Product codes

FDA product code

FFK: Lithotriptor, Electro-Hydraulic (FDA product code)

Follow FFK: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 876.4480
Review panel
Gastroenterology and Urology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
610 in the last 12 months
Median review time
90 daysthe latest 25 decisions
Recall records
210 in the last 12 months
Adverse event reports
10last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code FFK per year, by decision date, 61 since 1980. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20191
20200
20211
20220
20231
20240
20252
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K243032 EHL Probe (SCDG-AS, SCDG-BS, SCDG-CS)Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd. Mar 21, 2025175 days
K242888 Electrohydraulic Lithotriptor (TCS-B3-II)Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd. Mar 21, 2025179 days
K230488 EL27-Compact; Sterile EHL-ProbesWalz Elektronik GmbH Aug 31, 2023189 days
K202813 URO-TOUCH 9 French ProbeNorthgate Technologies, Inc. Jun 14, 2021263 days
K191124 Swiss LithoClast TrilogyE.M.S Electro Medical Systems S.A May 29, 201930 days
K162119 UreTron PF Series ProbeMed-Sonics Corp. Aug 29, 201628 days
K160198 UreTron MultiProbe Lithotripter SystemMed-Sonics Corp. Feb 23, 201626 days
K132795 CYBERWAND STERILE PROBE SETCybersonics, Inc. Nov 26, 201381 days
K130368 AUTOLITH TOUCH; URO TOUCHNorthgate Technologies, Inc. Nov 15, 2013274 days
K123038 SWISS LITHOBREAKERE.M.S Electro Medical Systems S.A Jan 16, 2013110 days

Recalls

Recall records for product code FFK per year, by the date the recall began. Each recalled product is one record; these 21 records belong to 10 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20171
20187
20190
20203
20211
20221
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1235-2022Class II Northgate Technologies, Inc.Product labeled incorrectly. Apr 13, 2022
Z-0333-2022Class II Northgate Technologies, Inc.The languages for Russian, Romanian, Slovak, and Czech have the power settings for high and low switched in the device. Oct 25, 2021
Z-1883-2020Class II Northgate Technologies, Inc.As part of an investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing. Mar 10, 2020
Z-1884-2020Class II Northgate Technologies, Inc.Investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing. Mar 10, 2020
Z-1885-2020Class II Northgate Technologies, Inc.Investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing. Mar 10, 2020

Adverse event reports (MAUDE)

Medical device reports received per year that name product code FFK. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20179
201818
201910
202039
202122
202213
202310
20248
20256
2026 (so far)10

Last 12 months by event type

  • Deaths0
  • Injuries4
  • Malfunctions6
  • Other or not stated0

Most reported device problems, last 12 months

  1. Unintended Electrical Shock3
  2. Activation Problem2
  3. Connection Problem2
  4. Adverse Event Without Identified Device or Use Problem1

In plain English: what the newest reports describe Written by AI

  • Metal particles or shavings were reported to get into the patient's body during kidney stone procedures. In two reports the metal entered the kidney or body cavity, and in one the surgeon removed it by hand; the reports do not say how the metal got there.

    2 reports: MW5182585, MW5182601

  • Reports describe the probe or cord not producing sparks or pulses. In one case a cord was replaced and a second probe tried, and the representative said cords may wear out and that holding the connection tightly helped; the procedure was then completed with a third probe. In another report the generator produced no pulse because an internal wire was disconnected.

    2 reports: MW5183514, 0001450997-2025-00001

  • Electric shocks to a physician's arm and hand were reported after a probe was used at high power during a scope procedure. The physician reported ongoing arm and hand pain, the patient was reported to be doing well, and the probe was not returned for evaluation.

    2 reports: 0001450997-2026-00003, 0001450997-2026-00004

  • The electrohydraulic probe (the part that makes sparks to break up stones) has been reported to cause electric shocks or sparking outside the patient's body. In one report a nurse was shocked when the probe was stepped on while being removed from a scope, and the report says the nurse was expected to fully recover.

    Report: 0001450997-2026-00006

  • A bile duct perforation (a tear in the duct that carries bile) was reported during stone breaking with the electrohydraulic device. The doctor was unsure how it happened, and a stent was placed; the patient was to be monitored to see whether the tear healed on its own.

    Report: 0001450997-2025-00005

Written by AI from the narratives of the 19 newest reports that have one, received May 30, 2024 to May 26, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
292in the FDA’s Global Unique Device Identification Database
Establishment listings
102registered establishments that list this product code

Product codes classified under the same regulation.

Watch FFK in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get FFK every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in FFK. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.