FDA product code
GFG: Bit, Surgical (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 878.4820
- Review panel
- General and Plastic Surgery
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 30 in the last 12 months
- Median review time
- 44 daysthe latest 3 decisions
- Recall records
- 170 in the last 12 months
- Adverse event reports
- 108last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code GFG per year, by decision date, 3 since 1982. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
Recalls
Recall records for product code GFG per year, by the date the recall began. Each recalled product is one record; these 17 records belong to 4 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 14 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1838-2017Class II | Smith & Nephew, Inc.Potential for the cutting blade to detach from the device and/or the distal drill head to fracture during retrograde reaming | Mar 15, 2017 |
| Z-1839-2017Class II | Smith & Nephew, Inc.Potential for the cutting blade to detach from the device and/or the distal drill head to fracture during retrograde reaming | Mar 15, 2017 |
| Z-1840-2017Class II | Smith & Nephew, Inc.Potential for the cutting blade to detach from the device and/or the distal drill head to fracture during retrograde reaming | Mar 15, 2017 |
| Z-1841-2017Class II | Smith & Nephew, Inc.Potential for the cutting blade to detach from the device and/or the distal drill head to fracture during retrograde reaming | Mar 15, 2017 |
| Z-1842-2017Class II | Smith & Nephew, Inc.Potential for the cutting blade to detach from the device and/or the distal drill head to fracture during retrograde reaming | Mar 15, 2017 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code GFG. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 28 |
| 2018 | 22 |
| 2019 | 20 |
| 2020 | 52 |
| 2021 | 68 |
| 2022 | 69 |
| 2023 | 52 |
| 2024 | 79 |
| 2025 | 110 |
| 2026 (so far) | 101 |
Last 12 months by event type
- Deaths0
- Injuries6
- Malfunctions102
- Other or not stated0
Most reported device problems, last 12 months
- Fracture65
- Break21
- Material Deformation6
- Material Fragmentation5
- Mechanical Jam4
- Failure to Cut3
In plain English: what the newest reports describe Written by AI
The tip of a screwdriver or bit snapped or sheared off while tightening a screw or inserting a peripheral screw, most often during implant placement. In several reports the broken tip was removed from the patient or did not stay inside, and in one the surgeon switched to a new bit. Reports say there was no patient harm or only a minor delay.
5 reports: 0001649390-2026-01124, 0001649390-2026-01042, 0001649390-2026-00761, 0001649390-2026-00722, 8030965-2026-07754
Instruments broke during use and left pieces that had to be removed from the patient. In one case a fragment was taken out with a grasper and the procedure continued with a back-up device. In another, the surgeon forced an instrument onto a guide pin that had not been removed, and it broke.
3 reports: 1219602-2026-02866, 1219602-2026-02259, 8030965-2026-10016
A cannulated (hollow-centred) drill or reamer did not cut bone properly. In one case it only spun, and in another it was described as dull and met resistance, so the surgeon used a new instrument or a tap instead.
2 reports: 50839481, 3006460162-2026-00044
A bit was bent while being pushed through bone during surgery. When the returned device was inspected, the manufacturer found it twisted and deformed.
Report: 8030965-2026-10045
A reamer broke and its pieces had to be retrieved. The manufacturer did not get the device back, so the cause could not be confirmed from photos, though excessive force was named as a possible cause. Its records showed the product met specifications before shipment.
Report: 1017294-2026-00149
Written by AI from the narratives of the 12 newest reports that have one, received Jul 15, 2026 to Oct 28, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 1,316in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 103registered establishments that list this product code
Other codes under 21 CFR 878.4820
Product codes classified under the same regulation.
Watch GFG in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get GFG every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in GFG. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:GFGRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:GFGRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:GFG · count decision_dateRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:GFGRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:GFG · count event_date_initiatedRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:GFG · count res_event_numberRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1838-2017" "Z-1839-2017" "Z-1840-2017" "Z-1841-2017" "Z-1842-2017")Retrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:GFG AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:GFG AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:GFG AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:GFGRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:GFGRetrieved Nov 29, 2026, 4:25 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:878.4820Retrieved Nov 28, 2026, 7:57 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.