Product codes

FDA product code

HWN: Instrument, Compression (FDA product code)

Follow HWN: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class I
Regulation
21 CFR 888.4540
Review panel
Orthopedic
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
40 in the last 12 months
Median review time
24 daysthe latest 4 decisions
Recall records
90 in the last 12 months
Adverse event reports
137last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code HWN per year, by decision date, 4 since 1976. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K854241 AOT EXTERNAL COMPRESS DEVICEAdvanced Ortho-Technology, Inc. Mar 3, 1986133 days
K792253 ARICULATED TENSION DEVICE SYNTHES TENSynthes (Usa) Dec 5, 197928 days
K770349 RIGID TEACHING ATTACHMENTRichard Wolf Medical Instruments Corp. Mar 7, 197713 days
K760970 FLORIO CPC BANDRichard'S Medical Equip., Inc. Nov 23, 197619 days

Recalls

Recall records for product code HWN per year, by the date the recall began. Each recalled product is one record; these 9 records belong to 1 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0493-2011 Zimmer Inc.The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, incre… Feb 16, 2010
Z-0494-2011 Zimmer Inc.The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, incre… Feb 16, 2010
Z-0495-2011 Zimmer Inc.The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, incre… Feb 16, 2010
Z-0496-2011 Zimmer Inc.The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, incre… Feb 16, 2010
Z-0497-2011 Zimmer Inc.The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, incre… Feb 16, 2010

Adverse event reports (MAUDE)

Medical device reports received per year that name product code HWN. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201729
201821
201937
202051
202188
202264
202382
202483
2025181
2026 (so far)122

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions137
  • Other or not stated0

Most reported device problems, last 12 months

  1. Break82
  2. Device-Device Incompatibility23
  3. Material Deformation9
  4. Material Twisted/Bent8
  5. Crack6
  6. Entrapment of Device5

In plain English: what the newest reports describe Written by AI

  • Handles on patella clamps, the tools used to press and seat the kneecap implant in knee replacement surgery, broke off while the surgeon was tightening them. Several reports describe this, and in some the broken handle or clamp had to be removed from the patient or sawed out, adding time to surgery.

    9 reports: 1818910-2026-17922, 1818910-2026-17822, 1818910-2026-15309, 1818910-2026-14759, 1818910-2026-14374, 1818910-2026-15995, 1818910-2026-16343, 1818910-2026-13130, 1818910-2026-14177

  • Rubber rings on the patella clamp were reported to be cracked, worn, or disintegrated. In one report the ring disintegrated before surgery and the sterility of the instrument could not be confirmed, while another said the clamp needed replacement.

    3 reports: 1818910-2026-14351, 1818910-2026-13644, 1818910-2026-14756

  • Impact tools used to seat knee implants, including femoral and tibial impactors and a trial poly extractor, broke during surgery in a report. The manufacturer's findings are not stated for these reports.

    2 reports: 1818910-2026-14177, 1818910-2026-15893

  • A screwdriver's teeth became worn, so it slipped while unscrewing screws and the locking mechanism did not work properly. One report says the screwdriver had been used only a few times.

    2 reports: 1526439-2026-01117, 8030965-2026-09952

  • A fragment of a wire used to hold a clavicle hook in place broke off during insertion and was left in the patient because it could not be retrieved from the screw hole. The report says the surgery was completed without delay, and no further information is available.

    Report: 8030965-2026-10351

Written by AI from the narratives of the 20 newest reports that have one, received Jul 30, 2026 to Oct 16, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
753in the FDA’s Global Unique Device Identification Database
Establishment listings
171registered establishments that list this product code

Watch HWN in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get HWN every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HWN. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.