FDA product code
IMM: Footrest, Wheelchair (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 890.3920
- Review panel
- Physical Medicine
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 20 in the last 12 months
- Median review time
- 72 daysthe latest 2 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 8last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code IMM per year, by decision date, 2 since 1982. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
Recalls
Recall records for product code IMM per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code IMM. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 8 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions8
- Other or not stated0
Most reported device problems, last 12 months
- Activation Problem3
- Break1
- Failure to Align1
- Fracture1
- Material Deformation1
- Material Separation1
In plain English: what the newest reports describe Written by AI
Several reports describe welds on the front rigging (the metal bar and hardware that hold the footrests to the chair) breaking, including a release lever that broke at its weld and a connector whose weld gave way. One report says the weld was incorrectly made out of the box.
5 reports: 9616091-2015-01616, 1525712-2015-02525, 9616091-2015-00394, 1531186-2014-06480, 3008262382-2015-00932
Several reports describe footplates (the plates the feet rest on) arriving or being used cracked, broken or split. One dealer reported receiving cracked plates repeatedly from a hospital.
4 reports: 1531186-2014-06931, 1531186-2014-06242, 1531186-2014-06244, 1531186-2015-00835
Reports describe parts coming loose or breaking during handling. One says a pin broke off while footrests were being fitted to a chair, another says the swing-away legs were damaged, another says the pads pulled through their mount with the screws still attached, and another says a footrest broke.
4 reports: 1531186-2014-06479, 1417592-2026-50305, 9616091-2015-01470, 1525712-2015-02661
Reports describe leg rests that would not lock in place, elevate, or fold properly. One report says the footrests could not reach the bracket on the frame, and another says a ratchet locking mechanism seized up and a cam broke.
3 reports: 1417592-2026-60317, 1531186-2014-06803, 1525712-2015-02750
In one report, a patient's feet got caught in the footrests while trying to stand, and she fell forward and fractured her pelvis, which needed surgery. In another, a split footplate caught a user's foot on rough terrain and the user fell forward; X-rays showed no broken bones, but the user reported pain.
2 reports: 3012316249-2023-00028, 1531186-2015-00835
Written by AI from the narratives of the 20 newest reports that have one, received Dec 12, 2014 to Oct 8, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 122in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 63registered establishments that list this product code
Other codes under 21 CFR 890.3920
Product codes classified under the same regulation.
Watch IMM in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get IMM every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in IMM. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:IMMRetrieved Nov 29, 2026, 7:43 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:IMMRetrieved Nov 29, 2026, 7:43 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:IMM · count decision_dateRetrieved Nov 29, 2026, 7:43 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:IMMRetrieved Nov 29, 2026, 7:43 PM UTC.
- device/recall · search product_code:IMM · count event_date_initiatedRetrieved Nov 29, 2026, 7:43 PM UTC.
- device/recall · search product_code:IMM · count res_event_numberRetrieved Nov 29, 2026, 7:43 PM UTC.
- device/event · search device.device_report_product_code:IMM AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:43 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:IMM AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:43 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:IMM AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:43 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:IMMRetrieved Nov 29, 2026, 7:43 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:IMMRetrieved Nov 29, 2026, 7:43 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:890.3920Retrieved Nov 28, 2026, 9:46 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil