FDA device news · Product code QEZ
FDA 510(k) clearance
AlphaVac Multipurpose Mechanical Aspiration System F1885 (K252509): AngioDynamics, Inc. 510(k) clearance
Queensburgy, NY
- Decision
- Substantially Equivalent
- Decision date
- Nov 24, 2025
- Received
- Aug 8, 2025
- 510(k) type
- Traditional
- Product code
- QEZAspiration Thrombectomy Catheter
- Review panel
- Cardiovascular
- Device class
- Class II
- Regulation
- 21 CFR 870.5150
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 108 daysreceipt to decision, calendar days
- Median in QEZ
- 89 daysthe latest 24 other decisions
- QEZ decisions
- 7in the last 12 months
- Applicant’s 510(k)s
- 88in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The AlphaVac Multipurpose Mechanical Aspiration (MMA) F1885 System is a single-use, catheter-based system used to remove thrombus (clots), emboli (material that has traveled through the blood), and other clots during minimally invasive percutaneous procedures.
- What it’s cleared for
- The Cannula is cleared for the non-surgical removal of thrombi or emboli from the vasculature and for aspiration and injection of contrast media and other fluids. It is intended for use in the venous system and for the treatment of pulmonary embolism. The Handle is a vacuum source for the system, and the Sheath is a conduit for inserting the Cannula/Obturator and other endovascular devices while minimizing blood loss.
- Compared with (the predicate)
- K213402, Triever24, Triever20
- How it differs
- The summary says the modified device has a changed proximal cannula end and handle that allow aspiration and injection of contrast media and other fluids, a sheath cleared to act as a conduit for endovascular devices while minimizing blood loss, and shorter tubing between the handle and collection bag.
- Testing the summary lists
- Sterilization and packaging integrity testing (ethylene oxide sterilization)
- Biocompatibility (ISO 10993: cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility)
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K252509 took 108 calendar days from receipt to decision. Of the latest 24 other decisions in product code QEZ, 11 took longer, 13 were faster; their median was 89 days.
Latest decisions in QEZ
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262864 | FLOWRUNNER Aspiration SystemExpanse Medical, Inc. · Traditional | Oct 26, 202673 days |
| K254035 | Mayumi™ Thrombectomy SystemMicroVention, Inc., d/b/a Terumo Neuro · Traditional | Sep 3, 2026261 days |
| K261105 | Powered Aventus Thrombectomy System™Inquis Medical · Traditional | Jul 27, 2026115 days |
| K262030 | Aventus Thrombectomy SystemInquis Medical · Special | Jul 15, 202628 days |
| K253589 | Liberant™ RX Aspiration CatheterMedtronic Interventional Vascular, Inc. · Traditional | Apr 23, 2026157 days |
| K254208 | FLOWRUNNER Aspiration SystemExpanse Medical, Inc. · Special | Mar 20, 202681 days |
| K253925 | Aventus Thrombectomy SystemInquis Medical · Traditional | Jan 15, 202638 days |
| K252509 | AlphaVac Multipurpose Mechanical Aspiration System F1885AngioDynamics, Inc. · Traditional | Nov 24, 2025108 days |
| K251312 | Vesalio Peripheral SystemVesalio, Inc. · Traditional | Oct 24, 2025179 days |
| K251488 | FLOWRUNNER Aspiration SystemExpanse Medical, Inc. · Special | Jul 14, 202561 days |
Other 510(k)s from AngioDynamics, Inc.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K263076 | AngioVac F1885AngioDynamics, Inc. · Traditional | Oct 30, 202659 days |
| K253106 | AngioVac CannulaAngioDynamics, Inc. · Special | Oct 24, 202530 days |
| K242687 | NanoKnife Generator (H78720300301US0); NanoKnife Single Electrode Activation Probe, 15 cm (H787204001090); NanoKnife Single Electrode Activation Probe, 25 cm (H787204001100); NanoKnife Single Electrode Probe Spacer (H787204003010)AngioDynamics, Inc. · Traditional | Dec 6, 202491 days |
| K240397 | AlphaVac MMA F1885 System (H787253020)AngioDynamics, Inc. · Traditional | Apr 1, 202452 days |
| K231945 | VenaCure EVLT NeverTouch Procedure Kits; VenaCure EVLT NeverTouch Direct Procedure KitsAngioDynamics, Inc. · Special | Jul 20, 202320 days |
| K223581 | Solero Microwave Tissue Ablation (MTA) System and AccessoriesAngioDynamics, Inc. · Special | Mar 27, 2023116 days |
| K221883 | Solero Microwave Tissue Ablation (MTA) System and AccessoriesAngioDynamics, Inc. · Special | Mar 7, 2023251 days |
| K213388 | AlphaVac Multipurpose Mechanical Aspiration (MMA) F1885 SystemAngioDynamics, Inc. · Traditional | Apr 4, 2022172 days |
| K213067 | Solero Microwave Tissue Ablation (MTA) System and AccessoriesAngioDynamics, Inc. · Special | Dec 3, 202171 days |
| K212386 | AngioVac F18 85AngioDynamics, Inc. · Traditional | Sep 30, 202159 days |
| K211081 | AlphaVac Multipurpose Mechanical Aspiration (MMA) SystemAngioDynamics, Inc. · Traditional | Jun 4, 202153 days |
| K202347 | UNIFUSE Infusion System with Cooper WireAngioDynamics, Inc. · Special | Sep 15, 202028 days |
Watch QEZ in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get QEZ every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in QEZ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K252509Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:QEZRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:QEZRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:QEZ AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"AngioDynamics, Inc."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil