FDA device news · Product code QEZ

FDA 510(k) clearance

Aventus Thrombectomy System (K262030): Inquis Medical 510(k) clearance

Menlo Park, CA

Follow QEZ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Jul 15, 2026
Received
Jun 17, 2026
510(k) type
Special
Product code
QEZAspiration Thrombectomy Catheter
Review panel
Cardiovascular
Device class
Class II
Regulation
21 CFR 870.5150
Third-party review
No
Summary or statement
Summary
Review time
28 daysreceipt to decision, calendar days
Median in QEZ
101 daysthe latest 24 other decisions
QEZ decisions
7in the last 12 months
Applicant’s 510(k)s
10in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The Aventus Thrombectomy System is a catheter-based manual aspiration system used to remove emboli and thrombi (blood clots) from pulmonary arteries and peripheral blood vessels, and to infuse fluids. A catheter is a thin flexible tube threaded into the body.
What it’s cleared for
The system is cleared for the non-surgical removal of emboli and thrombi from blood vessels, and for injection, infusion, or aspiration of contrast media and other fluids into or from a blood vessel. It is intended for use in the peripheral vasculature and for the treatment of pulmonary embolism. The summary does not specify the intended patient population beyond these uses.
Compared with (the predicate)
  • K253925, Aventus Thrombectomy System
How it differs
The summary says the subject device was modified to add ergonomic enhancements and a pressure sensor in the catheter handle, which provides an additional input for the sensing algorithm. The summary states the intended use and most technological characteristics are the same as the predicate.
Testing the summary lists
  • Sterilization (ISO 14937:2009; sterility assurance level of 10 to the minus 6) and bacterial endotoxins (USP <85>, AAMI ST72)
  • Distribution, packaging, and shelf-life (twelve-month labeled shelf life)
  • Software documentation and testing per FDA guidance
  • Electrical safety and EMC (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62366-1)

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K262030 took 28 calendar days from receipt to decision. Of the latest 24 other decisions in product code QEZ, 23 took longer, 1 took the same time; their median was 101 days.

Latest decisions in QEZ

NumberDevice and applicantDecision and review time
K262864 FLOWRUNNER Aspiration SystemExpanse Medical, Inc. · Traditional Oct 26, 202673 days
K254035 Mayumi™ Thrombectomy SystemMicroVention, Inc., d/b/a Terumo Neuro · Traditional Sep 3, 2026261 days
K261105 Powered Aventus Thrombectomy System™Inquis Medical · Traditional Jul 27, 2026115 days
K262030 Aventus Thrombectomy SystemInquis Medical · Special Jul 15, 202628 days
K253589 Liberant™ RX Aspiration CatheterMedtronic Interventional Vascular, Inc. · Traditional Apr 23, 2026157 days
K254208 FLOWRUNNER Aspiration SystemExpanse Medical, Inc. · Special Mar 20, 202681 days
K253925 Aventus Thrombectomy SystemInquis Medical · Traditional Jan 15, 202638 days
K252509 AlphaVac Multipurpose Mechanical Aspiration System F1885AngioDynamics, Inc. · Traditional Nov 24, 2025108 days
K251312 Vesalio Peripheral SystemVesalio, Inc. · Traditional Oct 24, 2025179 days
K251488 FLOWRUNNER Aspiration SystemExpanse Medical, Inc. · Special Jul 14, 202561 days

Other 510(k)s from Inquis Medical

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K261105 Powered Aventus Thrombectomy System™Inquis Medical · Traditional Jul 27, 2026115 days
K260091 Aventus Introducer SheathInquis Medical · Traditional Apr 13, 202691 days
K253925 Aventus Thrombectomy SystemInquis Medical · Traditional Jan 15, 202638 days
K251189 Aventus Thrombectomy SystemInquis Medical · Traditional Jun 13, 202557 days
K250668 Aventus Thrombectomy System (IM-2100)Inquis Medical · Special Mar 25, 202520 days
K250202 Aventus Thrombectomy SystemInquis Medical · Special Feb 26, 202533 days
K240426 Aventus Clot Management SystemInquis Medical · Traditional May 14, 202491 days
K240117 Aventus Thrombectomy System (IM-2001/IM-0002)Inquis Medical · Traditional Mar 15, 202459 days
K232730 Aventus Thrombectomy SystemInquis Medical · Traditional Nov 1, 202355 days

Watch QEZ in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get QEZ every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in QEZ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.