Jan 26, 2023205 days |
| K213081 | CLISIS SYSTEMS, Discovery RF180General Medical Merate S.P.A | Oct 26, 202133 days |
| K193262 | DR 600 with TomosynthesisAgfa N.V. | Mar 9, 2020104 days |
| K183275 | DR 800 with TomosynthesisAgfa N.V. | Feb 1, 201970 days |
| K132261 | DISCOVERY XR656 WITH VOLUMERAD (DIGITAL TOMOSYNTHESIS)Ge Medical Systems, LLC | Nov 18, 2013122 days |
Recalls
Recall records for product code IZF per year, by the date the recall began. Each recalled product is one record; these 6 records belong to 4 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 3 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0923-2026Class II | AGFA Healthcare Corp.It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on t… | Nov 18, 2025 |
| Z-1502-2020Class II | Agfa N.V.Tomosynthesis image acquisition sequence did not stop automatically after expected number of exposures. | Feb 10, 2020 |
| Z-3237-2018Class II | Philips Electronics North America CorporationThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fa… | Jul 16, 2018 |
| Z-3238-2018Class II | Philips Electronics North America CorporationThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fa… | Jul 16, 2018 |
| Z-3239-2018Class II | Philips Electronics North America CorporationThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fa… | Jul 16, 2018 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code IZF. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Last 12 months by event type
No reports received in the last 12 months.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 14in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 32registered establishments that list this product code
Watch IZF in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get IZF every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in IZF. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:IZFRetrieved Nov 28, 2026, 8:59 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:IZFRetrieved Nov 29, 2026, 7:26 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:IZF · count decision_dateRetrieved Nov 28, 2026, 8:59 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:IZFRetrieved Nov 29, 2026, 7:26 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:IZF · count event_date_initiatedRetrieved Nov 28, 2026, 8:59 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:IZF · count res_event_numberRetrieved Nov 28, 2026, 8:59 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0923-2026" "Z-1502-2020" "Z-3237-2018" "Z-3238-2018" "Z-3239-2018")Retrieved Nov 29, 2026, 7:26 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:IZF AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:26 PM UTC.
- device/event · search device.device_report_product_code:IZF AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:26 PM UTC.
- device/event · search device.device_report_product_code:IZF AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:26 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:IZFRetrieved Nov 28, 2026, 8:59 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:IZFRetrieved Nov 28, 2026, 8:59 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:892.1740Retrieved Nov 28, 2026, 8:59 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil