FDA device news · Product code HRS
FDA 510(k) clearance
Alians Elbow S (K261191): Newclip Technics 510(k) clearance
Haute-Goulaine, FR
- Decision
- Substantially Equivalent
- Decision date
- Jul 8, 2026
- Received
- Apr 13, 2026
- 510(k) type
- Traditional
- Product code
- HRSPlate, Fixation, Bone
- Review panel
- Orthopedic
- Device class
- Class II
- Regulation
- 21 CFR 888.3030
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 86 daysreceipt to decision, calendar days
- Median in HRS
- 81 daysthe latest 24 other decisions
- HRS decisions
- 60in the last 12 months
- Applicant’s 510(k)s
- 35in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The Alians Elbow S is a set of metal plates and screws used to hold broken bones or cut bones (osteotomies) in place while they heal. It is made of titanium alloy (Ti-6Al-4V ELI) and is designed for the distal humerus (the lower end of the upper arm bone near the elbow) and the proximal ulna (the upper end of the forearm bone near the elbow) in adults.
- What it’s cleared for
- The implants are cleared for fixation of fractures and osteotomies of the distal humerus and proximal ulna in adults.
- Compared with (the predicate)
- How it differs
- The summary says the indications for use are the same as the predicate devices. It states the device uses the same material as the primary predicate and is substantially equivalent in shape, operating principles, and dimensions. It does not describe any specific technical difference.
- Testing the summary lists
- Engineering analysis and comparative static and dynamic tests on plates (ASTM F382)
- No clinical studies were performed
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K261191 took 86 calendar days from receipt to decision. Of the latest 24 other decisions in product code HRS, 9 took longer, 15 were faster; their median was 81 days.
Latest decisions in HRS
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262939 | HERO Sternal Closure SystemHero Orthopedics, LLC · Traditional | Nov 2, 202673 days |
| K263602 | Arthrex Plates, Screws and Pins, and WashersArthrex, Inc. · Special | Oct 29, 202629 days |
| K261135 | Trochlear Salvage Plating SystemSkeletal Dynamics, Inc. · Traditional | Oct 26, 2026202 days |
| K262779 | ANTHEM™ Fracture SystemGlobus Medical, Inc. · Traditional | Oct 23, 202679 days |
| K263407 | Tempo Plating SystemPace Surgical · Special | Oct 23, 202628 days |
| K263269 | Park's Pectus SystemTdm Co., Ltd. · Special | Oct 15, 202627 days |
| K261611 | TDM Large Bone Plate and Screw SystemTdm Co., Ltd. · Special | Oct 2, 2026140 days |
| K262649 | Alians ForearmNewclip Technics · Traditional | Sep 28, 202660 days |
| K261892 | ProRib Fixation SystemCosta Surgical, Inc. · Traditional | Sep 23, 2026110 days |
| K262098 | EnHAnce PEEK Bunion SystemNvision Biomedical Technologies · Special | Sep 18, 202688 days |
Other 510(k)s from Newclip Technics
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262649 | Alians ForearmNewclip Technics · Traditional | Sep 28, 202660 days |
| K261511 | Xpert AnkleNewclip Technics · Traditional | Jul 28, 202682 days |
| K261038 | Xpert PFPNewclip Technics · Traditional | May 29, 202660 days |
| K253906 | Xpert HandNewclip Technics · Traditional | Mar 11, 202696 days |
| K250767 | Newclip Patient-matched instrumentation non sterile PSINewclip Technics · Traditional | Oct 1, 2025202 days |
| K250155 | Xpert KneeNewclip Technics · Traditional | Apr 23, 202592 days |
| K243912 | Newclip Patient-matched instrumentation non sterile PSINewclip Technics · Traditional | Feb 13, 202556 days |
| K240415 | Newclip Patient-matched instrumentation non sterile PSINewclip Technics · Traditional | Nov 7, 2024269 days |
| K241539 | Activmotion SNewclip Technics · Traditional | Jul 30, 202460 days |
| K221615 | Newclip Patient-matched instrumentation non sterile PSINewclip Technics · Traditional | Oct 25, 2023509 days |
| K221395 | Footmotion Plating SystemNewclip Technics · Traditional | Jul 1, 202249 days |
| K202803 | Activmotion S DTONewclip Technics · Traditional | Dec 16, 2021449 days |
Watch HRS in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get HRS every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in HRS. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K261191Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:HRSRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:HRSRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:HRS AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Newclip Technics"Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil