FDA device news · Product code HRS
FDA 510(k) clearance
I.T.S. PRS Phoenix II (K260054): I.T.S. GmbH 510(k) clearance
Laßnitzhöhe, AT
- Decision
- Substantially Equivalent
- Decision date
- Apr 24, 2026
- Received
- Jan 8, 2026
- 510(k) type
- Traditional
- Product code
- HRSPlate, Fixation, Bone
- Review panel
- Orthopedic
- Device class
- Class II
- Regulation
- 21 CFR 888.3030
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 106 daysreceipt to decision, calendar days
- Median in HRS
- 82 daysthe latest 25 other decisions
- HRS decisions
- 60in the last 12 months
- Applicant’s 510(k)s
- 8in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The I.T.S. PRS Phoenix II is a set of metal plates and screws used to hold broken pelvic bones in place while they heal. The plates are made of titanium and are shaped to fit the pelvis. The set includes symphysis, anterior column, posterior column, and other plate types, along with cortical and cancellous screws of several diameters.
- What it’s cleared for
- It is cleared to stabilize one or more pelvic bone fractures in adult patients. The listed uses include symphyseal disruptions, sacroiliac joint disruptions, acetabular and posterior wall fractures, pelvic ring fractures, and revision surgery for pseudoarthrosis, non-unions, and malunions. It is not intended for spinal use.
- Compared with (the predicate)
- How it differs
- The summary states the device is similar to its predicates in material, geometry, design, indications, and operational principles. It does not describe specific differences from the predicate devices.
- Testing the summary lists
- Engineering analysis (finite element analysis) per ASTM F382 to compare load capacity with the cleared device
- Screw testing per ASTM F543 compared to a 2024 FDA guideline for non-spinal metallic bone screws
- Sterility and bioburden (ISO 11137-1, -2, -3), endotoxin (Ph. Eur. 2.6.14 / USP <85>), and cytotoxicity (ISO 10993-5) on worst-case products
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K260054 took 106 calendar days from receipt to decision. Of the latest 25 other decisions in product code HRS, 5 took longer, 20 were faster; their median was 82 days.
Latest decisions in HRS
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262939 | HERO Sternal Closure SystemHero Orthopedics, LLC · Traditional | Nov 2, 202673 days |
| K263602 | Arthrex Plates, Screws and Pins, and WashersArthrex, Inc. · Special | Oct 29, 202629 days |
| K261135 | Trochlear Salvage Plating SystemSkeletal Dynamics, Inc. · Traditional | Oct 26, 2026202 days |
| K262779 | ANTHEM™ Fracture SystemGlobus Medical, Inc. · Traditional | Oct 23, 202679 days |
| K263407 | Tempo Plating SystemPace Surgical · Special | Oct 23, 202628 days |
| K263269 | Park's Pectus SystemTdm Co., Ltd. · Special | Oct 15, 202627 days |
| K261611 | TDM Large Bone Plate and Screw SystemTdm Co., Ltd. · Special | Oct 2, 2026140 days |
| K262649 | Alians ForearmNewclip Technics · Traditional | Sep 28, 202660 days |
| K261892 | ProRib Fixation SystemCosta Surgical, Inc. · Traditional | Sep 23, 2026110 days |
| K262098 | EnHAnce PEEK Bunion SystemNvision Biomedical Technologies · Special | Sep 18, 202688 days |
Other 510(k)s from I.T.S. GmbH
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260958 | I.T.S. Clavicle Plates with Angular Stability; I.T.S. Volar Radius Plate with Angular Stability; I.T.S. Humeral Head Plate with Angular Stability; I.T.S. FR.O.H. Calcaneus Repair System; I.T.S. Pilonplate with Angular Stability; I.T.S. Olecranonplate with Angular Stability; I.T.S. Straight Plate with Angular Stability; I.T.S. Screw System; I.T.S. PRS Sacral Rod System; I.T.S. Fibula Plate PROlock with Angular Stability; I.T.S. Distal Humeral Plates with Angular Stability; I.T.S. LRS (LI.T.S. GmbH · Traditional | Jun 16, 202685 days |
| K233134 | I.T.S. INS Proximal Femur NailI.T.S. GmbH · Traditional | Feb 12, 2024138 days |
| K210935 | I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix)I.T.S. GmbH · Traditional | Aug 31, 2022520 days |
| K142418 | I.T.S. Hand Locking Plates System - HLSI.T.S. GmbH · Traditional | Sep 26, 201429 days |
| K132945 | CFN-CANNULATED FEMUR NAIL, CTN-CANNULATED TIBIA NAIL, CHN-CANNULATED HUMERAL NAILI.T.S. GmbH · Traditional | Mar 7, 2014169 days |
| K131722 | I.T.S. EXTREMITY FIXATION SYSTEMSI.T.S. GmbH · Traditional | Aug 7, 201356 days |
| K130008 | DISTAL ULNA LOCKING PLATE (DUL), ULNA OSTEOTOMY LOCKING PLATE (UOL), 3.0 CANCELLOUS LOCKING SCREW, 3.0 CORTICAL LOCKINGI.T.S. GmbH · Traditional | Jun 25, 2013174 days |
Watch HRS in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get HRS every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in HRS. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K260054Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:HRSRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:HRSRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:HRS AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"I.T.S. GmbH"Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil