FDA product code
EXS: Urinometer, Electrical (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.1800
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 00 in the last 12 months
- Median review time
- Not availablethe latest 0 decisions
- Recall records
- 60 in the last 12 months
- Adverse event reports
- 3last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code EXS per year, by decision date. Review time is the days from receipt to decision.
No decisions on record for this product code.
Recalls
Recall records for product code EXS per year, by the date the recall began. Each recalled product is one record; these 6 records belong to 6 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 3 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0554-2024Class II | C.R. Bard IncBD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica device changed dur… | Nov 15, 2023 |
| Z-1293-2022Class II | C.R. Bard IncComplaints relating to urine output measurement accuracy. | May 19, 2022 |
| Z-0565-2018Class II | C.R. Bard, Inc.BMD has identified that some lots of CritiCore(R) monitors may experience failures involving urine output volume measurement fluctuations, core bladder tempera… | Nov 10, 2017 |
| Z-2243-2018Class II | C.R. Bard, Inc.Issues identified with the monitor including urine output measurement errors, temperature measurement errors and undesired alarms. | Nov 10, 2017 |
| Z-0488-2018Class II | C.R. Bard, Inc.The possibility exists that use of the Uros Automated Urine Output and Temperature Monitor may lead to the loss of urine output measurement data and/or urine o… | Sep 29, 2017 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code EXS. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 29 |
| 2018 | 16 |
| 2019 | 18 |
| 2020 | 13 |
| 2021 | 33 |
| 2022 | 226 |
| 2023 | 97 |
| 2024 | 13 |
| 2025 | 16 |
| 2026 (so far) | 3 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions3
- Other or not stated0
Most reported device problems, last 12 months
- Volume Accuracy Problem2
- Device Handling Problem1
- Failure to Sense1
In plain English: what the newest reports describe Written by AI
Several reports describe the hourly display freezing or grouping output wrongly. Output from several hours appears under one hour, or the hour label stays stuck on an earlier time, so nurses cannot see what was produced each hour. In one report the nurse averaged output by hand; a restart often fixed it, and the device sometimes lost earlier data.
4 reports: 1018233-2025-06929, 1018233-2025-04890, 1018233-2025-04949, 1018233-2025-04466
Several reports describe the device restarting or rebooting during use. In one report the restart was linked to a damaged load cell, the internal part that weighs the collected urine.
3 reports: 1018233-2025-04195, 1018233-2025-03031, 1018233-2025-01431
Some reports describe the device's internal control module or screen failing. One report says the screen stayed dark while the power light was on, and another says the control module could not pass a calibration step during a software upgrade and needed replacement.
2 reports: 1018233-2025-07049, 1018233-2025-07722
Reports describe parts of the connection between the device and the drainage bag failing. One report says a plastic cap came off and the bag was hung too high, causing urine to flow backward. Another says the tube sensor could not detect the catheter tubing. The manufacturer agreed to replace broken units free of charge in one case.
2 reports: 1018233-2026-00435, 1018233-2025-03508
A report describes the device being unable to hold power on battery. The biomedical staff found it stayed on only about 10 minutes after unplugging, not the full 60 minutes expected.
Report: 1018233-2025-04814
Written by AI from the narratives of the 19 newest reports that have one, received Dec 13, 2024 to Jul 9, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 17in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 8registered establishments that list this product code
Other codes under 21 CFR 876.1800
Product codes classified under the same regulation.
Watch EXS in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get EXS every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in EXS. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:EXSRetrieved Nov 29, 2026, 5:35 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:EXSRetrieved Nov 29, 2026, 5:35 PM UTC.
- device/510k · search product_code:EXS · count decision_dateRetrieved Nov 29, 2026, 5:35 PM UTC.
- device/recall · search product_code:EXSRetrieved Nov 29, 2026, 5:35 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:EXS · count event_date_initiatedRetrieved Nov 29, 2026, 5:35 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:EXS · count res_event_numberRetrieved Nov 29, 2026, 5:35 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0554-2024" "Z-1293-2022" "Z-0565-2018" "Z-2243-2018" "Z-0488-2018")Retrieved Nov 29, 2026, 5:35 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:EXS AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:35 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:EXS AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:35 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:EXS AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:35 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:EXSRetrieved Nov 29, 2026, 5:35 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:EXSRetrieved Nov 29, 2026, 5:35 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:876.1800Retrieved Nov 29, 2026, 7:17 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.