FDA product code
HWN: Instrument, Compression (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 888.4540
- Review panel
- Orthopedic
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 40 in the last 12 months
- Median review time
- 24 daysthe latest 4 decisions
- Recall records
- 90 in the last 12 months
- Adverse event reports
- 137last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code HWN per year, by decision date, 4 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K854241 | AOT EXTERNAL COMPRESS DEVICEAdvanced Ortho-Technology, Inc. | Mar 3, 1986133 days |
| K792253 | ARICULATED TENSION DEVICE SYNTHES TENSynthes (Usa) | Dec 5, 197928 days |
| K770349 | RIGID TEACHING ATTACHMENTRichard Wolf Medical Instruments Corp. | Mar 7, 197713 days |
| K760970 | FLORIO CPC BANDRichard'S Medical Equip., Inc. | Nov 23, 197619 days |
Recalls
Recall records for product code HWN per year, by the date the recall began. Each recalled product is one record; these 9 records belong to 1 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0493-2011 | Zimmer Inc.The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, incre… | Feb 16, 2010 |
| Z-0494-2011 | Zimmer Inc.The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, incre… | Feb 16, 2010 |
| Z-0495-2011 | Zimmer Inc.The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, incre… | Feb 16, 2010 |
| Z-0496-2011 | Zimmer Inc.The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, incre… | Feb 16, 2010 |
| Z-0497-2011 | Zimmer Inc.The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, incre… | Feb 16, 2010 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code HWN. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 29 |
| 2018 | 21 |
| 2019 | 37 |
| 2020 | 51 |
| 2021 | 88 |
| 2022 | 64 |
| 2023 | 82 |
| 2024 | 83 |
| 2025 | 181 |
| 2026 (so far) | 122 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions137
- Other or not stated0
Most reported device problems, last 12 months
- Break82
- Device-Device Incompatibility23
- Material Deformation9
- Material Twisted/Bent8
- Crack6
- Entrapment of Device5
In plain English: what the newest reports describe Written by AI
Handles on patella clamps, the tools used to press and seat the kneecap implant in knee replacement surgery, broke off while the surgeon was tightening them. Several reports describe this, and in some the broken handle or clamp had to be removed from the patient or sawed out, adding time to surgery.
9 reports: 1818910-2026-17922, 1818910-2026-17822, 1818910-2026-15309, 1818910-2026-14759, 1818910-2026-14374, 1818910-2026-15995, 1818910-2026-16343, 1818910-2026-13130, 1818910-2026-14177
Rubber rings on the patella clamp were reported to be cracked, worn, or disintegrated. In one report the ring disintegrated before surgery and the sterility of the instrument could not be confirmed, while another said the clamp needed replacement.
3 reports: 1818910-2026-14351, 1818910-2026-13644, 1818910-2026-14756
Impact tools used to seat knee implants, including femoral and tibial impactors and a trial poly extractor, broke during surgery in a report. The manufacturer's findings are not stated for these reports.
2 reports: 1818910-2026-14177, 1818910-2026-15893
A screwdriver's teeth became worn, so it slipped while unscrewing screws and the locking mechanism did not work properly. One report says the screwdriver had been used only a few times.
2 reports: 1526439-2026-01117, 8030965-2026-09952
A fragment of a wire used to hold a clavicle hook in place broke off during insertion and was left in the patient because it could not be retrieved from the screw hole. The report says the surgery was completed without delay, and no further information is available.
Report: 8030965-2026-10351
Written by AI from the narratives of the 20 newest reports that have one, received Jul 30, 2026 to Oct 16, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 753in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 171registered establishments that list this product code
Other codes under 21 CFR 888.4540
Product codes classified under the same regulation.
Watch HWN in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get HWN every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HWN. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:HWNRetrieved Nov 28, 2026, 8:47 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HWNRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HWN · count decision_dateRetrieved Nov 28, 2026, 8:47 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:HWNRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HWN · count event_date_initiatedRetrieved Nov 28, 2026, 8:47 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HWN · count res_event_numberRetrieved Nov 28, 2026, 8:47 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0493-2011" "Z-0494-2011" "Z-0495-2011" "Z-0496-2011" "Z-0497-2011")Retrieved Nov 29, 2026, 4:38 PM UTC.
- device/event · search device.device_report_product_code:HWN AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HWN AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HWN AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:HWNRetrieved Nov 28, 2026, 8:47 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:HWNRetrieved Nov 28, 2026, 8:47 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:888.4540Retrieved Nov 28, 2026, 11:20 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil