FDA product code
HTQ: Broach (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 888.4540
- Review panel
- Orthopedic
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 00 in the last 12 months
- Median review time
- Not availablethe latest 0 decisions
- Recall records
- 50 in the last 12 months
- Adverse event reports
- 455last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code HTQ per year, by decision date. Review time is the days from receipt to decision.
No decisions on record for this product code.
Recalls
Recall records for product code HTQ per year, by the date the recall began. Each recalled product is one record; these 5 records belong to 5 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2055-2017Class II | Greatbatch MedicalGreatbatch Medical has voluntarily initiated a Recall for the Linear Straight Broach Handle (Rasp Handle) and EZ Clean" Linear Broach Handle. Greatbatch Medica… | Apr 23, 2017 |
| Z-2734-2011 | Encore Medical, LpStrike plate may dislodge from the broach stem handle during impaction of the device. | Apr 21, 2011 |
| Z-1246-2009 | Zimmer Inc.The broach handle may fracture at the threaded tip during use, which would result in surgical intervention to retrieve the broach or the trial as the threaded… | Apr 29, 2009 |
| Z-1262-2009 | Zimmer Inc.The surface finish inside of the broach is not to specification and the instrument may corrode if cleaned and sterilized according to instructions. | Mar 13, 2009 |
| Z-2349-2008 | Stryker Howmedica Osteonics Corp.Rejuvenate Broach system was designed in a manner that may result in the potential for a medical calcar fracture. | Jan 25, 2008 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code HTQ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 225 |
| 2018 | 433 |
| 2019 | 460 |
| 2020 | 466 |
| 2021 | 550 |
| 2022 | 504 |
| 2023 | 576 |
| 2024 | 470 |
| 2025 | 310 |
| 2026 (so far) | 441 |
Last 12 months by event type
- Deaths0
- Injuries21
- Malfunctions434
- Other or not stated0
Most reported device problems, last 12 months
- Break340
- Device-Device Incompatibility92
- No Apparent Adverse Event27
- Material Deformation25
- Activation, Positioning or Separation Problem15
- Unintended Movement13
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 2,067in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 126registered establishments that list this product code
Other codes under 21 CFR 888.4540
Product codes classified under the same regulation.
Watch HTQ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get HTQ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HTQ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:HTQRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HTQRetrieved Nov 29, 2026, 4:38 PM UTC.
- device/510k · search product_code:HTQ · count decision_dateRetrieved Nov 29, 2026, 4:38 PM UTC.
- device/recall · search product_code:HTQRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HTQ · count event_date_initiatedRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HTQ · count res_event_numberRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2055-2017" "Z-2734-2011" "Z-1246-2009" "Z-1262-2009" "Z-2349-2008")Retrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HTQ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HTQ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HTQ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:HTQRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:HTQRetrieved Nov 29, 2026, 4:38 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:888.4540Retrieved Nov 28, 2026, 11:20 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.